To further improve the design of our direct-coupled SQUID magnetometers, we performed simulations of the current distribution and thus of the effective area and inductance for various distributions of the 16 parallel 50-mum-wide loops forming the pickup structure. It is shown that an optimization of the line distribution yields an increase of the effective area of only 2.2% but could be advantageous for the operation in time-varying magnetic fields. Based on the devices which were optimized with regard to their operation in magnetically unshielded environment, we have developed a single-channel second-order gradiometer system that allowed us to record real-time magnetocardiograms in a typical laboratory environment.
引用
收藏
页码:1315 / 1318
页数:4
相关论文
共 50 条
  • [1] Sensitive high-Tc SQUID magnetometers for unshielded operation
    DiIorio, M.S.
    Yang, K.-Y.
    Yoshizumi, S.
    Haupt, S.G.
    Haran, D.
    Koch, R.H.
    Milliken, F.P.
    Rozen, J.R.
    Lathrop, D.K.
    Kumar, S.
    Trammell III, H.S.
    IEEE Transactions on Applied Superconductivity, 1999, 9 (2 III): : 4428 - 4431
  • [2] Sensitive high-Tc SQUID magnetometers for unshielded operation
    DiIorio, MS
    Yang, KY
    Yoshizumi, S
    Haupt, SG
    Haran, D
    Koch, RH
    Milliken, FP
    Rozen, JR
    Lathrop, DK
    Kumar, S
    Trammell, HS
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1999, 9 (02) : 4428 - 4431
  • [3] Performance of high-Tc dc SQUID magnetometers for use in a magnetically disturbed environment
    Kim, IS
    Yu, KK
    Lee, YH
    Kim, KW
    Park, YK
    PHYSICA STATUS SOLIDI A-APPLIED RESEARCH, 2004, 201 (08): : 1969 - 1972
  • [4] High-Tc SQUID gradiometers for magnetocardiography in an unshielded environment
    Tralshawala, N
    Claycomb, JR
    Ying, Q
    Miller, JH
    Nesteruk, K
    Jackson, DR
    BIOMAG 96: PROCEEDINGS OF THE TENTH INTERNATIONAL CONFERENCE ON BIOMAGNETISM, VOLS I & II, 2000, : 134 - +
  • [5] High-Tc SQUID gradiometer system for magnetocardiography in an unshielded environment
    Liao, SH
    Hsu, SC
    Lin, CC
    Horng, HE
    Chen, JC
    Chen, MJ
    Wu, CH
    Yang, HC
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2003, 16 (12): : 1426 - 1429
  • [6] High-Tc SQUID magnetometers for use in moderate magnetically-shielded room
    Matsuda, M
    Ono, S
    Kato, K
    Matsuura, T
    Oyama, H
    Hayashi, A
    Hirano, S
    Kuriki, S
    Yokosawa, K
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2001, 11 (01) : 1323 - 1326
  • [7] DC SQUID Modulation Electronics for Operation With HTS DC SQUID Magnetometers in the Unshielded Environment
    Burmistrov, Evgeny V.
    Slobodchikov, V. Yu.
    Khanin, V. V.
    Maslennikov, Yu. V.
    Snigirev, O. V.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2009, 19 (03) : 206 - 209
  • [8] Slotted high-Tc dc SQUID magnetometers
    Jansman, ABM
    Izquierdo, M
    Flokstra, J
    Rogalla, H
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1999, 9 (02) : 3290 - 3293
  • [9] HIGH-TC SUPERCONDUCTING MULTILAYERS FOR SQUID MAGNETOMETERS
    CLARKE, J
    KINGSTON, JJ
    MIKLICH, AH
    WELLSTOOD, FC
    CHAR, K
    COLCLOUGH, MS
    LEE, LP
    ZAHARCHUK, G
    PHYSICA SCRIPTA, 1992, T42 : 51 - 56
  • [10] Crack detection for the graphite slab using the high-Tc SQUID in unshielded environment
    Jeng, JT
    Horng, HE
    Yang, HC
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2002, 372 : 174 - 177