Comparisons of lesion detectability in ultrasound images acquired using time-shift compensation and spatial compounding

被引:10
|
作者
Lacefield, JC [1 ]
Pilkington, WC
Waag, RC
机构
[1] Univ Western Ontario, Dept Elect & Comp Engn, London, ON N6A 5B9, Canada
[2] Univ Western Ontario, Dept Med Biophys, London, ON N6A 5B9, Canada
[3] Robarts Res Inst, Imaging Res Labs, London, ON N6A 5K8, Canada
[4] Univ Rochester, Dept Elect & Comp Sci, Rochester, NY 14627 USA
[5] Univ Rochester, Dept Elect & Comp Engn, Rochester, NY 14627 USA
[6] Univ Rochester, Dept Radiol, Rochester, NY 14627 USA
关键词
D O I
10.1109/TUFFC.2004.1386682
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The effects of aberration, time-shift compensation, and spatial compounding on the discrimination of positive-contrast lesions in ultrasound b-scan images are investigated using a two-dimensional (2-D) array system and tissue-mimicking phantoms. Images were acquired within an 8.8 x 12-mm(2) field of view centered on one of four statistically similar 4-mm diameter spherical lesions. Each lesion was imaged in four planes offset by successive 45degrees rotations about the central scan line. Images of the lesions were acquired using conventional geometric focusing through a water path, geometric focusing through a 35-mm thick distributed aberration phantom, and time-shift compensated transmit and receive focusing through the aberration phantom. The views of each lesion were averaged to form sets of water path, aberrated, and time-shift compensated 4:1 compound images and 16:1 compound images. The contrast ratio and detectability index of each image were computed to assess lesion differentiation. In the presence of aberration representative of breast or abdominal wall tissue, time-shift compensation provided statistically significant improvements of contrast ratio but did not consistently affect the detectability index, and spatial compounding significantly increased the detectability index but did not alter the contrast ratio. Time-shift compensation and spatial compounding thus provide complementary benefits to lesion detection.
引用
收藏
页码:1649 / 1659
页数:11
相关论文
共 7 条
  • [1] A portable measuring system for cross-sectional ultrasound images using spatial compounding and edge sharpening
    Tsubai, Masayoshi
    Fukumoto, Kiyotaka
    Fukuda, Osamu
    Muraki, Satoshi
    Ueno, Naohiro
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 2150 - +
  • [2] 3D spatial compounding of ultrasound images using image-based nonrigid registration
    Krücker, JF
    Meyer, CR
    LeCarpentier, GL
    Fowlkes, JB
    Carson, PL
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (09): : 1475 - 1488
  • [3] TIME-SHIFT COMPENSATION OF ULTRASONIC PULSE FOCUS DEGRADATION USING LEAST-MEAN-SQUARE ERROR-ESTIMATES OF ARRIVAL TIME
    LIU, DL
    WAAG, RC
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 95 (01): : 542 - 555
  • [4] CORRECTION OF ULTRASONIC WAVE-FRONT DISTORTION USING BACKPROPAGATION AND A REFERENCE WAVE-FORM METHOD FOR TIME-SHIFT COMPENSATION
    LIU, DL
    WAAG, RC
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 96 (02): : 649 - 660
  • [5] Compressive sensing of wind speed data of large-scale spatial structures with dedicated dictionary using time-shift strategy
    Wan, Hua-Ping
    Dong, Guan-Sen
    Luo, Yaozhi
    MECHANICAL SYSTEMS AND SIGNAL PROCESSING, 2021, 157
  • [6] Focus-shift compensation in high intensity focused ultrasound using the time reversal method
    Lin, Weijun (linwj@mail.ioa.ac.cn), 2017, Science Press (42):
  • [7] Dynamic 18F-FDG PET Images Simulation Using 4D-XCAT Phantom and Kinetic Modeling for Lesion Detectability Investigation and Scan Time Reduction Purpose
    Bezoubiri, Fethi
    Zidi, Tahar
    Kharfi, Fayçal