Effect of microseparation on contact mechanics in metal-on-metal hip replacementsA finite element analysis

被引:27
|
作者
Liu, Feng [1 ]
Williams, Sophie [1 ]
Fisher, John [1 ]
机构
[1] Univ Leeds, Inst Med & Biol Engn, Sch Mech Engn, Leeds, W Yorkshire, England
基金
英国工程与自然科学研究理事会; 英国生物技术与生命科学研究理事会; 英国惠康基金;
关键词
metal-on-metal; hip replacement; microseparation; rim contact; finite element; IN-VIVO DETERMINATION; FEMORAL-HEAD SIZE; RESURFACING ARTHROPLASTY; LOADING CONDITIONS; WEAR; BEARINGS; CUP; FAILURE; AGE; LUBRICATION;
D O I
10.1002/jbm.b.33313
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Some early failures of metal-on-metal (MoM) hip replacements associated with elevated wear have caused concerns for the use of this bearing combination. Simulator studies have shown that microseparation and its associated rim contact and edge loading may produce the most severe wear in MoM bearings. It is generally recognized that this high wear can be attributed to the high contact stress of the head on the rim of the cup. In this study, an improved finite element contact model that incorporates an elastic-perfectly plastic material property for cobalt-chrome alloy of the metal bearing was developed in an attempt to provide an accurate prediction of the stress and strain for the rim contact. The effects of the microseparation displacement (0.1-2 mm), cup inclination angle (25-65 degrees) and cup rim radius (0.5-4 mm) on the contact stress/strain were investigated. The results show that a translational displacement >0.1 mm under a load >0.5 kN can produce a highly concentrated contact stress at the surface of the cup rim which can lead to plastic deformation. This study also suggests that the magnitude of translational displacement was the major factor that determined the severity of the contact conditions and level of stress and strain under microseparation conditions. Future studies will address the effect of surgical translational and rotational malposition and component design on the magnitude of microseparation, contact stress and strain and severity of wear. (c) 2014 The Authors. Journal of Biomedical Materials Research Part B: Applied Biomaterials Published by Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 103B: 1312-1319, 2015.
引用
收藏
页码:1312 / 1319
页数:8
相关论文
共 50 条
  • [1] Contact mechanics analysis of metal-on-metal hip resurfacing prostheses
    Udofia, IJ
    Yew, A
    Jin, ZM
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2004, 218 (H5) : 293 - 305
  • [2] Analysis of contact mechanics in McKee-Farrar metal-on-metal hip implants
    Yew, A
    Jagatia, M
    Ensaff, H
    Jin, ZM
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2003, 217 (H5) : 333 - 340
  • [3] Contact mechanics of a novel metal-on-metal total hip replacement
    Besong, AA
    Lee, R
    Farrar, R
    Jin, ZM
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2001, 215 (H6) : 543 - 548
  • [4] A finite element simulation of MOM (metal-on-metal) hip implant
    Chaturvedi, Sumita
    Bharti, Prem Kumar
    Yadav, Saurabh Kumar
    Singh, Saumya
    LUBRICATION SCIENCE, 2019, 31 (05) : 210 - 217
  • [5] Microseparation, fluid pressure and flow in failures of metal-on-metal hip resurfacing arthroplasties
    Wroblewski, B. M.
    Siney, P. D.
    Fleming, P. A.
    BONE & JOINT RESEARCH, 2012, 1 (03): : 25 - 30
  • [6] Comparison of contact mechanics between a total hip replacement and a hip resurfacing with a metal-on-metal articulation
    Liu, F
    Udofia, IJ
    Jin, ZM
    Hirt, F
    Rieker, C
    Roberts, P
    Grigoris, P
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART C-JOURNAL OF MECHANICAL ENGINEERING SCIENCE, 2005, 219 (07) : 727 - 732
  • [7] The effect of cup orientation and coverage on contact mechanics and range of motion of metal-on-metal hip resurfacing arthroplasty
    Wang, Ling
    Williams, Sophie
    Udofia, Itoro
    Isaac, Graham
    Fisher, John
    Jin, Zhongmin
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2012, 226 (H11) : 877 - 886
  • [8] Prediction of contact mechanics in metal-on-metal Total Hip Replacement for parametrically comprehensive designs and loads
    Donaldson, Finn E.
    Nyman, Edward, Jr.
    Coburn, James C.
    JOURNAL OF BIOMECHANICS, 2015, 48 (10) : 1828 - 1835
  • [9] A finite element method comparison of wear in two metal-on-metal total hip prostheses
    Cosmi, F.
    Hoglievina, M.
    Fancellu, G.
    Martinelli, B.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H8) : 871 - 879
  • [10] Contact mechanics and elastohydrodynamic lubrication in a novel metal-on-metal hip implant with an aspherical bearing surface
    Meng, Qingen
    Gao, Leiming
    Liu, Feng
    Yang, Peiran
    Fisher, John
    Jin, Zhongmin
    JOURNAL OF BIOMECHANICS, 2010, 43 (05) : 849 - 857