摘要
We have developed a high T-c superconducting quantum interference device (SQUID) magnetometer utilizing a Cu pickup coil cooled to T = 77 K and a resonant-type coupling circuit. The thermal noise can be decreased by cooling the pickup coil, while the signal transferred to the SQUID can be increased by using the resonant-type coupling circuit. Four pickup coils with different radii were used in the experiment. For the pickup coil with a radius r = 15 mm and the number of turns n = 150, we obtained the magnetic field noise of S-B(1/2) = 26 fT/Hz(1/2) at the resonant frequency of f(r) = 6.37 kHz. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:1396 / 1400
页数:5
相关论文
共 26 条
  • [1] HTS SQUID Magnetometer Using Resonant Coupling of Cooled Cu Pickup Coil
    Enpuku, K.
    Hirakawa, S.
    Tsuji, Y.
    Momotomi, R.
    Matsuo, M.
    Yoshida, T.
    Kandori, A.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2011, 21 (03) : 514 - 517
  • [2] Magnetometer utilizing SQUID pico-voltmeter and cooled normal pickup coil
    Yang, TQ
    Yao, K
    Yamasaki, D
    Enpuku, K
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2005, 426 : 1596 - 1600
  • [3] High Tc superconducting quantum interference device magnetometer utilizing cooled copper pickup coil and resonant coupling circuit
    Yang, TQ
    Enpuku, K
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS, 2002, 41 (11A): : L1221 - L1223
  • [4] Performance of HTS SQUID using resonant coupling of cooled Cu pickup coil
    Enpuku, K.
    Hirakawa, S.
    Momotomi, R.
    Matsuo, M.
    Yoshida, T.
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2011, 471 (21-22): : 1234 - 1237
  • [5] Detection of magnetic nanoparticles utilizing cooled normal pickup coil and high Tc SQUID
    Enpuku, K.
    Nabekura, S.
    Tsuji, Y.
    Okada, S.
    Matsuo, M.
    Tsukamoto, A.
    Mizoguchi, T.
    Kandori, A.
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2009, 469 (15-20): : 1634 - 1637
  • [6] Detection of magnetic nanoparticle utilizing cooled pickup coil and SQUID picovoltmeter
    Tazoh, Kazuya
    Nabekura, Shinya
    Enpuku, Keiji
    Research Reports on Information Science and Electrical Engineering of Kyushu University, 2008, 13 (01): : 37 - 42
  • [7] A RESONANT-TYPE AMORPHOUS RIBBON MAGNETOMETER DRIVEN BY AN OPERATIONAL-AMPLIFIER
    TAKEUCHI, S
    HARADA, K
    IEEE TRANSACTIONS ON MAGNETICS, 1984, 20 (05) : 1723 - 1725
  • [8] RESONANT-TYPE AMORPHOUS RIBBON MAGNETOMETER DRIVEN BY AN OPERATIONAL AMPLIFIER.
    Takeuchi, S.
    Harada, K.
    IEEE Transactions on Magnetics, 1984, MAG-20 (05)
  • [9] Accurate circuit analysis of resonant-type left handed transmission lines with inter-resonator coupling
    Gil, I.
    Bonache, J.
    Gil, M.
    Garcia-Garcia, J.
    Martin, F.
    Marques, R.
    JOURNAL OF APPLIED PHYSICS, 2006, 100 (07)