Optimization of dose and image quality in pediatrics chest digital radiography

被引:0
|
作者
Deevband, Mohammad Reza [1 ]
Rahmani, Fereshteh [1 ]
Bardar, Meisam Ebrahimi [2 ]
Ghorbani, Mahdi [1 ]
机构
[1] Shahid Beheshti Univ Med Sci, Sch Med, Biomed Engn & Med Phys Dept, Tehran, Iran
[2] Mashhad Univ Med Sci, Fac Med, Dept Med Phys, Mashhad, Razavi Khorasan, Iran
关键词
Digital radiography; Pediatric; Patient dose; Image quality; Dose optimization; X-RAY EXAMINATIONS;
D O I
10.1016/j.jmir.2021.05.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: In this study, pediatric chest digital radiography (DR) is evaluated in response to the high volume of chest DR examinations and high radiosensitivity of children and young adults. The aim of the study is to optimize irradiation parameters in chest DR to have dose as low as reasonably achievable (ALARA) and simultaneously obtain improved or preserved image quality. Materials and methods: Homogeneous phantoms in terms of density, dimensions, and composition were constructed to produce equivalent chest phantoms with less than 5% error for the 5-10 and 1015-year-old age groups. The modulation transfer function (MTF), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured for both the reference and new conditions for the two age groups. Results: The results indicate that for the 5-10 year old age group, the optimized technique was 75 kVp and 6.3 mAs, in which the dose reduced 10% and the SNR and CNR increased by 0.4% and 6%, respectively, compared to the reference condition. For the 10-15-yearold age group, the 85 kVp and 5 mAs was close to the optimum condition, in which the dose reduced 37% and the SNR and CNR increased by 16% and 4%, respectively, compared to the reference condition. Conclusion: The introduced optimized conditions in this study are accompanied by lower dose and higher SNR and CNR; therefore, they Pour le groupe d'age 10-15 ans, 85 kVp et 5 mAs etait proche de la condition optimale, dans laquelle la dose a ete reduite de 37 % tandis que le SNR et CNR affichaient une augmentation respective de 16 % et 4 % par rapport a la condition de reference.
引用
收藏
页码:443 / 449
页数:7
相关论文
共 50 条
  • [1] Evaluation of dose-image-quality optimization in digital chest radiography
    Lee, S. C.
    Wang, J. N.
    Liu, S. C.
    Jiang, S. H.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 544 - 547
  • [2] Optimization of Dose and Image Quality for Computed Radiography and Digital Radiography
    John E. Aldrich
    Emerenciana Duran
    Pat Dunlop
    John R. Mayo
    Journal of Digital Imaging, 2006, 19 : 126 - 131
  • [3] Optimization of dose and image quality for computed radiography and digital radiography
    Aldrich, JE
    Duran, E
    Dunlop, P
    Mayo, JR
    JOURNAL OF DIGITAL IMAGING, 2006, 19 (02) : 126 - 131
  • [4] Dose reduction in digital chest radiography and perceived image quality
    Kroft, L. J. M.
    Veldkamp, W. J. H.
    Mertens, B. J. A.
    Van Delft, J-P A.
    Geleijns, J.
    BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (960): : 984 - 988
  • [5] Dose-image quality optimisation in digital chest radiography
    Doyle, P
    Martin, CJ
    Gentle, D
    RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 269 - 272
  • [6] Analyses of patient dose and image quality for chest digital radiography
    Tsai, H. Y.
    Yang, C. H.
    Huang, K. M.
    Li, M. J.
    Tung, C. J.
    RADIATION MEASUREMENTS, 2010, 45 (3-6) : 722 - 725
  • [7] Digital chest radiography image quality assessment with dose reduction
    Grewal, R. K.
    Young, N.
    Collins, L.
    Karunaratne, N.
    Sabharwal, R.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) : 71 - 80
  • [8] Digital chest radiography image quality assessment with dose reduction
    R. K. Grewal
    N. Young
    L. Collins
    N. Karunaratne
    R. Sabharwal
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 71 - 80
  • [9] Quality and dose optimization in canine chest radiography using a digital radiography system
    Santana de Souza, Sergio Augusto
    Fattori Alves, Allan Felipe
    Mamprim, Maria Jaqueline
    Pina, Diana Rodrigues
    RADIATION PHYSICS AND CHEMISTRY, 2022, 195
  • [10] A phantom for dose-image quality optimization in chest radiography
    Vassileva, J
    BRITISH JOURNAL OF RADIOLOGY, 2002, 75 (898): : 837 - 842