Permanent pacemaker use after cardiac transplantation: Completing the audit cycle

被引:7
|
作者
Holt, ND
Tynan, MM
Scott, CD
Parry, G
Dark, JH
McComb, JM
机构
[1] Regional Cardiothoracic Centre, Freeman Hospital, Newcastle upon Tyne
[2] Regional Cardiothoracic Centre, Freeman Hospital
关键词
audit; pacemakers; cardiac transplantation;
D O I
10.1136/hrt.76.5.435
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective-To determine the effects of delaying permanent pacemaker implantation in cardiac transplant recipients from less than three weeks to three weeks or more post transplantation-a change prompted by an earlier audit. Design-Retrospective review of resting 12 lead electrocardiograms and prospective 24 hour ambulatory electrocardiograms. Comparison of pacemaker usage before (period 1) and after (period 2) the policy change in November 1990. Setting-Outpatient department, supra-regional cardiopulmonary transplant unit. Patients-All 30 consecutive orthotopic cardiac transplant recipients who received a permanent pacemaker within one month of transplantation between May 1985 and August 1995. Main outcome measures-Presence of pacing on the 12 lead electrocardiogram and during 24 hour ambulatory electrocardiogram monitoring (pacemaker programmed to 50 beats per minute). Results-16/152 (10.5%) cardiac transplant recipients received permanent pacemakers in period 1 compared with 14/180 (7.8%) in period 2 (P = NS). Evidence of pacing was seen on 12 lead electrocardiograms at three months in 37.5% recipients in period 1 compared with 78.6% in period 2 (P = 0.03). At six months pacemaker usage had declined to 18.8% in period 1 and 35.7% in period 2 and at three years to 13.3% in period 1 and 40% in period 2 (P = NS for both). 21% patients in period 1 pa;ced on ambulatory 24 hour monitoring compared with 38.5% in period 2 (P = NS). Conclusions-Delaying permanent pacemaker implantation to three weeks or more after cardiac transplantation reduced the proportion of permanent pacemaker implantations, slightly but not significantly. There was a significant increase in permanent pacemaker usage at three months post transplantation with trends towards increased usage at later times, suggesting more appropriate selection of patients for permanent pacing.
引用
收藏
页码:435 / 438
页数:4
相关论文
共 50 条
  • [1] Permanent pacemaker implantation after cardiac transplantation: Extra cost of a conservative policy
    Holt, ND
    Parry, G
    Tynan, MM
    Dark, JH
    McComb, JM
    HEART, 1996, 76 (05) : 439 - 441
  • [2] Use of ultrasound for central line insertion: completing the audit cycle
    Dondapati, S.
    Bhagwat, A.
    Uppugonduri, S.
    ANAESTHESIA, 2019, 74 : 73 - 73
  • [3] Antibiotic use in lower limb trauma - completing the audit cycle
    Marsden, N.
    Yarrow, J.
    Rahman, S.
    Wilson-Jones, N.
    Pallister, I.
    Hemington-Gorse, S.
    BRITISH JOURNAL OF SURGERY, 2014, 101 : 70 - 70
  • [4] Predictors of Permanent Pacemaker Implant after Cardiac Transplantation: A Retrospective Analysis of a Contemporary Cohort
    Moser, Ashley R.
    St Clair, Katherine R.
    Farid, Talha A.
    House, John A.
    Wimmer, Alan P.
    Borkon, Michael
    Kao, Andrew
    Aggarwal, Sanjeev
    Magalski, Anthony
    Austin, Bethany A.
    JOURNAL OF CARDIAC FAILURE, 2014, 20 (08) : S31 - S32
  • [5] An audit of audits: are we completing the cycle?
    Gnanalingham, J
    Gnanalingham, MG
    Gnanalingham, KK
    JOURNAL OF THE ROYAL SOCIETY OF MEDICINE, 2001, 94 (06) : 288 - 289
  • [6] PERMANENT CARDIAC PACEMAKER
    TORRESANI, J
    JAUSSERAN, JM
    BERNARD, Y
    SCHWEIZERISCHE MEDIZINISCHE WOCHENSCHRIFT, 1972, 102 (12) : 387 - +
  • [7] Cardiac allograft vasculopathy and donor age affecting permanent pacemaker implantation after heart transplantation
    Roest, Stefan
    Manintveld, Olivier C.
    Kolff, Marit A. E.
    Akca, Ferdi
    Veenis, Jesse F.
    Constantinescu, Alina A.
    Brugts, Jasper J.
    Birim, Ozcan
    van Osch-Gevers, Lennie M.
    Szili-Torok, Tamas
    Caliskan, Kadir
    ESC HEART FAILURE, 2022, 9 (02): : 1239 - 1247
  • [8] Cardiac pacemaker implantation after heart transplantation
    Castagno, Davide
    Boffini, Massimo
    Troiano, Francesca
    Barbero, Cristina
    Simonato, Erika
    Pidello, Stefano
    Ponzo, Federica
    Budano, Carlo
    Golzio, Piergiorgio
    Anselmino, Matteo
    Ferraris, Enrico
    Bissolino, Arianna
    Giustetto, Carla
    De Ferrari, Gaetano Maria
    Rinaldi, Mauro
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2019, 21 (0J) : J143 - J144
  • [9] PERMANENT PACING AFTER CARDIAC TRANSPLANTATION
    SCOTT, CD
    MCCOMB, JM
    DARK, JH
    BEXTON, RS
    BRITISH HEART JOURNAL, 1993, 69 (05): : 399 - 403
  • [10] PERMANENT PACING AFTER CARDIAC TRANSPLANTATION
    HEINZ, G