Comparison of transmission and internal reflection infrared spectra of cocaine

被引:0
|
作者
Koulis, CV
Reffner, JA
Bibby, AM
机构
[1] Illinois State Police, Joliet Forens Sci Lab, Joliet, IL 60432 USA
[2] SensIR Technol, Danbury, CT USA
关键词
forensic science; chemistry; infrared spectroscopy; attenuated total reflection; internal reflection spectroscopy; transmission spectroscopy; cocaine salt; cocaine base; drug analysis; dispersion;
D O I
暂无
中图分类号
DF [法律]; D9 [法律]; R [医药、卫生];
学科分类号
0301 ; 10 ;
摘要
Comparing the infrared transmission spectrum of cocaine HCl to its attenuated total reflection (ATR) spectrum has raised questions about the use of ATR spectra for forensic drug analysis. Whenever infrared spectra are collected using different modes or sample preparation methods, small variations in peak intensity ratios or peak positions are possible. These variations in infrared spectra are small and do not interfere with qualitative analysis, but they can cause confusion when unrecognized as normal effects of the different spectroscopic techniques. Comparison of the absorption and ATR spectra of cocaine hydrochloride illustrates the type of differences that can be expected. These differences are ex plained by the fundamental differences in the collection techniques. For the best quantitative results, only spectra collected by the same technique should be compared.
引用
收藏
页码:822 / 829
页数:8