Tantalum effects on the ICP-OES determination of trace elements in tantalum powder

被引:0
|
作者
Anil, G [1 ]
Reddy, MRP [1 ]
Kumar, A [1 ]
Prakash, TL [1 ]
机构
[1] Ctr Mat Elect Technol, IDA, Hyderabad 500051, Andhra Pradesh, India
关键词
D O I
暂无
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In this study, the trace elements Zr, Mo, Cu, Mn, Mg, Co, W, Ca, B, and Al are determined in tantalum powder, using a 0.64-m monochromator inductively coupled plasma spectrometer. The effects of tantalum in the determination of trace level impurities were studied by determining these elements in solutions containing tantalum from 1000 to 20,000 mg/L. As the concentration of tantalum is increased, the Zr, Mo, Co, W, and Al decreases in intensity counts while Cu and Mn showed an increase in intensity counts. In the case of boron, there was no change in intensity. For Ca and Mg, +1 states showed an increase in intensity and +2 states showed a decrease. The relative standard intensities of trace elements in the tantalum matrix were calculated from 1000 to 20,000 mg/L tantalum solutions. Spectral line profiles of the trace elements and spectral overlaps of tantalum at 1000 mg/L for interference-free line selection is presented and discussed.
引用
收藏
页码:185 / 189
页数:5
相关论文
共 50 条
  • [1] Matrix effects of tantalum in the determination of Cr, Fe, Nb, Ni, and Ti in tantalum powder using ICP-OES
    Anil, G
    Reddy, MRP
    Kumar, A
    Prakash, TL
    ATOMIC SPECTROSCOPY, 2002, 23 (04) : 119 - 124
  • [2] Determination of silicon in tantalum and its compounds using ICP-OES
    Anil, G
    Reddy, MRP
    Kumar, A
    Prakash, TL
    CHEMICAL PAPERS, 2004, 58 (03): : 195 - 199
  • [3] ICP-OES determination of niobium, tantalum, and titanium at trace to percentage levels in varying geological matrices
    Saran, R
    Khorge, CR
    Premadas, A
    Kumar, V
    ATOMIC SPECTROSCOPY, 2004, 25 (05) : 226 - 231
  • [4] Determination of Trace Elements in Turkish Wines by ICP-OES and HG-ICP-OES
    Aydin, Isil
    Yuksel, Uyan
    Guzel, Remziye
    Ziyadanogullari, Berrin
    Aydin, Firat
    ATOMIC SPECTROSCOPY, 2010, 31 (02) : 67 - 71
  • [5] ICP-OES determination of trace elements in salinated water samples
    Mitko, K
    Bebek, M
    ATOMIC SPECTROSCOPY, 1999, 20 (06) : 217 - 223
  • [6] Comparison of ICP-MS and ICP-OES in the Determination of Trace Elements in Water
    Cidu, R.
    Atomic Spectroscopy, 17 (04):
  • [8] Analysis of tantalum loaded tributyl phosphate for the content of major, minor and trace elements using microwave decomposition and ICP-OES
    Anil, G
    Reddy, MRP
    Sharma, MK
    Kumar, A
    Prakash, TL
    CHEMIA ANALITYCZNA, 2004, 49 (04): : 459 - 465
  • [9] Method of low tantalum amounts determination in niobium and its compounds by ICP-OES technique
    Smolik, Marek
    Turkowska, Magdalena
    TALANTA, 2013, 115 : 184 - 189
  • [10] Determination of tantalum and nitrogen for semiconductor copper metallization technology by ICP-OES and its application to XRF
    Onuma, M
    Takenaka, M
    Yabuki, M
    Hayashi, M
    BUNSEKI KAGAKU, 2003, 52 (06) : 475 - 480