A Zr-89 solution was measured by means of liquid scintillation counting techniques in order to determine the activity concentration. Two methods were used: the CIEMAT/NIST efficiency tracing method with H-3 as a tracer, and the triple-to-double coincidence ratio method. The counting efficiencies were computed with a stochastic model. The very detailed investigation showed that a few corrections are particularly important: Asymmetries in the photodetector responses as well as the backscattering of high-energy gamma rays must be taken into account. Corresponding corrections have therefore been applied. In addition, a detailed uncertainty analysis was carried out and the uncertainties compared with those determined by other research groups.The activity concentrations obtained from the two methods agree well and a combined result was used to establish calibration factors for ionization chambers, which are important secondary standardization instruments. The ionization chambers were combined with a new high-precision current measurement device to provide outstanding linearity. Measurement data from one chamber were used to determine the half-life, which was found to be T-1/2=(78.373 & PLUSMN; 0.023) h.