Evaluation and Calibration of a Binocular Infrared Pupillometer for Measuring Relative Afferent Pupillary Defect

被引:14
|
作者
Shwe-Tin, Audrey [1 ]
Smith, Guy T. [1 ]
Checketts, Daniel
Murdoch, Ian E. [2 ]
Taylor, Daniel [3 ]
机构
[1] Great Western Hosp, Dept Ophthalmol, Swindon SN3 6BB, Wilts, England
[2] Inst Ophthalmol, London, England
[3] Procyon Instruments Ltd, London, England
关键词
D O I
10.1097/WNO.0b013e31823f45e5
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: Binocular infrared pupillometry allows an estimate of the relative afferent pupillary defect (RAPD), designated the pupillometric RAPD (pRAPD). We calibrated the pRAPD of a commercially available pupillometer against neutral density filters (NDFs) of known attenuation. The performance of the pupillometer using its own proprietary algorithm is assessed and compared to that of alternative algorithms. Methods: The pRAPDs of 50 healthy volunteers were measured with each of 4 filters of known attenuation: 00, 0.3, 0.6, and 0.9 log units, positioned unilaterally in the light stimulus pathway. The filter values were plotted against the pupillometer output, and the slope and intercept were used to determine a calibration factor. Corrected pRAPD results were used to assess physiological ranges of pRAPD. The sensitivity and specificity to 0.3 log unit differences between increasing filter densities using receiver operator characteristic (ROC) curves. Results: The calibrated physiological pRAPD ranged from 0 to 0.22 log units. The area under the ROC curve for detecting unilateral simulated pRAPD of 0.3 log units, the simulated disease progression from 0.3 to 0.6 log units, and a further progression from 0.6 to 0.9 log units by NDFs was 0.99 (95% confidence interval [CI], 0.95-1.00), 0.86 (95% CI, 0.78-0.92), and 0.79 (95% CI, 0.70-0.87), respectively. The optimum discrimination was for detecting a unilateral simulated pRAPD of 0.3 log units; sensitivity and specificity was 98% (95% CI, 88%-99%). Conclusion: The commercially available pupillometer detects the RAPD induced by the NDFs with high sensitivity and specificity. The results suggest that it is best for detecting unilateral early disease but potentially useful for assessing progression of disease.
引用
收藏
页码:111 / 115
页数:5
相关论文
共 50 条
  • [1] Relative Afferent Pupillary Defect Detected In Asymmetric Macular Disease Using An Automated Binocular Pupillometer
    Rahman, Shiraaz
    Grewal, Dilraj Singh
    Bhat, Pooja
    Fallor, Morgan
    Volpe, Nicholas J.
    Mirza, Rukhsana
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [2] Relative afferent pupillary defect, or afferent pupillary defect?
    Papalkar, D
    Francis, IC
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2006, 34 (02): : 193 - 193
  • [3] RELATIVE AFFERENT PUPILLARY DEFECT
    THOMPSON, HS
    CORBETT, JJ
    ARCHIVES OF NEUROLOGY, 1985, 42 (06) : 516 - 516
  • [4] Infrared pupillary asymmetry (IPA) for detection of a relative afferent pupillary defect (RAPD)
    Meneguette, N. S.
    De Carvalho, E.
    Petzold, A.
    MULTIPLE SCLEROSIS JOURNAL, 2018, 24 : 372 - 372
  • [5] Relative Afferent Pupillary Defect
    Manjunathan, Sujatha
    Saini, Lokesh
    NEW ENGLAND JOURNAL OF MEDICINE, 2022, 387 (08): : E15 - E15
  • [6] RELATIVE AFFERENT PUPILLARY DEFECT IN A LESION OF THE PRETECTAL AFFERENT PUPILLARY PATHWAY
    JOHNSON, RE
    BELL, RA
    CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 1987, 22 (05): : 282 - 284
  • [7] VARIABILITY OF THE RELATIVE AFFERENT PUPILLARY DEFECT
    KARDON, R
    MOORE, P
    SELKY, A
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1994, 35 (04) : 2059 - 2059
  • [9] VARIABILITY OF THE RELATIVE AFFERENT PUPILLARY DEFECT
    KAWASAKI, A
    MOORE, P
    KARDON, RH
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1995, 120 (05) : 622 - 633
  • [10] The detection of a relative afferent pupillary defect
    Glazer-Hockstein, C