Numerical methods of mode selection in continuous-wave cavity ring-down spectroscopy

被引:4
|
作者
Wang Jin-Duo [1 ,2 ]
Yu Jin [1 ,2 ]
Mo Ze-Qiang [1 ,2 ,3 ]
He Jian-Guo [1 ,2 ,3 ]
Dai Shou-Jun [1 ,2 ]
Meng Jing-Jing [1 ]
Wang Xiao-Dong [1 ,2 ]
Liu Yang [1 ,3 ]
机构
[1] Chinese Acad Sci, Acad Optoelect, Beijing 100094, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
[3] Chinese Acad Sci, Key Lab Computat Optic Imaging Technol, Beijing 100094, Peoples R China
关键词
cavity ring-down spectroscopy; cavity misadjustment; multimode decay; measurement sensitivity; DIODE-LASER; HIGH-POWER;
D O I
10.7498/aps.68.20190844
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
In continuous-wave cavity ring-down spectroscopy (CW-CRDS), the measurement sensitivity is seriously affected by the multimode excitations in the ring-down cavity. The using of an intracavity aperture is a common way to restrain the excitation of high-order modes, thus leading the laser power to additionally lose and the signal-to-noise ratio to degrade. In this paper, two numerical methods, named " trigger threshold method" and "curve fitness method" , are proposed for selecting the mode in which the decays excited by the high-order modes can be removed. The laser coupling efficiency between the incident laser and the oscillating fundamental or high-order modes is studied in a misadjusted ring-down cavity. It is found that with a misadjusted ring-down cavity, the laser energy is partially coupled into the high-order modes, and the coupling energy increases with the extent of the cavity misadjustment increasing. In this case, the ring-down decaying traces excited by these high-order modes are different from and much shorter than those excited by the fundamental mode, which are respectively called "bad decays" and "good decays" in this paper. Both the fundamental mode and the high-order modes can reach the threshold in the case of low triggering threshold selection and result in the components of both good and bad decays in the output ring-down curves. When the trigger threshold rises, the bad decays are effectively restrained by the deficient coupling into the high-order modes. Thus raising the trigger threshold is an effective method to restrain bad decays for the mode selection. Another approach is to consider the time spent on turning off the laser injection since the fitting goodness of good decays is better than that of bad decays. In this paper this characteristic is also used to separate the good decays from the bad ones. These two methods are demonstrated in the CW-CRDS experiments. The results show that the sensitivity of the CW-CRDS instrument can be greatly improved by one order of magnitude in the trigger threshold method with the minimum of Allan deviations gradually approaching to a constant, while the acquisition rate of the ring-down decays slows down with the increase of the trigger threshold. The results also explain the relationship between single sampling and averaged sampling, which presents an answer to the question about the sequence choice between averaging and fitting. A numerical model is proposed to estimate the probability of good decays versus the trigger threshold, which can be used to choose appropriate trigger threshold for CW-CRDS experiment. The applicable conditions and the limitations of these two methods in CW-CRDS for trace gas detection are also discussed in the paper.
引用
收藏
页数:11
相关论文
共 23 条
  • [1] Cavity ring-down spectroscopy: Experimental schemes and applications
    Berden, G
    Peeters, R
    Meijer, G
    [J]. INTERNATIONAL REVIEWS IN PHYSICAL CHEMISTRY, 2000, 19 (04) : 565 - 607
  • [2] Absolute frequencies of water lines near 790 nm with 10-11 accuracy
    Chen, J.
    Hua, T. -P.
    Tao, L. -G.
    Sun, Y. R.
    Liu, A. -W.
    Hu, S. -M.
    [J]. JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2018, 205 : 91 - 95
  • [3] Simultaneous measurements of s- and p-polarization reflectivity with a cavity ring-down technique employing no polarization optics
    Cui, Hao
    Li, Bincheng
    Han, Yanling
    Wang, Jing
    Gao, Chunming
    Wang, Yafei
    [J]. CHINESE OPTICS LETTERS, 2017, 15 (05)
  • [4] Simultaneous mapping of reflectance, transmittance and optical loss of highly reflective and anti-reflective coatings with two-channel cavity ring-down technique
    Cui, Hao
    Li, Bincheng
    Xiao, Shilei
    Han, Yanling
    Wang, Jing
    Gao, Chunming
    Wang, Yafei
    [J]. OPTICS EXPRESS, 2017, 25 (05): : 5807 - 5820
  • [5] Analysis and alignment of the light path of Gauss beam matched to the fundamental mode of an optical resonator
    Cui Li-Hong
    Zhao Wei-Ning
    Yan Chang-Xiang
    [J]. ACTA PHYSICA SINICA, 2015, 64 (22)
  • [6] Grand Y. L., 1990, J OPT SOC AM B, V7, P1251
  • [7] Study on Cavity Ring-Down Spectroscopy in High Power Ion Source
    Hu Chun-dong
    Yan Jing-yang
    Wang Yan
    Liang Li-zhen
    [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38 (02) : 346 - 351
  • [8] Noise in cavity ring-down spectroscopy caused by transverse mode coupling
    Huang, Haifeng
    Lehmann, Kevin K.
    [J]. OPTICS EXPRESS, 2007, 15 (14): : 8745 - 8759
  • [9] Ultrasensitive, real-time trace gas detection using a high-power, multimode diode laser and cavity ringdown spectroscopy
    Karpf, Andreas
    Qiao, Yuhao
    Rao, Gottipaty N.
    [J]. APPLIED OPTICS, 2016, 55 (16) : 4497 - 4504
  • [10] Lehmann KK, 2009, FRONTIERS OF MOLECULAR SPECTROSCOPY, P623, DOI 10.1016/B978-0-444-53175-9.00018-0