Feasibility and Safety of MR-Guided Focused Ultrasound Lesioning in the Setting of Deep Brain Stimulation

被引:4
|
作者
Dallapiazza, Robert [1 ]
Khaled, Mohamad [1 ]
Eames, Matthew [4 ]
Snell, John [4 ]
Lopes, M. Beatriz [2 ]
Wintermark, Max [3 ]
Elias, W. Jeffrey [1 ]
机构
[1] Univ Virginia, Dept Neurosurg, Charlottesville, VA USA
[2] Univ Virginia, Dept Neuropathol, Charlottesville, VA USA
[3] Univ Virginia, Dept Radiol, Charlottesville, VA USA
[4] Focused Ultrasound Surg Fdn, Charlottesville, VA USA
关键词
Focused ultrasound; Thalamotomy; Magnetic resonance imaging compatibility; Deep brain stimulation; Movement disorders; Thermography; ESSENTIAL TREMOR; DBS; SYSTEMS; DAMAGE; SKULL;
D O I
10.1159/000368908
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Background: Patients treated with deep brain stimulation (DBS) often develop symptom progression. If safe, focused ultrasound (FUS) lesioning could be used for patients unable to undergo further DBS surgery. Objective: To test the feasibility and safety of MR-guided FUS surgery in the setting of a previously implanted DBS system. Methods:Three preclinical experiments were designed to test feasibility and safety. Hydrogels were implanted with an electrode, and FUS lesions were targeted adjacently. Cadavers were implanted with a thalamic electrode, and FUS lesions were targeted in the contralateral thalamus. Finally, DBS systems were implanted in swine, and FUS lesioning was targeted to the contralateral thalamus, MRI was used to assess the treatments, and histological analyses were performed at 2 days and at 1 month. Results: In gel experiments and cadavers, FUS resulted in target heating to 29-32 degrees C without any heating at the electrode. In animal experiments, there were no FUS-related MRI signal changes near the electrode. Histological analysis showed typical FUS lesions with no evidence of damage surrounding the electrode tracts. Conclusions: FUS is feasible in the setting of a preimplanted DBS device. There was minimal heating of the device during the procedure and no apparent FUS-related tissue injury. (C) 2015 S. Karger AG, Basel
引用
收藏
页码:140 / 146
页数:7
相关论文
共 50 条
  • [1] Development of an MR-Guided Focused Ultrasound (MRgFUS) Lesioning Approach for the Fornix in the Rat Brain
    Cornelssen, Carena
    Payne, Allison
    Parker, Dennis L.
    Alexander, Matthew
    Merrill, Robb
    Senthilkumar, Sharayu
    Christensen, Jacob
    Wilcox, Karen S.
    Odeen, Henrik
    Rolston, John D.
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2024, 50 (06): : 920 - 926
  • [2] Stereotactic MR-guided focused ultrasound deep brain lesioning: the resurrection of posteroventral pallidotomy and thalamotomy for Parkinson's disease?
    Kinfe, Thomas M.
    ACTA NEUROCHIRURGICA, 2017, 159 (07) : 1367 - 1369
  • [3] Stereotactic MR-guided focused ultrasound deep brain lesioning: the resurrection of posteroventral pallidotomy and thalamotomy for Parkinson’s disease?
    Thomas M. Kinfe
    Acta Neurochirurgica, 2017, 159 : 1367 - 1369
  • [4] MR-guided focused ultrasound pallidotomy for Parkinson's disease: safety and feasibility
    Eisenberg, Howard M.
    Krishna, Vibhor
    Elias, W. Jeffrey
    Cosgrove, G. Rees
    Gandhi, Dheeraj
    Aldrich, Charlene E.
    Fishman, Paul S.
    JOURNAL OF NEUROSURGERY, 2021, 135 (03) : 792 - 798
  • [5] MR-Guided Focused Ultrasound
    Mukherji, Suresh K.
    Bencardino, Jenny T.
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2024, 32 (04) : 0xii - 0xii
  • [6] MR-guided focused ultrasound
    Gedroyc, Wladyslaw M. W.
    Anstee, Ann
    EXPERT REVIEW OF MEDICAL DEVICES, 2007, 4 (04) : 539 - 547
  • [7] MR-Guided Focused Ultrasound Thalamotomy in the Setting of Aneurysm Clip
    Odeen, Henrik
    Shah, Lubdha M.
    Rieke, Viola
    Parker, Dennis L.
    Rahimpour, Shervin
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2024, : 302 - 304
  • [8] Deep brain stimulation, lesioning, focused ultrasound: update on utility
    Reddy, Akshay
    Hosseini, Mohammad Reza
    Patel, Aashay
    Sharaf, Ramy
    Reddy, Vishruth
    Tabarestani, Arman
    Lucke-Wold, Brandon
    AIMS NEUROSCIENCE, 2023, 10 (02) : 87 - 108
  • [9] MR-guided focused ultrasound ablation therapy: Safety control
    Wu, T
    Felmlee, JP
    RADIOLOGY, 2001, 221 : 168 - 168
  • [10] MR-GUIDED FOCUSED ULTRASOUND SURGERY
    CLINE, HE
    SCHENCK, JF
    HYNYNEN, K
    WATKINS, RD
    SOUZA, SP
    JOLESZ, FA