Insensitivity of the routine dimethylglyoxime test for detecting release of nickel from earrings

被引:19
|
作者
Pönkä, A [1 ]
Ekman, A [1 ]
机构
[1] Helsinki City Ctr Environm, Helsinki 00530, Finland
关键词
dimethylglyoxime test; nickel; piercing earrings; atomic absorption spectrometry; artificial sweat;
D O I
10.1016/S0048-9697(98)00345-3
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The Helsinki City Centre of the Environment tested two methods, dimethylglyoxime (DMG) and atomic absorption spectrometry (AAS), for detecting nickel release in piercing earrings. The DMG screening test was performed in two slightly different ways, with and without ethanol and heat prehandling. All 30 pairs of earrings tested, totalling 66 objects, were negative. However, according to our AAS test, 25 of the 66 objects (38%) released greater than or equal to 0.05% of nickel, the mean amount being 2.1% and the maximum 12%. When measured by AAS after artificial sweat treatment, 11 objects released more than 0.5 mu g/cm(2) per week of nickel, the mean amount being 3.4 mu g/cm(2) and the range < 0.1-84 mu g/cm(2). After this sweat treatment, nine of the objects (14%) were positive in DMG tests. These findings indicate that the DMG test is unreliable for detecting nickel release from jewelry. Quality control of consumer items should be performed by laboratories that have quantitative analysis methods for such investigations. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:161 / 165
页数:5
相关论文
共 50 条
  • [1] Assessment of nickel release from earrings randomly purchased in China and Thailand using the dimethylglyoxime test
    Hamann, Carsten R.
    Hamann, Dathan J.
    Hamann, Quinlan J.
    Hamann, Curtis P.
    Boonchai, Waranya
    Li, Lin-Feng
    Thyssen, Jacob P.
    CONTACT DERMATITIS, 2010, 62 (04) : 232 - 240
  • [2] Nickel Release from Laptop Detected by Dimethylglyoxime (DMG) Test
    Amala, Deepika
    Nayak, Sudhir U. K.
    Shenoi, Shrutakirthi D.
    INDIAN JOURNAL OF DERMATOLOGY, 2021, 66 (06) : 696 - 697
  • [3] Nickel release from metallic earrings: A survey of the Danish market and validation of the nickel spot test
    Wennervaldt, Michael
    Ahlstrom, Malin G.
    Menne, Torkil
    Thyssen, Jacob P.
    Johansen, Jeanne D.
    CONTACT DERMATITIS, 2021, 85 (02) : 178 - 185
  • [4] RELEASE OF NICKEL BY PERMANENT WAVE LIQUIDS, SHOWN BY THE DIMETHYLGLYOXIME TEST
    MENEZESBRANDAO, F
    CONTACT DERMATITIS, 1979, 5 (06) : 406 - 406
  • [5] NICKEL RELEASE FROM EAR PIERCING KITS AND EARRINGS
    FISCHER, T
    FREGERT, S
    GRUVBERGER, B
    RYSTEDT, I
    CONTACT DERMATITIS, 1984, 10 (01) : 39 - 41
  • [6] DIMETHYLGLYOXIME STICK TEST FOR EASIER DETECTION OF NICKEL
    SHORE, RN
    BINNICK, S
    ARCHIVES OF DERMATOLOGY, 1977, 113 (12) : 1734 - 1734
  • [7] Sensitivity and specificity of the nickel spot (dimethylglyoxime) test
    Thyssen, Jacob P.
    Skare, Lizbet
    Lundgren, Lennart
    Menne, Torkil
    Johansen, Jeanne D.
    Maibach, Howard I.
    Liden, Carola
    CONTACT DERMATITIS, 2010, 62 (05) : 279 - 288
  • [8] DETECTION OF NICKEL BY DIMETHYLGLYOXIME SPOT-TEST
    FISHER, AA
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1975, 234 (04): : 429 - 430
  • [9] THE DIMETHYLGLYOXIME TEST IN THE PREVENTION AND MANAGEMENT OF NICKEL DERMATITIS
    FISHER, AA
    CUTIS, 1990, 46 (06): : 467 - 468
  • [10] Nickel release from earrings purchased in the united states: the San Francisco earring study
    Thyssen, Jacob Pontoppidan
    Maibach, Howard Ira
    JOURNAL OF THE AMERICAN ACADEMY OF DERMATOLOGY, 2008, 58 (06) : 1000 - 1005