Comparison of various current electronic apex locators to determine the working length using the clearing technique

被引:8
|
作者
Altunbas, D. [1 ]
Kustarci, A. [2 ]
Arslan, D. [3 ]
Er, K. [2 ]
Kocak, S. [4 ]
机构
[1] Cumhuriyet Univ, Sch Dent, Dept Endodont, Sivas, Turkey
[2] Akdeniz Univ, Sch Dent, Dept Endodont, TR-07058 Antalya, Turkey
[3] Bezmialem Vakif Univ, Sch Dent, Dept Endodont, Istanbul, Turkey
[4] Bulent Ecevit Univ, Sch Dent, Dept Endodont, Zonguldak, Turkey
关键词
Clearing technique; electronic apex locator; endodontics; major foramen; ROOT-CANAL LENGTH; IN-VITRO EVALUATION; MEASUREMENT DEVICES; POSTERIOR TEETH; APICAL FORAMEN; VIVO ACCURACY; DENTAPORT ZX; RAYPEX; EX-VIVO; FREQUENCY;
D O I
10.4103/1119-3077.151769
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Correct working length determination is an important step for successful endodontic therapy. Objective: The objective was to compare in vitro the accuracy of three electronic apex locators (EALs) (DentaPort ZX [Morita Co., Tokyo, Japan], SIROEndo Pocket [Sirona Dental Systems, NY, USA], and Rootor [Meta Biomed, Cheongwon-gun, Korea]) in detecting the major foramen using the clearing technique. Materials and Methods: Forty-five human extracted single-rooted teeth with mature apices were used for the study and divided into three groups of 15 teeth each. All teeth were embedded in an alginate model, and the electronic measurements were taken following the manufacturers orientations. Then, the teeth were cleared and photographed under a stereomicroscope with a digital camera. The distance from the file tip to the major foramen was measured using image analysis software program. Statistical analysis was performed using the Kruskal-Wallis, Mann-Whitney U, and Chi-square tests at a significance level of 0.05. Results: The mean distances from the file tip to the major foramen were 0.164 0.292, 0.162 0.234, 0.341 0.166 mm in the DentaPort ZX, SIROEndo Pocket, and Rootor groups, respectively. Statistical analysis showed that there was a significant difference between SIROEndo Pocket and Rootor (P < 0.05). However, no significant difference was found between DentaPort ZX and other EALs (P > 0.05). Conclusion: DentaPort ZX located the major foramen with 100% accuracy within the range of 0.5 mm. However, the accuracy of the SIROEndo Pocket and Rootor in locating the major foramen within 0.5 mm was 73.3% and 86.7%, respectively. All EALs showed an acceptable determination of the major foramen within the range of 0.5 mm.
引用
收藏
页码:359 / 363
页数:5
相关论文
共 50 条
  • [1] In vitro comparison of four different electronic apex locators to determine the major foramen using the clearing technique
    Altunbas, D.
    Kustarci, A.
    Arslan, D.
    Er, K.
    NIGERIAN JOURNAL OF CLINICAL PRACTICE, 2014, 17 (06) : 706 - 710
  • [2] Ex vivo comparison of three electronic apex locators to determine working length in endodontic treatment
    Miguita, Kenner Bruno
    de Araujo, Roberta Aranha
    De Martin, Alexandre Sigrist
    da Silveira Bueno, Carlos Eduardo
    Cunha, Rodrigo Sanches
    ENDO-ENDODONTIC PRACTICE TODAY, 2011, 5 (04): : 281 - 284
  • [3] An Invitro Evaluation of the Accuracy of Two Electronic Apex Locators to Determine Working Length in the Presence of Various Irrigants
    Ashwini, Prasad B.
    Harshit, Srivastava
    Aastha, Srivastava Arora
    Deepak, Raisingani
    ETHIOPIAN JOURNAL OF HEALTH SCIENCES, 2016, 26 (05) : 457 - 462
  • [4] In vivo comparison of working length determination with two electronic apex locators
    Wrbas, K. T.
    Ziegler, A. A.
    Altenburger, M. J.
    Schirrmeister, J. F.
    INTERNATIONAL ENDODONTIC JOURNAL, 2007, 40 (02) : 133 - 138
  • [5] Comparison of working length determination with radiographs and four electronic apex locators
    Vieyra, J. P.
    Acosta, J.
    INTERNATIONAL ENDODONTIC JOURNAL, 2011, 44 (06) : 510 - 518
  • [6] Comparison of working length determination with radiographs and two electronic apex locators
    Vieyra, J. P.
    Acosta, J.
    Mondaca, J. M.
    INTERNATIONAL ENDODONTIC JOURNAL, 2010, 43 (01) : 16 - 20
  • [7] An ex vivo comparison of working length determination by 3 electronic apex locators
    Pascon, Elizeu Alvaro
    Marrelli, Massimo
    Congi, Orsola
    Ciancio, Rosetta
    Miceli, Federica
    Versiani, Marco Aurelio
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2009, 108 (03): : E147 - E151
  • [8] Ex vivo evaluation of the ability of four different electronic apex locators to determine the working length in teeth with various foramen diameters
    Ebrahim, A. K.
    Wadachi, R.
    Suda, H.
    AUSTRALIAN DENTAL JOURNAL, 2006, 51 (03) : 258 - 262
  • [9] Comparison of the Accuracies of CBCT, Radiography, and Four Electronic Apex Locators in Working Length Determination
    Ozata, Merve Yeniceri
    Falakaloglu, Seda
    Kaya, Sadullah
    MAKARA JOURNAL OF HEALTH RESEARCH, 2022, 26 (01): : 61 - 66
  • [10] The use of micro-computed tomography to determine the accuracy of electronic working length with two apex locators
    Suguro, Hisashi
    Nishihara, Anna
    Tamura, Takahito
    Nakamura, Takeshi
    Toyama, Yurika
    Takeichi, Osamu
    JOURNAL OF ORAL SCIENCE, 2021, 63 (02) : 167 - 169