Intraoperative detection of stuck leaflet after implantation of a mechanical aortic valve: a case report

被引:1
|
作者
Miyata, Kazuto [1 ]
Shigematsu, Sayaka [1 ]
机构
[1] New Heart Watanabe Inst, Dept Anesthesia, Suginami Ku, Hamadayama 3-19-11, Tokyo 1680065, Japan
关键词
Stuck leaflet; Mechanical aortic valve; Suture materials; SURGERY;
D O I
10.1186/s40981-020-0313-3
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background Mechanical aortic valve dysfunction is a rare but potentially fatal complication. It is usually caused by thrombosis, vegetation, and pannus overgrowth. However, it very rarely occurs immediately after the cardiopulmonary bypass weaning period. Case presentation We describe a case of stuck leaflet after the implantation of a mechanical aortic valve because of residual suture materials interfering with mechanical aortic valve leaflet closure, which is a very rare cause and has not been reported until now. Conclusion The residual suture materials for mechanical aortic valve replacement could cause mechanical valve dysfuction indicated as stuck leafet.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Intraoperative detection of stuck leaflet after implantation of a mechanical aortic valve: a case report
    Kazuto Miyata
    Sayaka Shigematsu
    JA Clinical Reports, 6
  • [2] Intermittent stuck valve after aortic valve replacement with a mechanical valve A case report
    Luo, Wenzong
    Wang, Xinxin
    Li, Jing
    Mu, Yun
    Ni, Yiming
    MEDICINE, 2017, 96 (09)
  • [3] Stuck leaflet after transcatheter aortic valve implantation due to calcium chunk in a patient with bicuspid valve
    Kawamura, Hiroto
    Yoshizaki, Toru
    Sato, Akira
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2024, 8 (06)
  • [4] A stuck leaflet after balloon post-dilatation in transcatheter aortic valve implantation with a SAPIEN-3 ultra RESILIA valve: a case report
    Yamazoe, Shinji
    Ogawa, Yasuhiro
    Kano, Naoaki
    Mamiya, Keita
    Kawaguchi, Katsuhiro
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2025, 9 (01)
  • [5] Embolization of aortic valve leaflet during valve-in-valve transcatheter aortic valve implantation: a case report
    Kinthala, Sudhakar
    Saththasivam, Poovendran
    Ankam, Abistanand
    Sattur, Sudhakar
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2020, 4 (01)
  • [6] Blockage of a Mechanical Aortic Valve Leaflet with BioGlue: A Case Report
    Gabrijelcic, Tone
    HEART SURGERY FORUM, 2012, 15 (06): : E310 - E312
  • [7] Leaflet thrombosis after valve-in-valve transcatheter aortic valve implantation: a case series
    Aktuerk, Dincer
    Mirsadraee, Saeed
    Quarto, Cesare
    Davies, Simon
    Duncan, Alison
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2020, 4 (04) : 1 - 6
  • [8] An Unexpected Finding Stuck Leaflet After Transapical Mitral Valve-in-Valve Implantation
    D'Onofrio, Augusto
    Gallo, Michele
    Tarantini, Giuseppe
    Cucchini, Umberto
    Pittarello, Demetrio
    Gerosa, Gino
    JACC-CARDIOVASCULAR INTERVENTIONS, 2014, 7 (12) : S187 - S189
  • [9] Acute myocardial infarction associated with prosthetic valve leaflet thrombosis after transcatheter aortic valve implantation: a case report
    Loria, Johannes Rotta Detto
    Thiele, Holger
    Abdel-Wahab, Mohamed
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2020, 4 (06)
  • [10] Stuck valve leaflet detected by intraoperative transesophageal echocardiography
    Kumano, H
    Suehiro, S
    Shibata, T
    Hattori, K
    Kinoshita, H
    ANNALS OF THORACIC SURGERY, 1999, 67 (05): : 1484 - 1485