In vitro comparison of the hemocompatibility of two centrifugal left ventricular assist devices

被引:15
|
作者
Zayat, Rashad [1 ]
Moza, Ajay [1 ]
Grottke, Oliver [2 ]
Grzanna, Tim [1 ]
Fechter, Tamara [2 ]
Motomura, Tadashi [4 ]
Schmidt-Mewes, Christian [4 ]
Breuer, Thomas [3 ]
Autschbach, Udiger [1 ]
Rossaint, Rolf [2 ]
Goetzenich, Andreas [1 ]
Bleilevens, Christian [2 ]
机构
[1] RWTH Univ Hosp, Fac Med, Dept Thorac & Cardiovasc Surg, Aachen, Germany
[2] RWTH Univ Hosp, Fac Med, Dept Anesthesiol, Aachen, Germany
[3] RWTH Univ Hosp, Fac Med, Dept Intens & Intermediate Care, Aachen, Germany
[4] Evaheart Inc, Houston, TX USA
来源
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY | 2019年 / 157卷 / 02期
关键词
Left ventricular assist device; circulatory mock-loop; HeartMate; 3; EVAHEART; haemolysis; hemocompatibility; VON-WILLEBRAND-FACTOR; CONTINUOUS-FLOW; HEARTMATE II; P-SELECTIN; ACTIVATION; EVAHEART; BLOOD; COAGULATION; MULTICENTER; PUMP;
D O I
10.1016/j.jtcvs.2018.07.085
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: Shear stress from left ventricular assist devices induces von Willebrand factor degradation and platelet dysfunction, leading to nonsurgical bleeding. We characterized the hemostatic changes induced by 2 centrifugal left ventricular assist devices, the HeartMate 3 (Abbott Inc, Chicago, Ill) and the EVAHEART (Evaheart Inc, Houston, Tex), for comparison. Methods: Whole blood from 8 healthy volunteers was used ex vivo. Blood from the same donor was used for 6 hours of circulation in a miniature mock-loop system consisting of 2 identical extracorporeal circuits to compare the following experimental settings: (1) optimal revolutions per minute (rpm) for the HeartMate 3 (n = 4; 5000 rpm) and the EVAHEART (n = 4; 2500 rpm) and (2) equal rpm (3000 rpm for the HeartMate 3 and EVAHEART, n = 4 vs n = 4). For both settings, blood flow was adjusted to 1 mock-loop filling volume per minute (HeartMate 3 = 82 mL/min, EVAHEART = 100 mL/min). A panel of coagulation markers was analyzed to investigate hemostatic changes. Results: The free plasma hemoglobin concentration was significantly lower in the EVAHEART compared with the HeartMate 3 after 6 hours of mock-loop circulation under both settings (optimal: 37 +/- 31 vs 503 +/- 173 mg/dL, P <. 0001; equal: 27 +/- 4 vs 139 +/- 135 mg/dL, P = .024). Loss of von Willebrand factor high-molecular-weight multimers occurred in both left ventricular assist devices and settings, but the von Willebrand factor: activity/von Willebrand factor: antigen ratio after 6 hours was significantly lower in optimal settings for the HeartMate 3 (P = .009). The thrombin-antithrombin complex level was significantly lower with the EVAHEART for both settings (P <. 0001). Conclusions: The EVAHEART left ventricular assist device caused less hemolysis, resulted in lower coagulation activation, and provided better preservation of von Willebrand factor functional activity compared with the HeartMate 3 device. These findings prove that left ventricular assist device design plays a major role in minimizing blood damage during left ventricular assist device support.
引用
收藏
页码:591 / +
页数:13
相关论文
共 50 条
  • [1] Comparison of the Hemocompatibility of an Axial and a Centrifugal Left Ventricular Assist Device in an In Vitro Test Circuit
    Borchers, Patrick
    Winnersbach, Patrick
    Kraemer, Sandra
    Beckers, Christian
    Buhl, Eva Miriam
    Leonhardt, Steffen
    Rossaint, Rolf
    Walter, Marian
    Breuer, Thomas
    Bleilevens, Christian
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (12)
  • [2] A tale of two centrifugal left ventricular assist devices
    Schroder, Jacob N.
    Milano, Carmelo A.
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2017, 154 (03): : 850 - 852
  • [3] The CentriMag Centrifugal Blood Pump as a Benchmark for In Vitro Testing of Hemocompatibility in Implantable Ventricular Assist Devices
    Chan, Chris H. H.
    Pieper, Ina Laura
    Hambly, Rebecca
    Radley, Gemma
    Jones, Alyssa
    Friedmann, Yasmin
    Hawkins, Karl M.
    Westaby, Stephen
    Foster, Graham
    Thornton, Catherine A.
    ARTIFICIAL ORGANS, 2015, 39 (02) : 93 - 101
  • [4] Retrospective Hemolysis Comparison Between Patients With Centrifugal Biventricular Assist and Left Ventricular Assist Devices
    Stepanenko, Alexander
    Krabatsch, Thomas
    Hennig, Ewald
    Kaufmann, Friedrich
    Jurmann, Beate
    Dranishnikov, Nikolay
    Lehmkuhl, Hans B.
    Pasic, Miralem
    Weng, Yuguo
    Hetzer, Roland
    Potapov, Evgenij V.
    ASAIO JOURNAL, 2011, 57 (05) : 382 - 387
  • [5] In Vitro Hemocompatibility Evaluation of Modified Rotary Left to Right Ventricular Assist Devices in Pulmonary Flow Conditions
    Chan, Chris Hoi Houng
    Nandakumar, Deepika
    Balletti, Nicoletta
    Horobin, Jarod
    Wu, Eric Lee-Jan
    Bouquet, Mahe
    Stephens, Andrew
    Pauls, Jo P.
    Tansley, Geoff
    Fraser, John F.
    Simmonds, Michael J.
    Gregory, Shaun D.
    ASAIO JOURNAL, 2020, 66 (06) : 637 - 644
  • [6] Hemocompatibility and biophysical interface of left ventricular assist devices and total artificial hearts
    Nascimbene, Angelo
    Bark, David
    Smadja, David M.
    BLOOD, 2024, 143 (08) : 661 - 672
  • [7] Evaluation of the Hemocompatibility of the Direct Oral Anticoagulant Apixaban in Left Ventricular Assist Devices
    Shah, Palak
    Looby, Mary
    Dimond, Matthew
    Bagchi, Pramita
    Shah, Bhruga
    Isseh, Iyad
    Rollins, Allman T.
    Abdul-Aziz, Ahmad A.
    Kennedy, Jamie
    Tang, Daniel G.
    Klein, Katherine M.
    Casselman, Samantha
    Vermeulen, Christen
    Sheaffer, Wendy
    Snipes, Meredith
    Sinha, Shashank S.
    O'Connor, Christopher M.
    JACC-HEART FAILURE, 2024, 12 (09) : 1540 - 1549
  • [8] Comparison of Exercise Hemodynamics in Patients Supported with Centrifugal and Axial Flow Left Ventricular Assist Devices
    Moss, N.
    Parikh, A.
    Rakita, V.
    Lala, A.
    Mitter, S.
    Roldan, J.
    Cuca, A.
    Barghash, M.
    Contreras, J.
    Pinney, S.
    Anyanwu, A.
    Mancini, D. M.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2019, 38 (04): : S152 - S152
  • [9] Characteristics of strokes associated with centrifugal flow left ventricular assist devices
    Inamullah, Ovais
    Chiang, Yuting P.
    Bishawi, Muath
    Weiss, Martin
    Lutz, Michael W.
    Blue, Laura J.
    Feng, Wayne
    Milano, Carmelo A.
    Luedke, Matthew
    El Husseini, Nada
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [10] Heat Generation in Axial and Centrifugal Flow Left Ventricular Assist Devices
    Yost, Gardner
    Joseph, Christine Rachel
    Royston, Thomas
    Tatooles, Antone
    Bhat, Geetha
    ASAIO JOURNAL, 2016, 62 (06) : 684 - 689