Safety and Efficacy of Leadless Pacemakers: A Systematic Review and Meta-Analysis

被引:47
|
作者
Ngo, Linh [1 ,2 ,3 ]
Nour, Daniel [2 ]
Denman, Russell A. [2 ]
Walters, Tomos E. [1 ,4 ]
Haqqani, Haris M. [1 ,2 ]
Woodman, Richard J. [5 ]
Ranasinghe, Isuru [1 ,2 ]
机构
[1] Univ Queensland, Sch Clin Med, Brisbane, Qld, Australia
[2] Prince Charles Hosp, Dept Cardiol, Brisbane, Qld, Australia
[3] E Hosp, Cardiovasc Ctr, Hanoi, Vietnam
[4] St Vincents Private Hosp Northside, Brisbane, Qld, Australia
[5] Flinders Univ S Australia, Flinders Ctr Epidemiol & Biostat, Adelaide, SA, Australia
来源
关键词
efficacy; leadless pacemaker; meta-analysis; safety; systematic review; TRICUSPID REGURGITATION; CARDIAC-PACEMAKER; IMPLANTATION; PERFORMANCE; COMPLICATIONS; EXPERIENCE;
D O I
10.1161/JAHA.120.019212
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Leadless pacemaker is a novel technology, and evidence supporting its use is uncertain. We performed a systematic review and meta-analysis to examine the safety and efficacy of leadless pacemakers implanted in the right ventricle. Methods and Results We searched PubMed and Embase for studies published before June 6, 2020. The primary safety outcome was major complications, whereas the primary efficacy end point was acceptable pacing capture threshold (<= 2 V). Pooled estimates were calculated using the Freedman-Tukey double arcsine transformation. Of 1281 records screened, we identified 36 observational studies of Nanostim and Micra leadless pacemakers, with most (69.4%) reporting outcomes for the Micra. For Micra, the pooled incidence of complications at 90 days (n=1608) was 0.46% (95% CI, 0.08%-1.05%) and at 1 year (n=3194) was 1.77% (95% CI, 0.76%-3.07%). In 5 studies with up to 1-year follow-up, Micra was associated with 51% lower odds of complications compared with transvenous pacemakers (3.30% versus 7.43%; odds ratio [OR], 0.49; 95% CI, 0.34-0.70). At 1 year, 98.96% (95% CI, 97.26%-99.94%) of 1376 patients implanted with Micra had good pacing capture thresholds. For Nanostim, the reported complication incidence ranged from 6.06% to 23.54% at 90 days and 5.33% to 6.67% at 1 year, with 90% to 100% having good pacing capture thresholds at 1 year (pooled result not estimated because of the low number of studies). Conclusions Most studies report outcomes for the Micra, which is associated with a low risk of complications and good electrical performance up to 1-year after implantation. Further data from randomized controlled trials are needed to support the widespread adoption of these devices in clinical practice.
引用
收藏
页数:22
相关论文
共 50 条
  • [1] Efficacy and Safety of Leadless Pacemakers for Atrioventricular Synchronous Pacing: A Systematic Review and Meta-Analysis
    Wu, Sijin
    Jin, Yuanhao
    Lu, Wenzhao
    Chen, Zhongli
    Dai, Yan
    Chen, Keping
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (07)
  • [2] EFFICACY AND SAFETY OF LEADLESS PACEMAKER: AN UPDATED SYSTEMATIC REVIEW AND META-ANALYSIS
    Menezes, Antonio, Jr.
    Oliveira, Vinicius
    Rivera, Andre
    Oliveira, Izadora
    Nishikubo, Maria Elisa
    Sousa, Andre
    Serpa, Frans
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 83 (13) : 269 - 269
  • [3] Comparison of Safety of Leadless Pacemakers and Transvenous Pacemakers: A Meta-Analysis
    Gangannapalle, Mahesh
    Monday, Obinna
    Rawat, Anurag
    Nwoko, Ugonna A.
    Mandal, Arun Kumar
    Babur, Maham
    Khan, Tayyaba J.
    Palleti, Sujith K.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (09)
  • [4] The Effectiveness and Safety of Leadless Pacemakers: An Updated Meta-Analysis
    Oliveira, Vinicius Martins Rodrigues
    Rivera, Andre
    Oliveira, Izadora Caiado
    de Sousa, Andre Maroccolo
    Nishikubo, Maria Elisa Passos
    Serpa, Frans
    Menezes Junior, Antonio de Silva
    CURRENT CARDIOLOGY REPORTS, 2024, 26 (08) : 789 - 799
  • [5] Comparative assessment of safety with leadless pacemakers compared to transvenous pacemakers: a systemic review and meta-analysis
    Shtembari, Jurgen
    Shrestha, Dhan Bahadur
    Awal, Shila
    Raut, Anuradha
    Gyawali, Pratik
    Abe, Temidayo
    Patel, Nimesh K.
    Deshmukh, Abhishek
    Voruganti, Dinesh
    Bhave, Prashant Dattatraya
    Whalen, Patrick
    Pothineni, Naga Venkata K.
    Shantha, Ghanshyam
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2023, 66 (09) : 2165 - 2175
  • [6] Comparative assessment of safety with leadless pacemakers compared to transvenous pacemakers: a systemic review and meta-analysis
    Jurgen Shtembari
    Dhan Bahadur Shrestha
    Shila Awal
    Anuradha Raut
    Pratik Gyawali
    Temidayo Abe
    Nimesh K. Patel
    Abhishek Deshmukh
    Dinesh Voruganti
    Prashant Dattatraya Bhave
    Patrick Whalen
    Naga Venkata K. Pothineni
    Ghanshyam Shantha
    Journal of Interventional Cardiac Electrophysiology, 2023, 66 : 2165 - 2175
  • [7] Safety and Efficacy Outcomes of Micra Leadless Pacemakers: An Updated Systematic Review and Meta-Analaysis
    Haseeb, Shahan
    Ansari, Umair A.
    Chhetri, Chandra
    Masoud, Ahmed T.
    Sohail, Hamza
    Haseeb, Shawal
    Munir, Shafia
    Zulqarnain, Muhammad
    Goyal, Rajat
    CIRCULATION, 2023, 148
  • [8] CLINICAL OUTCOMES OF LEADLESS VERSUS CONVENTIONAL PACEMAKERS: A META-ANALYSIS AND SYSTEMATIC REVIEW
    Ferreira, Andre Luiz Carvalho
    Gonzalez, Maria E. Benitez
    Quezada, Miguel Treminio
    Guida, Camila Mota
    Lima, Ana Emanuela
    Neves, Henrique Alexsander
    De Carvalho, Guilherme Dagostin
    Baddi, Ajay
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 83 (13) : 44 - 44
  • [9] Intracardiac leadless pacemaker: long term safety over traditional pacemakers. a systematic review and meta-analysis
    Gharehdaghi, S.
    Engh, M.
    Veres, B.
    Vigh, N.
    Fehervari, P.
    Hegyi, P.
    Zima, E.
    Duray, G.
    EUROPEAN HEART JOURNAL, 2024, 45
  • [10] Efficacy and safety of tigecycline: a systematic review and meta-analysis
    Yahav, Dafna
    Lador, Adi
    Paul, Mical
    Leibovici, Leonard
    JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2011, 66 (09) : 1963 - 1971