Fast collimated neutron flux measurement using stilbene scintillator and flashy analog-to-digital converter in JT-60U

被引:24
|
作者
Ishikawa, M.
Itoga, T.
Okuji, T.
Nakhostin, M.
Shinohara, K.
Hayashi, T.
Sukegawa, A.
Baba, M.
Nishitani, T.
机构
[1] Tohoku Univ, Ctr Cyclotron & Radioisotope, Sendai, Miyagi 9808578, Japan
[2] Japan Atom Energy Agcy, Ibaraki 3110193, Japan
[3] Tohoku Univ, Ctr Cyclotron & Radioisotope, Sendai, Miyagi 9808578, Japan
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2006年 / 77卷 / 10期
关键词
D O I
10.1063/1.2221927
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A line-integrated neutron emission profile is routinely measured using the radial neutron collimator system in JT-60U tokamak. Stilbene neuron detectors (SNDs), which combine a stilbene organic crystal scintillation detector (SD) with an analog neutron-gamma pulse shape discrimination (PSD) circuit, have been used to measure collimated neutron flux. Although the SND has many advantages as a neutron detector, the maximum count rate is limited up to similar to 1 x 10(5) counts/s due to the analog PSD circuit. To overcome this issue, a digital signal processing system (DSPS) using a flash analog-to-digital converter (Acqiris DC252, 8 GHz, 10 bits) has been developed at Cyclotron and Radioisotope Center in Tohoku University. In this system anode signals from photomultiplier of the SD are directory stored and digitized. Then, the PSD between neutrons and gamma rays is performed using software. The DSPS has been installed in the vertical neutron collimator system in JT-60U and applied to deuterium experiments. It is confirmed that the PSD is sufficiently performed and collimated neutron flux is successfully measured with count rate up to similar to 5 x 10(5) counts/s without the effect of pileup of detected pulses. The performance of the DSPS as a neutron detector, which supersedes the SND, is demonstrated. (c) 2006 American Institute of Physics.
引用
收藏
页数:3
相关论文
共 29 条
  • [1] First measurement of neutron emission profile on JT-60U using Stilbene neutron detector with neutron-gamma discrimination
    Ishikawa, M
    Nishitani, T
    Morioka, A
    Takechi, M
    Shinohara, K
    Shimada, M
    Miura, Y
    Nagami, M
    Kaschuck, YA
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (12): : 4237 - 4242
  • [2] Development of fast measurement system of neutron emission profile using a digital signal processing technique in JT-60U
    Ishikawa, M.
    Itoga, T.
    Okuji, T.
    Nakhostin, M.
    Shinohara, K.
    Baba, M.
    Nishitani, T.
    BURNING PLASMA DIAGNOSTICS, 2008, 988 : 295 - +
  • [3] Fast neutron-gamma discrimination on neutron emission profile measurement on JT-60U
    Ishii, K.
    Shinohara, K.
    Ishikawa, M.
    Baba, M.
    Isobe, M.
    Okamoto, A.
    Kitajima, S.
    Sasao, M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (10):
  • [4] INFRARED TV MEASUREMENT OF FAST-ION LOSS ON JT-60U
    TOBITA, K
    NEYATANI, Y
    KUSAMA, Y
    TAKEUCHI, H
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01): : 594 - 596
  • [5] First measurement of time-resolved neutron yield on JT-60U using a microfission chamber
    Hayashi, T
    Nishitani, T
    Ishikawa, M
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10): : 3575 - 3577
  • [6] High repetition rate neutron flux array using digital-signal processing and scintillators for study of high performance plasma at JT-60U
    Shinohara, K.
    Okuji, T.
    Ishikawa, M.
    Baba, M.
    Itoga, T.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (10):
  • [7] ABSOLUTE CALIBRATION OF THE JT-60U NEUTRON MONITORS USING A CF-252 NEUTRON SOURCE
    NISHITANI, T
    TAKEUCHI, H
    KONDOH, T
    ITOH, T
    KURIYAMA, M
    IKEDA, Y
    IGUCHI, T
    BARNES, CW
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (11): : 5270 - 5278
  • [8] Density fluctuation measurement using motional Stark effect optics in JT-60U
    Suzuki, T.
    Fujita, T.
    Oyama, N.
    Isayama, A.
    Matsunaga, G.
    Oikawa, T.
    Asakura, N.
    Takechi, M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (10):
  • [9] Precision timing measurement of phototube pulses using a flash analog-to-digital converter
    Bennett, J. V.
    Kornicer, M.
    Shepherd, M. R.
    Ito, M. M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 622 (01): : 225 - 230
  • [10] First measurements of triton burnup neutron spectra using a natural diamond detector on JT-60U
    Isobe, M
    Nishitani, T
    Krasilnikov, AV
    Kaneko, J
    Sasao, M
    FUSION ENGINEERING AND DESIGN, 1997, 34-35 : 573 - 576