Long term results of the titanium clip prosthesis

被引:17
|
作者
Gostian, Antoniu-Oreste [1 ]
Kouame, Jean-Marc [1 ]
Bremke, Martin [1 ]
Ortmann, Magdalene [2 ]
Huettenbrink, Karl Bernd [1 ]
Beutner, Dirk [1 ]
机构
[1] Univ Cologne, Fac Med, Dept Otorhinolaryngol Head & Neck Surg, D-50924 Cologne, Germany
[2] Jean Uhrmacher Inst Clin ENT Res, Geibelstr 29-31, D-50931 Cologne, Germany
关键词
Partial ossiculoplasty; Clip-prosthesis; Longterm results; Tympanoplasty; OSSICULAR RECONSTRUCTION; OSSICULOPLASTY; IMPLANTS; TYMPANOPLASTY; OPERATIONS; FAILURE; CHAIN; TORPS; PORPS;
D O I
10.1007/s00405-016-4174-3
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
The objective of the study is to evaluate the long term results of ossiculoplasty using the clip titanium partial ossicular replacement prosthesis. This study retrospectively reviews the partial ossiculoplasty conducted using clip titanium partial ossicular replacement prosthesis at a tertiary referral center. Audiometric outcomes and intraoperative findings were postoperatively measured from revision surgery of 47 ears (20 women, 27 men, mean age 43 years) averaging 6.5 years. The overall air-bone gap decreased from 25.7 dB preoperatively to 16.8 dB 6.5 years postoperatively (p aecurrency sign 0.001, eta (p) (2) = 0.210). An air-bone gap of < 20 dB was present in 28 % of ears preoperatively and increased to 72 % postoperatively. In revision (n = 30) and primary tympanoplasties (n = 17), the preoperative air-bone gaps were reduced from 28.9 and 20.1 to 18.7 dB (p aecurrency sign 0.001, eta (p) (2) = 0.240) and 13.2 dB (p = 0.033, eta (p) (2) = 0.192), respectively. In canal wall down (n = 15) procedures and tympanoplasties with intact canal wall (n = 32), the mean air-bone gaps diminished from 28.9 to 18.1 dB (p = 0.02, eta (p) (2) = 0.245) and 24.2 to 16.1 dB (p aecurrency sign 0.001, eta (p) (2) = 0.221), respectively. In our own revision tympanoplasties (n = 8) and second look operations (n = 6), we found that the prostheses were safe to remove without any deleterious effects. Two prostheses were dislocated from the stapes' head due to recurrent cholesteatoma. Therefore, it can be concluded that ossiculoplasty using the clip partial ossicular replacement prosthesis allows for good and reliable long term hearing results. Also, the flexible strips reveal no adverse effects on the encompassed stapes' head and do not complicate revision surgery.
引用
收藏
页码:4257 / 4266
页数:10
相关论文
共 50 条
  • [1] Long term results of the titanium clip prosthesis
    Antoniu-Oreste Gostian
    Jean-Marc Kouame
    Martin Bremke
    Magdalene Ortmann
    Karl Bernd Hüttenbrink
    Dirk Beutner
    European Archives of Oto-Rhino-Laryngology, 2016, 273 : 4257 - 4266
  • [2] The Titanium Angular Clip Prosthesis for Ossicular Chain Reconstruction: Long-Term Results
    Stuermer, Konrad Johannes
    Huettenbrink, Karl-Bernd
    Luers, Jan Christoffer
    OTOLOGY & NEUROTOLOGY, 2020, 41 (06) : 806 - 809
  • [3] Titanium clip prosthesis
    Hüttenbrink, KB
    Zahnert, T
    Wüstenberg, EG
    Hofmann, G
    OTOLOGY & NEUROTOLOGY, 2004, 25 (04) : 436 - 442
  • [4] Titanium Angular Clip: A New Prosthesis for Reconstruction of the Long Process of the Incus
    Huttenbrink, Karl-Bernd
    Luers, Jan Christoffer
    Beutner, Dirk
    OTOLOGY & NEUROTOLOGY, 2009, 30 (08) : 1186 - 1190
  • [5] Good long term results with a cemented straight femoral shaft prosthesis made of titanium
    Eingartner, C
    Ihm, A
    Maurer, F
    Volkmann, R
    Weise, K
    Weller, S
    UNFALLCHIRURG, 2002, 105 (09): : 804 - 810
  • [6] Short and Long-Term Outcomes of Titanium Clip Ossiculoplasty
    Kahue, Charissa N.
    O'Connnell, Brendan P.
    Dedmon, Matthew M.
    Haynes, David S.
    Rivas, Alejandro
    OTOLOGY & NEUROTOLOGY, 2018, 39 (06) : e453 - e460
  • [7] LONG-TERM RESULTS WITH THE DURAMADRE PROSTHESIS
    GRINFELD, R
    RONCORONI, A
    FABRYKANT, F
    REVISTA MEDICA DE CHILE, 1981, 109 (06) : R8 - R8
  • [8] LONG-TERM RESULTS WITH THE KAUFMAN PROSTHESIS
    GRAHAM, SD
    CARSON, CC
    ANDERSON, EE
    JOURNAL OF UROLOGY, 1982, 128 (02): : 328 - 330
  • [9] Total Ossiculoplasty: Short- and Long-Term Results Using a Titanium Prosthesis With Footplate Shoe
    Fayad, Jose N.
    Ursick, Joseph
    Brackmann, Derald E.
    Friedman, Rick A.
    OTOLOGY & NEUROTOLOGY, 2014, 35 (01) : 108 - 113
  • [10] LONG TERM RESULTS AND COMPLICATIONS OF THE UROLUME SPHINCTER PROSTHESIS
    Shenot, Patrick
    Das, Akhil
    NEUROUROLOGY AND URODYNAMICS, 2010, 29 (02) : 321 - 321