Influence of corneal thickness on comparative intraocular pressure measurements with goldmann and non-contact tonometers in keratoconus

被引:0
|
作者
Silih, MS [1 ]
Hawlina, M [1 ]
机构
[1] Univ Eye Clin, Med Ctr, Ljubljana 1000, Slovenia
关键词
Goldmann tonometry; non-contact tonometry; corneal thickness; keratoconus;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: The influence of corneal thickness and curvature on the difference between intraocular pressure (IOP) measurements obtained with non-contact (NCT) and those with the Goldmann applanation tonometer (GAT) was studied in patients with keratoconus. Materials and methods: We examined 113 patients with keratoconus. IOP was measured by the Canon TX 10 non-contact tonometer and the Goldmann tonometer, corneal curvature and thickness were obtained by Humphrey Automatic Refractometer Keratometer and Orbscan Version 3.0 Bausch & Lomb Surgical. Results: The IOP measured by NCT was significantly lower than that measured by GAT. The mean pachymetry of the thinnest point was 423.15 +/- 98.43 mum for the right eyes and 426.7 +/- 93.88 mum for the left eyes. The difference between NCT-GAT and corneal thickness showed a significant negative correlation (r = -0.427, p < 0.0001; t = -3.677, p < 0.0001). Values of NCT measurements were significantly increasing with corneal thickness (F = 6.505, p < 0.0001 for right eyes and F = 4.37, p = 0.004 for left eyes), whilst GAT measurements did not show a significant influence of the corneal thickness. The keratometry had no effect on the difference between NCT-GAT measurements (t = 1.090, p = 0.278). Conclusions: The thin cornea has more influence on the measurement with NCT than GAT. The relative precision of NCT compared with GAT seems to be influenced by the corneal thickness.
引用
收藏
页码:843 / 847
页数:5
相关论文
共 50 条