On-Site Measurement of Trace-Level Sulfide in Natural Waters by Vapor Generation and Microchannel Collection

被引:26
|
作者
Toda, Kei [1 ]
Kuwahara, Haruka [1 ]
Ohira, Shin-Ichi [1 ]
机构
[1] Kumamoto Univ, Dept Chem, Kumamoto 8608555, Japan
基金
日本学术振兴会;
关键词
ATMOSPHERIC HYDROGEN-SULFIDE; GAS-PHASE CHEMILUMINESCENCE; FIELD INSTRUMENT; SOLID SAMPLES; SULFUR; SEDIMENTS; ANALYZER; LIQUID; CHROMATOGRAPHY; METHANETHIOL;
D O I
10.1021/es2006226
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Aqueous sulfide plays an important role in the environment even at low concentrations. However, it is unstable, which means field samples cannot be transported to the laboratory for analysis without fixation. In this work, a novel method was developed to determine trace levels of sulfide on site. This method is based on vapor generation and collection in a special microchannel device followed by fluorescence measurement (VG-mu GAS). The microchannel scrubber gave a high enrichment factor, and a high sensitivity was achieved, which allowed measurement of nanomolar (nM) levels of sulfide. The theoretical approach to vapor generation for several compounds is discussed to evaluate the applicability of the method to these analytes, and compounds having a low Henry's law constant (<1 M atm(-1)) are suitable to measure by VG-mu GAS. Under optimized conditions, concentrations of 1.0-100 nM of sulfide could be measured. The sulfide contents of hot spring, aquarium, pond, and seawater were successfully measured by this method. Nanomolar levels of sulfide could be measured on site without loss of analyte, and results were obtained instantly in the field, both of which are advantageous for effective field surveys. The method was also applied to field measurements of aqueous sulfide in the Ariake Sea and Lake Baikal.
引用
收藏
页码:5622 / 5628
页数:7
相关论文
共 10 条
  • [1] On-site and high-resolution spectrophotometric measurement of total dissolved sulfide in natural waters
    Li, Peng
    Li, Songtao
    Yuan, Dongxing
    Lin, Kunde
    SCIENCE OF THE TOTAL ENVIRONMENT, 2023, 863
  • [2] Improving the measurement of total dissolved sulfide in natural waters: A new on-site flow injection analysis method
    Li, Peng
    Yuan, Dongxing
    Huang, Yongming
    Lin, Kunde
    SCIENCE OF THE TOTAL ENVIRONMENT, 2022, 829
  • [3] Simple Field Device for Measurement of Dimethyl Sulfide and Dimethylsulfoniopropionate in Natural Waters, Based on Vapor Generation and Chemiluminescence Detection
    Nagahata, Takanori
    Kajiwara, Hidetaka
    Ohira, Shin-Ichi
    Toda, Kei
    ANALYTICAL CHEMISTRY, 2013, 85 (09) : 4461 - 4467
  • [4] Trace-level determination of pharmaceutical residues by LC-MS/MS in natural and treated waters. A pilot-survey study
    M. D. Hernando
    E. Heath
    M. Petrovic
    D. Barceló
    Analytical and Bioanalytical Chemistry, 2006, 385 : 985 - 991
  • [5] Trace-level determination of pharmaceutical residues by LC-MS/MS in natural and treated waters.: A pilot-survey study
    Hernando, M. D.
    Heath, E.
    Petrovic, M.
    Barcelo, D.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 385 (06) : 985 - 991
  • [6] Direct Determination of Trace Antimony in Natural Waters by Photochemical Vapor Generation ICPMS: Method Optimization and Comparison of Quantitation Strategies
    Gao, Ying
    Sturgeon, Ralph E.
    Mester, Zoltan
    Hon, Xiandeng
    Zheng, Chengbin
    Yang, Lu
    ANALYTICAL CHEMISTRY, 2015, 87 (15) : 7996 - 8004
  • [7] On-site monitoring of dissolved Sb species in natural waters by an automatic system using flow injection coupled with hydride generation atomic fluorescence spectrometer
    Chen, Luodan
    Zhong, Ziyun
    Wu, Rongkun
    Lin, Qinglin
    Gong, Zhenbin
    Yuan, Dongxing
    TALANTA, 2024, 274
  • [8] A Cu/Nafion/Bi electrode for on-site monitoring of trace heavy metals in natural waters using anodic stripping voltammetry: An alternative to mercury-based electrodes
    Legeai, S
    Vittori, O
    ANALYTICA CHIMICA ACTA, 2006, 560 (1-2) : 184 - 190
  • [9] A portable/miniaturized analytical kit for on-site analysis: Chemical vapor generation-visual colorimetric and smartphone RGB dual-mode for detection of sulfide ion in water and food additives
    Jiang, Chenxi
    Chen, Jihong
    Tang, Jiayuan
    Xiao, Jing
    Xu, Fengzhou
    Luo, Hong
    Huang, Ke
    Zou, Zhirong
    FOOD CHEMISTRY, 2024, 444
  • [10] A simple and rapid head space-single drop microextraction-'spectro-pipette' (HS-SDME-SP) method for the on-site measurement of arsenic species in natural waters
    Leng, Geng
    Lin, Lin
    Worsfold, Paul J.
    Xu, Wenbo
    Luo, Xin
    Chang, Le
    Li, Weiyi
    Zhang, Xiaoyan
    Xia, Chuanxi
    MICROCHEMICAL JOURNAL, 2021, 168