A high-throughput liquid chromatography-electrospray ionization mass spectrometric (LC-ESI-MS) method for screening of sirolimus on dried blood spots (DBS) was developed and validated. It involves solvent extraction of a punch of DBS followed by reversed-phase LC on a relatively new monolithic column consisting of a silica rod with bimodal pore structure and detection by ESI-MS. The run time was less than 3 min with a very low backpressure at a flow rate of 0.5 mL/min. The method can analyze more than 100 samples in an 8 h working day, including sample preparation. The assay was linear from 1 to 100 ng/mL. The mean recovery was 92.42%. The mean inter-day and intra-day precisions were 1.23 and 1.41%, respectively. The developed method is simple, rapid and useful for clinical applications. Copyright (C) 2010 John Wiley & Sons, Ltd.
机构:
Indian Inst Chem Technol, Div Analyt Chem, Hyderabad 500607, Andhra Pradesh, IndiaIndian Inst Chem Technol, Div Analyt Chem, Hyderabad 500607, Andhra Pradesh, India
Rao, R. Nageswara
Raju, S. Satyanarayana
论文数: 0引用数: 0
h-index: 0
机构:
Indian Inst Chem Technol, Organ Div 1, Hyderabad 500607, Andhra Pradesh, IndiaIndian Inst Chem Technol, Div Analyt Chem, Hyderabad 500607, Andhra Pradesh, India
Raju, S. Satyanarayana
Vali, R. Mastan
论文数: 0引用数: 0
h-index: 0
机构:
Indian Inst Chem Technol, Div Analyt Chem, Hyderabad 500607, Andhra Pradesh, IndiaIndian Inst Chem Technol, Div Analyt Chem, Hyderabad 500607, Andhra Pradesh, India
Vali, R. Mastan
Sarma, V. U. M.
论文数: 0引用数: 0
h-index: 0
机构:
Indian Inst Chem Technol, Organ Div 1, Hyderabad 500607, Andhra Pradesh, IndiaIndian Inst Chem Technol, Div Analyt Chem, Hyderabad 500607, Andhra Pradesh, India