Design of a Dedicated Five Degree-of-Freedom Magnetic Resonance Imaging Compatible Robot for Image Guided Prostate Biopsy

被引:10
|
作者
Chen, Longquan [1 ]
Paetz, Torben [1 ]
Dicken, Volker [1 ]
Krass, Scheherazade [1 ]
Al Issawi, Jumana [1 ]
Ojdanic, Darko [1 ]
Krass, Stefan [1 ]
Tigelaar, Gerrit [2 ]
Sabisch, Jan [2 ]
van Poelgeest, Auguste [3 ]
Schaechtele, Jonathan [3 ]
机构
[1] Fraunhofer MEVIS, D-28359 Bremen, Germany
[2] Soteria Med BV, NL-6825 MC Arnhem, Netherlands
[3] Fraunhofer IPA, D-68167 Mannheim, Germany
关键词
MANIPULATOR; SYSTEM;
D O I
10.1115/1.4029506
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In order to improve the current clinical application of magnetic resonance (MR)-guided prostate biopsies, a new, fully magnetic resonance imaging (MRI)-compatible solution has been developed. This solution consists of a five degree-of-freedom (5DOF) pneumatic robot, a programmable logic controller (PLC), and a software application for visualization and robot control. The robot can be freely positioned on the MR table. For the calibration of the robot and MR coordinate system, the robot's needle guide (NG) is used. The software application supports the calibration with image segmentation and graphic overlays and guides the user through the interventional planning process. After selecting a target point, the application calculates the needed movements via solving the kinematics of the robot and translating the adjustment into commands for the PLC driving the step motors of the robot. In case further adjustments are required, the software also allows for manual control of the robot, to position the NG according to the acquired MR images.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Transperineal magnetic resonance image guided prostate biopsy
    D'Amico, AV
    Tempany, CM
    Cormack, R
    Hata, N
    Jinzaki, M
    Tuncali, K
    Weinstein, M
    Richie, JP
    JOURNAL OF UROLOGY, 2000, 164 (02): : 385 - 387
  • [2] An image-guided magnetic resonance-compatible surgical robot
    Sutherland, Garnette R.
    Latour, Isabelle
    Greer, Alexander D.
    Fielding, Tim
    Feil, Georg
    Newhook, Perry
    NEUROSURGERY, 2008, 62 (02) : 286 - 292
  • [3] Magnetic resonance imaging-guided transperineal prostate biopsy
    Chen, Jie
    Lin, Zhengyu
    Chen, Jin
    Lin, Qingfeng
    Chen, Jian
    Yan, Yuan
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (02) : 394 - 397
  • [5] Magnetic resonance imaging-guided prostate biopsy——A review of literature
    Kulthe Ramesh Seetharam Bhat
    Srinivas Samavedi
    Marcio Covas Moschovas
    Fikret Fatih Onol
    Shannon Roof
    Travis Rogers
    Vipul RPatel
    Ananthakrishnan Sivaraman
    Asian Journal of Urology, 2021, (01) : 105 - 116
  • [6] Magnetic Resonance Imaging-Guided Prostate Biopsy: Present and Future
    Kim, Chan Kyo
    KOREAN JOURNAL OF RADIOLOGY, 2015, 16 (01) : 90 - 98
  • [7] Robot for Magnetic Resonance Imaging Guided Focal Prostate Laser Ablation
    Squires, Alexander
    Xu, Sheng
    Seifabadi, Reza
    Chen, Yue
    Agarwal, Harsh
    Bernardo, Marcelino
    Negussie, Ayele
    Pinto, Peter
    Choyke, Peter
    Wood, Bradford
    Tse, Zion Tsz Ho
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2016, 10 (03):
  • [8] Environmental Impact of Prostate Magnetic Resonance Imaging and Transrectal Ultrasound Guided Prostate Biopsy
    Leapman, Michael S.
    Thiel, Cassandra L.
    Gordon, Ilyssa O.
    Nolte, Adam C.
    Perecman, Aaron
    Loeb, Stacy
    Overcash, Michael
    Sherman, Jodi D.
    EUROPEAN UROLOGY, 2023, 83 (05) : 463 - 471
  • [9] Screening and Detection Advances in Magnetic Resonance Image-Guided Prostate Biopsy
    Stephenson, Samuel K.
    Chang, Edward K.
    Marks, Leonard S.
    UROLOGIC CLINICS OF NORTH AMERICA, 2014, 41 (02) : 315 - +
  • [10] Multiparametric magnetic resonance imaging and multiparametric magnetic resonance imaging-guided biopsy in the diagnostic pathway of prostate cancer
    Chaloupka, Michael
    Apfelbeck, Maria
    Pfitzinger, Paulo
    Bischoff, Robert
    Lellig, Ekaterina
    Rath, Lukas
    Schlenker, Boris
    Stief, Christian G.
    Clevert, Dirk-Andre
    RADIOLOGE, 2020, 60 (Suppl 1): : 63 - 69