Assessing a single targeted next generation sequencing for human leukocyte antigen typing protocol for interoperability, as performed by users with variable experience

被引:10
|
作者
Gandhi, Manish J. [1 ]
Ferriola, Deborah [2 ]
Lind, Curt [2 ]
Duke, Jamie L. [2 ]
Huynh, Anh [2 ]
Papazoglou, Anna [2 ]
Mackiewicz, Kate [2 ]
Christiansen, Mette [3 ]
Dong, Wei [4 ]
Hsu, Susan [4 ]
Thomas, Dawn [5 ]
Schneider, Brittany [1 ]
Pierce, Erin [6 ]
Kearns, Jane [6 ]
Kamoun, Malek [6 ]
Monos, Dimitri [2 ,6 ]
Askar, Medhat [5 ,7 ]
机构
[1] Mayo Clin, Rochester, MN USA
[2] Childrens Hosp Philadelphia, Immunogenet Lab, Philadelphia, PA 19104 USA
[3] Aarhus Univ Hosp, Dept Clin Immunol, Aarhus, Denmark
[4] Amer Red Cross, Penn Jersey Reg, Philadelphia, PA USA
[5] Cleveland Clin, Cleveland, OH 44106 USA
[6] Univ Penn, Perelman Sch Med, Dept Pathol & Lab Med, Philadelphia, PA 19104 USA
[7] Baylor Univ, Med Ctr, Waco, TX 76798 USA
关键词
Next generation sequencing; Human leukocyte antigen typing; High resolution typing; HLA; NGS; HIGH-RESOLUTION; HLA; NGS;
D O I
10.1016/j.humimm.2017.07.012
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Background: A simplified protocol for HLA-typing-by NGS, developed for use with the Illumina MiSeq, was performed by technologists with varying NGS experience to assess accuracy and reproducibility. Methods: Technologists from six laboratories typed the same 16 samples at HLA-A, B, C, DRB1, and DQB1. The protocol includes long range PCR, library preparation, and paired-end 250 bp sequencing. Two indexing strategies were employed: locus-specific indexing whereby each locus was tagged uniquely and sample-specific indexing whereby all 5 loci for a sample were pooled prior to library preparation. Sequence analysis was performed with two software packages, Target HLA (Omixon) and NGSengine (GenDx). Results: The average number of sequence reads per library was 387,813; however, analysis was limited to 40,000 reads for locus-indexed libraries and 200,000 reads for sample-indexed libraries resulting in an average depth of coverage of 1444 reads per locus. Sufficient reads for genotype analysis were obtained for 98.4% of libraries. Genotype accuracy was >97% in pooled amplicons and >99% in individual amplicons by both software analysis. Inter-laboratory reproducibility was 99.7% and only cause of discordance was cross-contamination of a single amplicon. Conclusions: This NGS HLA-typing protocol is simple, reproducible, scalable, highly accurate and amenable to clinical testing. (C) 2017 American Society for Histocompatibility and Immunogenetics. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:642 / 648
页数:7
相关论文
共 50 条
  • [1] Human Leukocyte Antigen Typing by Next-Generation Sequencing
    Profaizer, Tracie
    Kumanovics, Attila
    CLINICS IN LABORATORY MEDICINE, 2018, 38 (04) : 565 - +
  • [2] Introduction of Human Leukocyte Antigen Typing by Next Generation Sequencing in Pakistan
    Tipu, Hamid Nawaz
    Noman, Muhammad
    JCPSP-JOURNAL OF THE COLLEGE OF PHYSICIANS AND SURGEONS PAKISTAN, 2020, 30 (09): : 994 - 994
  • [3] Assessment of human leukocyte antigen typing and chimerism testing combined on a single next generation sequencing run
    Willis, A.
    Lind, C.
    Zimmerman, A.
    Hsiao, T.
    Kmiec, B.
    HUMAN IMMUNOLOGY, 2023, 84 : 193 - 193
  • [4] ACCURATE TYPING OF HUMAN LEUKOCYTE ANTIGEN (HLA) BY NEXT-GENERATION EXOME SEQUENCING.
    Liu, Chang
    Yang, Xiao
    Duffy, Brian
    Mohanakumar, T.
    Mitra, Robi D.
    Zody, Michael C.
    Pfeifer, John D.
    HUMAN IMMUNOLOGY, 2013, 74 : 11 - 11
  • [5] Targeted Next- Generation Sequencing for Human Leukocyte Antigen Typing in a Clinical Laboratory Metrics of Relevance and Considerations for Its Successful Implementation
    Gandhi, Manish J.
    Ferriola, Deborah
    Huang, Yanping
    Duke, Jamie L.
    Monos, Dimitri
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2017, 141 (06) : 806 - 812
  • [6] Comparison of Sequence Specific Primers in the Next Generation Sequencing in Human Leukocyte Antigen Typing for Transplant Recipients
    Tipu, Hamid Nawaz
    JCPSP-JOURNAL OF THE COLLEGE OF PHYSICIANS AND SURGEONS PAKISTAN, 2020, 30 (11): : 1138 - 1142
  • [7] Next-generation sequencing: the solution for high-resolution, unambiguous human leukocyte antigen typing
    Lind, C.
    Ferriola, D.
    Mackiewicz, K.
    Heron, S.
    Rogers, M.
    Slavich, L.
    Walker, R.
    Hsiao, T.
    McLaughlin, L.
    D'Arcy, M.
    Gai, X.
    Goodridge, D.
    Sayer, D.
    Monos, D.
    HUMAN IMMUNOLOGY, 2010, 71 (10) : 1033 - 1042
  • [8] Benchmarking the Human Leukocyte Antigen Typing Performance of Three Assays and Seven Next-Generation Sequencing-Based Algorithms
    Liu, Ping
    Yao, Minya
    Gong, Yu
    Song, Yunjie
    Chen, Yanan
    Ye, Yizhou
    Liu, Xiao
    Li, Fugen
    Dong, Hua
    Meng, Rui
    Chen, Hao
    Zheng, Aiwen
    FRONTIERS IN IMMUNOLOGY, 2021, 12
  • [9] Performance Characteristics and Validation of Next-Generation Sequencing for Human Leucocyte Antigen Typing
    Weimer, Eric T.
    Montgomery, Maureen
    Petraroia, Rosanne
    Crawford, John
    Schmitz, John L.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2016, 18 (05): : 668 - 675
  • [10] A national proficiency scheme for human leucocyte antigen typing by next-generation sequencing
    Lin, Guigao
    Zhang, Kuo
    Li, Jinming
    CLINICA CHIMICA ACTA, 2022, 533 : 85 - 88