摘要
This paper reports on a low-cost, quantitative, pointof-care solution for the early detection of nitrite, a common biomarker for urinary tract infections (UTIs). In a healthy individual, nitrite is not found in the urine. However, a subject with a suspected UTI will produce nitrite in their urine since the bacteria present will convert nitrate into nitrite. Traditionally, nitrite is monitored by urinary dipsticks that are either read by eye or using a reflectance spectrophotometer. Both methods provide a semiquantitative positive or negative result at best. In this paper, we described a novel, affordable, portable transmission-based colorimeter for the quantitative measurement of nitrite. A unique permutation of the Griess reaction was optimized for the clinical detection of nitrite in urine and is reported. By using nitrite spiked in a salt buffer, artificial, and human urine samples, the performance of the colorimeter was evaluated against dipsticks read using two commercial dipstick analyzers, Urisys 1100 (Roche Diagnostics) and Clinitek Status+ (Siemens Medical Solutions). The colorimeter was able to detect the clinically relevant range of nitrite from 0.78 to 200 mu M in a salt buffer. The detection limit in artificial urine was determined as 1.6 mu M, which is similar to 16x more sensitive than commercial dipstick reflectance analyzers, enabling the possibility for earlier detection of urinary infections. The colorimeter is assembled using off-the-shelf components (<$80) and controlled by a smartphone application via low-energy bluetooth. It has a built-in color correction algorithm and is designed to enable for a turbidity correction in samples containing bacteria or other cellular debris as well. The mobile application can display the nitrite concentration for a single sample or display the results over a period of time. Tracking urinalysis results longitudinally can help identify trends such as increases in nitrite concentrations over an individual's baseline and identify possible infections earlier. While the detection of nitrite was showcased here, this portable analyzer can be expanded to other colorimetric-based chemistries to detect a panel of biomarkers, which can improve the overall sensitivity and specificity of the desired assay.
引用
收藏
页码:11126 / 11134
页数:9
相关论文
共 50 条
  • [1] Quantitative Point-of-Care Colorimetric Assay Modeling Using a Handheld Colorimeter
    Abels, Kristen
    Salvo-Halloran, Elizabeth M.
    White, Dawn
    Ali, Monsur
    Agarwal, Nisha R.
    Leung, Vincent
    Ali, Muntakim
    Sidawi, Mariam
    Capretta, Alfredo
    Brennan, John D.
    Nease, Jake
    Filipe, Carlos D. M.
    ACS OMEGA, 2021, 6 (34): : 22439 - 22446
  • [2] Quantitative Point-of-Care Pathogen Detection in Septicemia
    Tran, Nam K.
    Wisner, David
    Albertson, Timothy
    Cohen, Stuart
    Greenhalgh, David
    Galante, Joseph
    Phan, Ho
    Kost, Gerald J.
    POINT OF CARE, 2008, 7 (03): : 107 - 110
  • [3] Toward quantitative assessment of deformational plagiocephaly and brachycephaly at the point-of-care
    Seifabadi, Reza
    Aalamifar, Fereshteh
    Hezaveh, Seyed Hossein
    Kocabalkanli, Can
    Wilburn, Kelly
    Linguraru, Marius George
    JOURNAL OF MEDICAL IMAGING, 2021, 8 (02)
  • [4] Metal-organic framework-based ratiometric point-of-care testing for quantitative visual detection of nitrite
    Liang, Maosheng
    Gao, Yifan
    Sun, Xiaoling
    Kong, Rong-Mei
    Xia, Lian
    Qu, Fengli
    JOURNAL OF HAZARDOUS MATERIALS, 2024, 469
  • [5] Metal-organic framework-based ratiometric point-of-care testing for quantitative visual detection of nitrite
    Liang, Maosheng
    Gao, Yifan
    Sun, Xiaoling
    Kong, Rong-Mei
    Xia, Lian
    Qu, Fengli
    Journal of Hazardous Materials, 2024, 469
  • [6] Moving Toward Point-of-care?
    Blow, Nathan S.
    BIOTECHNIQUES, 2011, 50 (01) : 11 - 11
  • [7] Electrochemical aptasensors for pathogenic detection toward point-of-care diagnostics
    Kizilkurtlu, Ahmet Akif
    Demirbas, Erhan
    Agel, Hatice Esra
    BIOTECHNOLOGY AND APPLIED BIOCHEMISTRY, 2023, 70 (04) : 1460 - 1479
  • [8] Design of a Quantitative Readout in a Point-of-Care Device for Cisplatin Detection
    Ragusa, Edoardo
    Zunino, Rodolfo
    Mastronardi, Valentina
    Moglianetti, Mauro
    Pompa, Pier P.
    Gastaldo, Paolo
    IEEE SENSORS LETTERS, 2022, 6 (11)
  • [9] Toward Point-of-care Uropathogen Detection Using SERS Active Filters
    Dryden, Simon D.
    Anastasova, Salzitsa
    Satta, Giovanni
    Thompson, Alex J.
    Leff, Daniel R.
    Darzi, Ara
    OPTICAL DIAGNOSTICS AND SENSING XX: TOWARD POINT-OF-CARE DIAGNOSTICS, 2020, 11247
  • [10] Toward point-of-care microchip profiling of proteins
    Osiri, John K.
    Shadpour, Hamed
    BIOANALYSIS, 2010, 2 (10) : 1745 - 1754