Thermal calibration procedure for internal infrared laser deflection apparatus -: art. no. 094905

被引:5
|
作者
Perpiñà, X
Jordà, X
Madrid, F
Flores, D
Hidalgo, S
Vellvehi, M
Mestres, N
机构
[1] CSIC, IMB, CNM, Barcelona 08193, Spain
[2] CSIC, ICMAB, Barcelona 08193, Spain
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2005年 / 76卷 / 09期
关键词
D O I
10.1063/1.2039687
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A procedure to calibrate the temperature measurements in the transient regime of internal IR-laser deflection (IIR-LD) apparatus is presented. For this purpose, a thermal test chip (TTC), whose behavior is analytically well described by a simple model, is used. During the calibration process, the TTC is thermally excited during short heating times (up to 250 mu s), estimating its internal temperature profile by IIR-LD measurements. Afterwards, experimental results and model predictions are compared. Good agreement between theory and experiment is found when a temperature rise ranging from 0 to 1.4 K is measured. The presented procedure can be also used to thermally calibrate optical probing apparatus for measuring the thermal behavior of power devices, as well as to determine thermal parameters of other materials, such as SiC and GaN. In particular, it should be very useful for the determination of the thermo-optical coefficient (partial derivative n/partial derivative T)(C) of such materials. In the present work, a value for (partial derivative n/partial derivative T)(C) of 2.0x10(-4) K-1 is found in silicon, which agrees with literature reported values. The benefits of this method are its simplicity, accuracy, and low time consumption. (c) 2005 American Institute of Physics.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Calibration research of laser speedometer - art. no. 683513
    Cui, Yan-mei
    Li, Tao
    Leng, He
    Shi, Hui-sheng
    INFRARED MATERIALS, DEVICES, AND APPLICATIONS, 2007, 6835 : 83513 - 83513
  • [2] Scanning laser apparatus in medicine - art. no. 62511D
    Leontieva, Natali V.
    Savinov, Igor P.
    Efimenko, Vladimir T.
    Efimenko, Aleksej V.
    Lasers for Measurements and Information Transfer 2005, 2005, 6251 : D2511 - D2511
  • [3] Temperature uncertainty of IR Thermal Imager calibration - art. no. 620509
    Peacock, G. Raymond
    Thermosense XXVIII, 2006, 6205 : 20509 - 20509
  • [4] Analysis of infrared interferometry with laser noises - art. no. 67230M
    Cao Xiangke
    Zhang Rongzhu
    3RD INTERNATIONAL SYMPOSIUM ON ADVANCED OPTICAL MANUFACTURING AND TESTING TECHNOLOGIES: OPTICAL TEST AND MEASUREMENT TECHNOLOGY AND EQUIPMENT, PARTS 1-3, 2007, 6723 : M7230 - M7230
  • [5] Photodetachment of H- in a strong infrared laser field -: art. no. 243001
    Reichle, R
    Helm, H
    Kiyan, IY
    PHYSICAL REVIEW LETTERS, 2001, 87 (24)
  • [6] Infrared laser beam temporal fluctuations:: characterization and filtering -: art. no. 054203
    López-Alonso, JM
    Monacelli, B
    Alda, J
    Boreman, GD
    OPTICAL ENGINEERING, 2005, 44 (05) : 1 - 10
  • [7] Polarimetric scene modeling in the thermal infrared - art. no. 66820C
    Gartley, M. G.
    Brown, S. D.
    Goodenough, A. D.
    Sanders, N. J.
    Schott, J. R.
    POLARIZATION SCIENCE AND REMOTE SENSING III, 2007, 6682 : C6820 - C6820
  • [8] Maximum power of quantum dot laser versus internal loss - art. no. 073107
    Asryan, LV
    APPLIED PHYSICS LETTERS, 2006, 88 (07)
  • [9] β factor in a random laser -: art. no. 047601
    van Soest, G
    Lagendijk, A
    PHYSICAL REVIEW E, 2002, 65 (04):
  • [10] Ferrofluids as thermal ratchets -: art. no. 060602
    Engel, A
    Müller, HW
    Reimann, P
    Jung, A
    PHYSICAL REVIEW LETTERS, 2003, 91 (06) : 1 - 060602