Comparison of the recommendations of the AAPM TG-51 and TG-51 addendum reference dosimetry protocols

被引:2
|
作者
McCaw, Travis J. [1 ,2 ,3 ]
Hwang, Min-Sig [2 ,3 ]
Jang, Si Young [2 ,3 ]
Huq, M. Saiful [2 ,3 ]
机构
[1] Univ Wisconsin, Dept Human Oncol, Madison, WI 53706 USA
[2] Univ Pittsburgh, Dept Radiat Oncol, Inst Canc, Pittsburgh, PA 15260 USA
[3] UPMC, Ctr Canc, Pittsburgh, PA 15260 USA
来源
关键词
photon beam reference dosimetry; TG-51; addendum; PHOTON;
D O I
10.1002/acm2.12110
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work quantified differences between recommendations of the TG-51 and TG-51 addendum reference dosimetry protocols. Reference dosimetry was performed for flattened photon beams with nominal energies of 6, 10, 15, and 23 MV, as well as flattening-filter free (FFF) beam energies of 6 and 10 MV, following the recommendations of both the TG-51 and TG-51 addendum protocols using both a Farmer (R) ionization chamber and a scanning ionization chamber with calibration coefficients traceable to absorbed dose-to-water (D-w) standards. Differences in D-w determined by the two protocols were 0.1%-0.3% for beam energies with a flattening filter, and up to 0.2% and 0.8% for FFF beams measured with the scanning and Farmer (R) ionization chambers, respectively, due to k(Q) determination, volume-averaging correction, and collimator jaw setting. Combined uncertainty was between 0.91% and 1.2% (k = 1), varying by protocol and detector.
引用
收藏
页码:140 / 143
页数:4
相关论文
共 50 条
  • [1] Comparison of the TG-51 and TG-51 Addendum Calibration Protocols
    McCaw, T.
    Hwang, M.
    Jang, S.
    Huq, M.
    MEDICAL PHYSICS, 2016, 43 (06) : 3575 - 3575
  • [2] Clinical reference dosimetry: Comparison between AAPM TG-21 and TG-51 protocols
    Ding, GX
    Cygler, JE
    Kwok, CB
    MEDICAL PHYSICS, 2000, 27 (06) : 1217 - 1225
  • [3] Comparison of Two Photon Dosimetry Protocols: AAPM TG-51 & IPSM 1990
    Castrillon, S. Vargas
    Cutanda, F.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [4] Reference dosimetry in clinical high-energy electron beams: Comparison of the AAPM TG-51 and AAPM TG-21 dosimetry protocols
    Huq, MS
    Song, HJ
    Andreo, P
    Houser, CJ
    MEDICAL PHYSICS, 2001, 28 (10) : 2077 - 2087
  • [5] Reference dosimetry in clinical high-energy photon beams: Comparison of the AAPM TG-51 and AAPM TG-21 dosimetry protocols
    Huq, MS
    Andreo, P
    MEDICAL PHYSICS, 2001, 28 (01) : 46 - 54
  • [6] AAPM WGTG51 Report 374: Guidance for TG-51 reference dosimetry
    Muir, Bryan
    Culberson, Wesley
    Davis, Stephen
    Kim, Grace Gwe-Ya
    Lee, Sung-Woo
    Lowenstein, Jessica
    Renaud, James
    Sarfehnia, Arman
    Siebers, Jeffrey
    Tantot, Laurent
    Tolani, Naresh
    MEDICAL PHYSICS, 2022, 49 (11) : 6739 - 6764
  • [7] TG-51 Addendum - Overview and Implementation
    McEwen, M.
    MEDICAL PHYSICS, 2012, 39 (06) : 4006 - 4006
  • [8] The physics of the TG-51 dosimetry protocol
    Rogers, D. W. O.
    MEDICAL PHYSICS, 2007, 34 (06) : 2581 - 2582
  • [9] Addendum to the AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon beams
    McEwen, Malcolm
    DeWerd, Larry
    Ibbott, Geoffrey
    Followill, David
    Rogers, David W. O.
    Seltzer, Stephen
    Seuntjens, Jan
    MEDICAL PHYSICS, 2014, 41 (04)
  • [10] TG-51 reference dosimetry for the Halcyon™: A clinical experience
    Lloyd, Samantha A. M.
    Lim, Tze Yee
    Fave, Xenia
    Flores-Martinez, Everardo
    Atwood, Todd F.
    Moiseenko, Vitali
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (04): : 98 - 102