Study on catalytic adsorptive stripping voltammetry of trace cobalt at bismuth film electrodes

被引:110
|
作者
Krolicka, A
Bobrowski, A [1 ]
Kalcher, K
Mocak, J
Svancara, I
Vytras, K
机构
[1] AGH Univ Sci & Technol, Fac Mat Sci & Ceram, Dept Bldg Mat, PL-30059 Krakow, Poland
[2] Karl Franzens Univ Graz, Inst Chem Analyt Chem, A-8010 Graz, Austria
[3] Slovak Univ Technol Bratislava, Fac Chem & Food Technol, Dept Analyt Chem, SK-91237 Bratislava, Slovakia
[4] Univ Pardubice, Fac Chem Technol, Dept Analyt Chem, CZ-53210 Pardubice, Czech Republic
关键词
bismuth film electrode; catalytic adsorptive stripping voltammetry; cobalt; dimethylglyoxime; nitrite;
D O I
10.1002/elan.200302763
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Bismuth film electrodes (BiFE) prepared by plating a glassy carbon support ex situ from a solution containing 0.5 M LiBr and 1 M HCl was tested for the determination of cobalt traces using adsorptive stripping voltammetry (AdSV) and catalytic adsorptive stripping voltammetry (CAdSV). AdSV enables one to determine cobalt as the Co(II)-dimethylglyoximate (Co-DMG) complex down to 0.3 mug/L Co. Addition of NaNO2 to the solution containing dimethylglyoxime in ammonia buffer provides a 15-fold enhancement of the voltammetric signal of cobalt due to the catalytic effect occurring during the reduction of Co(II)-DMG. Utilization of the bismuth film electrode under optimized conditions assures a stable catalytic adsorptive stripping voltammetric response for Co, with an extremely high sensitivity (1.75 muA/(mug/L)), good precision (RSD=3%), and a low detection limit (0.07 mug/L Co with 60 s of adsorptive accumulation). The results of the measurement of the catalytic cobalt response in the presence of DMG and nitrite under hydrodynamic amperometric conditions have shown the BiFE to be an attractive and suitable tool for the rapid determination of cobalt at low mug/L level in both batch and flow systems.
引用
收藏
页码:1859 / 1863
页数:5
相关论文
共 50 条
  • [1] Bismuth film electrodes for adsorptive stripping voltammetry of trace nickel
    Wang, J
    Lu, JM
    ELECTROCHEMISTRY COMMUNICATIONS, 2000, 2 (06) : 390 - 393
  • [2] Catalytic adsorptive stripping voltammetry at film electrodes
    Bobrowski, Andrzej
    Zarebski, Jerzy
    CURRENT ANALYTICAL CHEMISTRY, 2008, 4 (03) : 191 - 201
  • [3] Catalytic adsorptive stripping voltammetric measurements of trace vanadium at bismuth film electrodes
    Wang, Joseph
    Lu, Donglai
    Thongngamdee, Sompong
    Lin, Yuehe
    Sadik, Omowunmi A.
    TALANTA, 2006, 69 (04) : 914 - 917
  • [4] Trace determination of chromium by square-wave adsorptive stripping voltammetry on bismuth film electrodes
    Chatzitheodorou, E
    Economou, A
    Voulgaropoulos, A
    ELECTROANALYSIS, 2004, 16 (21) : 1745 - 1754
  • [5] Determination of traces of cobalt in zinc matrix by catalytic adsorptive stripping voltammetry at bismuth film electrode
    Korolczuk, Mieczyslaw
    Grabarczyk, Malgorzata
    Moroziewicz, Aleksandra
    ELECTROANALYSIS, 2007, 19 (19-20) : 2155 - 2158
  • [6] Catalytic Adsorptive Stripping Voltammetry of Cobalt in the Presence of Nitrite at an In Situ Plated Bismuth Film Electrode
    Rutyna, Iwona
    Korolczuk, Mieczyslaw
    ELECTROANALYSIS, 2011, 23 (03) : 637 - 641
  • [7] Determination of trace titanium(IV) by differential pulse catalytic adsorptive stripping voltammetry at a bismuth film electrode
    Cui, Hong
    Sun, Wenhua
    Li, Xiangjun
    Zou, Hong
    Yuan, Zhuobin
    ANALYTICAL METHODS, 2013, 5 (07) : 1784 - 1789
  • [8] Determination of subnanomolar concentrations of cobalt by adsorptive stripping voltammetry at a bismuth film electrode
    Mieczyslaw Korolczuk
    Aleksandra Moroziewicz
    Malgorzata Grabarczyk
    Analytical and Bioanalytical Chemistry, 2005, 382 : 1678 - 1682
  • [9] Determination of subnanomolar concentrations of cobalt by adsorptive stripping voltammetry at a bismuth film electrode
    Korolczuk, M
    Moroziewicz, A
    Grabarczyk, M
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2005, 382 (07) : 1678 - 1682
  • [10] High sensitive determination of trace amount of cobalt by catalytic adsorptive stripping voltammetry
    Bahrami, S.
    Abbasi, S.
    Ghorbani, Y. A.
    Miran-Beigi, A. A.
    RUSSIAN JOURNAL OF ELECTROCHEMISTRY, 2009, 45 (02) : 208 - 214