Interlayer spray ionization mass spectrometry for the simple direct analysis of low amounts of sample

被引:10
|
作者
Chen, Jin [1 ]
Tang, Fei [1 ]
Guo, Cheng'an [1 ]
Huo, Xinming [1 ]
Zhang, Sichun [2 ]
Wang, Xiaohao [1 ]
机构
[1] Tsinghua Univ, Dept Precis Instrument, State Key Lab Precis Measurement Technol & Instru, Beijing 100084, Peoples R China
[2] Tsinghua Univ, Beijing Key Lab Microanalyt Methods & Instrumenta, Dept Chem, Beijing 100084, Peoples R China
基金
中国国家自然科学基金;
关键词
Ionization source; Trace analysis; Sandwich structure; Interlayer spray; Mass spectrometry; PAPER SPRAY; ELECTROSPRAY-IONIZATION; QUANTITATIVE-ANALYSIS; NANO-ELECTROSPRAY; DRUGS; MS;
D O I
10.1007/s00216-016-9600-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Interlayer spray is proposed as a convenient ionization source for direct analysis by mass spectrometry. Two slices of non-absorbent substrate hold the liquid sample to form a sandwich structure. By applying a high voltage to the sample, spray is generated at the tip of the substrate. The sampling procedure can be operated easily in an open condition and the spray is processed in a semi-enclosed condition, which leads to a relatively stable process. An ultralow amount (< 2 mu L) of the liquid sample can be analyzed without dilution, which ensures that the natural concentration and properties of the target are maintained. Less influence from the substrate is achieved compared with the spray methods based on porous absorbent materials, which results in a sensitivity enhancement of large molecule samples. It is demonstrated that the interlayer spray is applicable for the analysis of various compounds, including therapeutic drugs, peptides, and proteins. Good linearity can be obtained at a concentration as low as 50 ng/mL in the quantitative analysis for imatinib. We also show the ability to identify the chemical residuals on surfaces with high sensitivity by the "wipe-spray" method, which is useful for the fast screening of illicit substances. Interlayer spray working with mass spectrometry provides a promising method for direct analysis in an ambient environment.
引用
收藏
页码:5005 / 5012
页数:8
相关论文
共 50 条
  • [1] Interlayer spray ionization mass spectrometry for the simple direct analysis of low amounts of sample
    Jin Chen
    Fei Tang
    Cheng’an Guo
    Xinming Huo
    Sichun Zhang
    Xiaohao Wang
    Analytical and Bioanalytical Chemistry, 2016, 408 : 5005 - 5012
  • [2] Vibrating tip spray ionization mass spectrometry for direct sample analysis
    Yao, Ya-Nan
    Wu, Lin
    Di, Dandan
    Yuan, Zi-Cheng
    Hu, Bin
    JOURNAL OF MASS SPECTROMETRY, 2019, 54 (09): : 772 - 779
  • [3] Direct analysis of herbicides by paper spray ionization mass spectrometry
    Reeber, Steven L.
    Gadi, Sneha
    Huang, Sung-Ben
    Glish, Gary L.
    ANALYTICAL METHODS, 2015, 7 (23) : 9808 - 9816
  • [4] Paper spray ionization devices for direct, biomedical analysis using mass spectrometry
    Yang, Qian
    Wang, He
    Maas, Jeffrey D.
    Chappell, William J.
    Manicke, Nicholas E.
    Cooks, R. Graham
    Ouyang, Zheng
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2012, 312 : 201 - 207
  • [5] Ionization Suppression and Recovery in Direct Biofluid Analysis Using Paper Spray Mass Spectrometry
    Vega, Carolina
    Spence, Corina
    Zhang, Chengsen
    Bills, Brandon J.
    Manicke, Nicholas E.
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2016, 27 (04) : 726 - 734
  • [6] Rapid analysis of Callicarpa L. using direct spray ionization mass spectrometry
    Liu, Jingjing
    Gu, Zhixin
    Yao, Shouzuo
    Zhang, Zhaohui
    Chen, Bo
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2016, 124 : 93 - 103
  • [7] Teflon Spray Ionization Mass Spectrometry
    Narayanan, Rahul
    Song, Xiaowei
    Chen, Hao
    Zare, Richard N.
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2020, 31 (02) : 234 - 239
  • [8] Comprehensive comparison of ambient mass spectrometry with desorption electrospray ionization and direct analysis in real time for direct sample analysis
    Li, Lei
    Wang, Qiqin
    Li, Wen
    Yao, Ya-Nan
    Wu, Lin
    Hu, Bin
    TALANTA, 2019, 203 : 140 - 146
  • [9] Direct Biological Sample Analyses by Laserspray Ionization Miniature Mass Spectrometry
    Zhai, Yanbing
    Liu, Siyu
    Gao, Lijuan
    Hu, Lili
    Xu, Wei
    ANALYTICAL CHEMISTRY, 2018, 90 (09) : 5696 - 5702
  • [10] Conductive Polymer Spray Ionization Mass Spectrometry for Biofluid Analysis
    Song, Xiaowei
    Chen, Hao
    Zare, Richard N.
    ANALYTICAL CHEMISTRY, 2018, 90 (21) : 12878 - 12885