PurposeTo determine the optimal flip angle (FA) for cardiac cine imaging that maximizes myocardial signal and blood-myocardium contrast. MethodsBloch equation simulations of stationary myocardium and flowing blood with an imperfect slice profile were compared to in vivo measurements of blood and myocardium signal-to-noise ratio (SNR) and blood-myocardium contrast-to-noise ratio (CNR) in healthy subjects (N=10) in the short-axis and four-chamber views and in patients (N=7) in the three-chamber imaging plane. ResultsLeft ventricular (LV) and right ventricular (RV) blood SNR and blood-myocardium CNR increases with increasing FA up to approximate to 105 degrees in the short-axis view. A similar trend is seen in the RV four-chamber view, but a marked SNR difference between the LV and RV blood appears for FA>75 degrees, especially during systole. Notable RV and LV SNR and CNR differences are also evident in the three-chamber view due to the predominant LV in-plane flow versus RV through-plane flow. ConclusionVery high blood-myocardium CNR can be obtained with a FA of approximate to 105 degrees in the short-axis plane and approximate to 75 degrees in the three-chamber and four-chamber imaging planes. However, if through-plane flow is limited, as may occur for patients with low ejection fraction or low heart rates, then the FA may be limited to approximate to 75 degrees. Magn Reson Med 73:1095-1103, 2015. (c) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:1095 / 1103
页数:9
相关论文
共 50 条
  • [1] Flip angle optimization for balanced SSFP: Cardiac cine imaging following the application of standard extracellular contrast agent (gadobutrol)
    Kuetting, Daniel L. R.
    Dabir, Darius
    Luetkens, Julian
    Feisst, Andreas
    Homsi, Rami
    Thomas, Daniel
    Schild, Hans H.
    Sprinkart, Alois M.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2018, 47 (01) : 255 - 261
  • [2] Free-Breathing Variable Flip Angle Balanced SSFP Cardiac Cine Imaging with Reduced SAR at 3T
    Srinivasan, Subashini
    Kroeker, Randall M.
    Gabriel, Simon
    Plotnik, Adam
    Godinez, Sergio R.
    Hu, Peng
    Halnon, Nancy
    Finn, J. Paul
    Ennis, Daniel B.
    MAGNETIC RESONANCE IN MEDICINE, 2016, 76 (04) : 1210 - 1216
  • [3] Variable Flip Angle Balanced Steady-State Free Precession for Lower SAR or Higher Contrast Cardiac Cine Imaging
    Srinivasan, Subashini
    Ennis, Daniel B.
    MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (03) : 1035 - 1043
  • [4] Improved myocardial tagging contrast in cine balanced SSFP images
    Ibrahim, El-Sayed H.
    Stuber, Matthias
    Schar, Michael
    Osman, Nael F.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 24 (05) : 1159 - 1167
  • [5] Design and Use of Variable Flip Angle Schedules in Transient Balanced SSFP Subtractive Imaging
    Smith, Travis
    Zun, Zungho
    Wong, Eric C.
    Nayak, Krishna S.
    MAGNETIC RESONANCE IN MEDICINE, 2010, 63 (02) : 537 - 542
  • [6] Prediction of myocardial signal during CINE balanced SSFP imaging
    Kyunghyun Sung
    Hsu-Lei Lee
    Houchun H. Hu
    Krishna S. Nayak
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2010, 23 : 85 - 91
  • [7] Prediction of myocardial signal during CINE balanced SSFP imaging
    Sung, Kyunghyun
    Lee, Hsu-Lei
    Hu, Houchun H.
    Nayak, Krishna S.
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2010, 23 (02) : 85 - 91
  • [8] A technique for improving tag contrast persistence in SSFP MRI imaging using adaptive flip angle
    Ibrahim, ESH
    Osman, NF
    2004 2ND IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1 AND 2, 2004, : 1051 - 1054
  • [9] Slice encoding for the reduction of outflow signal artifacts in cine balanced SSFP imaging
    Ali, Fadil
    Bydder, Mark
    Han, Hui
    Wang, Da
    Ghodrati, Vahid
    Gao, Chang
    Prosper, Ashley
    Nguyen, Kim-Lien
    Finn, J. Paul
    Hu, Peng
    MAGNETIC RESONANCE IN MEDICINE, 2021, 86 (04) : 2034 - 2048
  • [10] Balanced SSFP Transient Imaging Using Variable Flip Angles for a Predefined Signal Profile
    Worters, Pauline W.
    Hargreaves, Brian A.
    MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (05) : 1405 - 1413