Medical beam monitor -: Pre-clinical evaluation and future applications

被引:5
|
作者
Frais-Koelbl, Helmut [1 ]
Griesmayer, Erich
Schreiner, Thomas
Georg, Dietmar
Pernegger, Heinz
机构
[1] Univ Appl Sci, Wiener Neustadt, Austria
[2] Fotec Forsch & Technol Transfer GmbH, Wiener Neustadt, Austria
[3] Med Univ, Dept Radiotherapy Radiobiol, Vienna, Austria
[4] CERN, Geneva, Switzerland
关键词
CVD diamond; semiconductor radiation detectors; radiotherapy dosimetry; ion therapy;
D O I
10.1016/j.nima.2007.07.038
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Future medical ion beam applications for cancer therapy which are based on scanning technology will require advanced beam diagnostics equipment. For a precise analysis of beam parameters we want to resolve time structures in the range of microseconds to nanoseconds. A prototype of an advanced beam monitor was developed by the University of Applied Sciences Wiener Neustadt and its research subsidiary Fotec in co-operation with CERN RD42, Ohio State University and the Jozef Stefan Institute in Ljubljana. The detector is based on polycrystalline Chemical Vapor Deposition (pCVD) diamond substrates and is equipped with readout electronics up to 2GHz analog bandwidth. In this paper we present the design of the pCVD-detector system and results of tests performed in various particle accelerator based facilities. Measurements performed in clinical high energy photon beams agreed within 1.2% with results obtained by standard ionization chambers. (C) 2007 Published by Elsevier B.V.
引用
收藏
页码:103 / 106
页数:4
相关论文
共 50 条
  • [1] Medical imaging applications in pre-clinical and clinical medicine
    Wiebe, LI
    PROCEEDINGS OF THE 3RD INTERNATIONAL CONFERENCE ON ISOTOPES: ISOTOPE PRODUCTION AND APPLICATIONS IN THE 21ST CENTURY, 2000, : 203 - 209
  • [2] MOXALACTAM - PRE-CLINICAL EVALUATION
    CHABBERT, YA
    SEMAINE DES HOPITAUX, 1983, 59 (26): : 1943 - 1946
  • [3] Beam monitor detectors for medical applications
    Nesteruk, Konrad P.
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2014, 19 : S32 - S36
  • [4] Medical physics in pre-clinical research
    Helbich, Thomas
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2014, 24 (04): : 271 - 271
  • [5] Pre-clinical irradiators: Advancements and applications
    Jaffray, D.
    Stewart, J.
    Weersink, R.
    Lindsay, P.
    Wang, K.
    Shah, D.
    DaCosta, R.
    Hill, R.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S140 - S140
  • [6] Pre-clinical in vitro evaluation of fluoroquinolones
    Dalhoff, A
    JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2001, 47 : 4 - 4
  • [7] PRE-CLINICAL EVALUATION OF ORTHOPEDIC IMPLANTS
    PIOTROWSKI, G
    BUTLER, JH
    RICHARDSON, DC
    JOURNAL OF MECHANICAL DESIGN-TRANSACTIONS OF THE ASME, 1980, 102 (04): : 655 - 662
  • [8] PRE-CLINICAL EVALUATION OF ANTINEPHRITIC DRUGS
    SUZUKI, Y
    ITO, M
    NAGAMATSU, T
    JOURNAL OF PHARMACOBIO-DYNAMICS, 1983, 6 (03): : S52 - S52
  • [9] THE PRE-CLINICAL EVALUATION OF ABORTIFACIENT DRUGS
    ELGER, W
    KOSUB, B
    KORTE, R
    POGGEL, HA
    TERATOLOGY, 1981, 24 (02) : A28 - A28
  • [10] PRE-CLINICAL SAFETY EVALUATION OF CEFTAZIDIME
    CAPELEDWARDS, K
    LOSTY, CR
    TUCKER, ML
    PRATT, DAH
    JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 1981, 8 : 237 - 239