Simple liquid chromatography method for the rapid simultaneous determination of prednisolone and cortisol in plasma and urine using hydrophilic lipophilic balanced solid phase extraction cartridges

被引:68
|
作者
AbuRuz, S
Millership, J
Heaney, L
McElnay, J
机构
[1] Queens Univ Belfast, Sch Pharm, Clin & Practice Res Grp, Belfast BT9 7BL, Antrim, North Ireland
[2] Belfast City Hosp, Reg Resp Ctr, Belfast BT9 7AD, Antrim, North Ireland
关键词
steroids; prednisolone; cortisol;
D O I
10.1016/j.jchromb.2003.09.044
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This article describes the development and validation of a simple solid phase extraction (SPE) and HPLC method for the extraction and the specific determination of prednisolone and hydrocortisone (cortisol) in both plasma and urine using one washing step with Oasis' hydrophilic lipophilic balanced (HLB) cartridges (1 ml/30 mg, 30 mum). Recoveries of prednisolone and cortisol from plasma and urine exceeded 82%. The limit of quantification (LOQ) in plasma and urine was 9.9 and 6.7 ng/ml for cortisol, respectively, and 11.6 and 8.0 ng/ml for prednisolone, respectively. The intraday and interday precision (measured by CV%) for both prednisolone and cortisol in both plasma and urine was always less than 7%. The accuracy (measured by relative error %) for both prednisolone and cortisol in both plasma and urine was always less than 8%. The advantages of the developed method are the use of a one step washing SPE utilising HLB cartridges which do not suffer the drying out problems of conventional SPE cartridges and the time saving when compared with solvent extraction (SE), in addition to the simultaneous determination of prednisolone and cortisol in both plasma and urine. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:193 / 201
页数:9
相关论文
共 50 条