A simulation study of irregular respiratory motion and its dosimetric impact on lung tumors

被引:27
|
作者
Mutaf, Y. D. [1 ]
Scicutella, C. J. [1 ]
Michalski, D. [1 ]
Fallon, K. [1 ]
Brandner, E. D. [1 ]
Bednarz, G. [1 ]
Huq, M. S. [1 ]
机构
[1] Univ Pittsburgh, Dept Radiat Oncol, Med Ctr, Pittsburgh, PA 15232 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 03期
关键词
RADIOTHERAPY; DISTRIBUTIONS; PNEUMONITIS; POSITION; CANCER; LIVER;
D O I
10.1088/0031-9155/56/3/019
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study is aimed at providing a dosimetric evaluation of the irregular motion of lung tumors due to variations in patients' respiration. Twenty-three lung cancer patients are retrospectively enrolled in this study. The motion of the patient clinical target volume is simulated and two types of irregularities are defined: characteristic and uncharacteristic motions. Characteristic irregularities are representative of random fluctuations in the observed target motion. Uncharacteristic irregular motion is classified as systematic errors in determination of the target motion during the planning session. Respiratory traces from measurement of patient abdominal motion are also used for the target motion simulations. Characteristic irregular motion was observed to cause minimal changes in target dosimetry with the largest effect of 2.5% +/- 0.9% (1 sigma) reduction in the minimum target dose (D-min) observed for targets that move 2 cm on average and exhibiting 50% amplitude variations within a session. However, uncharacteristic irregular motion introduced more drastic changes in the clinical target volume (CTV) dose; 4.1% +/- 1.7% reduction for 1 cm motion and 9.6% +/- 1.7% drop for 2 cm. In simulations with patients' abdominal motion, corresponding changes in target dosimetry were observed to be negligible (<0.1%). Only uncharacteristic irregular motion was identified as a clinically significant source of dosimetric uncertainty.
引用
收藏
页码:845 / 859
页数:15
相关论文
共 50 条
  • [1] Dosimetric Effect of Irregular Respiratory Motion on Radiotherapy of Lung Tumors
    Mutaf, Y. D.
    Scicutella, C. J.
    Michalski, D.
    Brandner, E. D.
    Fallon, K.
    Bednarz, G.
    Huq, S. M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S607 - S607
  • [2] An Advanced Deformable Lung Phantom for Analyzing the Dosimetric Impact of Respiratory Motion
    Shin, D.
    Kang, S.
    Kim, D.
    Kim, T.
    Kim, K.
    Cho, M.
    Suh, T.
    MEDICAL PHYSICS, 2016, 43 (06) : 3638 - 3638
  • [3] Is Respiratory Gating More Prone to Dosimetric Errors Due to Irregular Respiratory Motion?
    Mutaf, Y.
    Scicutella, C.
    Michalski, D.
    Brandner, E.
    Fallon, K.
    Bednarz, G.
    Huq, M.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [4] Dosimetric effects of respiratory motion during stereotactic body radiation therapy of lung tumors
    Sarudis, Sebastian
    Karlsson, Anna
    Nyman, Jan
    Back, Anna
    ACTA ONCOLOGICA, 2022, 61 (08) : 1004 - 1011
  • [5] Quantitative Analysis of Dosimetric Effects Due to Irregular Respiratory Motion in Stereotactic Body Radiotherapy of Lung Cancer
    Mutaf, Y.
    Michalski, D.
    Yang, Y.
    Li, X.
    Li, T.
    Bednarz, G.
    Heron, D.
    Huq, M.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [6] Dosimetric impact of respiratory motion during boron neutron capture therapy for lung cancer
    Wu, Shaojuan
    Geng, Changran
    Tang, Xiaobin
    Bortolussi, Silva
    Han, Yang
    Shu, Diyun
    Gong, Chunhui
    Zhang, Xinxin
    Tian, Feng
    RADIATION PHYSICS AND CHEMISTRY, 2020, 168
  • [7] Dosimetric impact of respiratory motion during boron neutron capture therapy for lung cancer
    Wu, Shaojuan
    Geng, Changran
    Tang, Xiaobin
    Bortolussi, Silva
    Han, Yang
    Shu, Diyun
    Gong, Chunhui
    Zhang, Xinxin
    Tian, Feng
    Radiation Physics and Chemistry, 2020, 168
  • [8] Dosimetric study of the different techniques to deal with respiratory motion for lung stereotactic radiotherapy
    Paumier, A.
    Crespeau, A.
    Krhili, S.
    Georgin-Mege, M.
    Tuchais, C.
    Mesgouez, J.
    Cellier, P.
    Lisbona, A.
    Denis, F.
    Autret, D.
    CANCER RADIOTHERAPIE, 2012, 16 (04): : 263 - 271
  • [9] Dosimetric study of the impact of breathing using a dynamic prototype simulating the respiratory motion
    Masset, H.
    Gavignet, E.
    Gschwind, R.
    Dumas, J-L.
    Makovicka, L.
    Bosset, J. F.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S507 - S508
  • [10] Evaluation of the Dosimetric Impact of Respiratory-Induced Motion
    St-Pierre, C.
    Tremblay, D.
    Lavoie, C.
    Archambault, L.
    MEDICAL PHYSICS, 2011, 38 (06) : 3467 - +