Fabrication and performance of a μRWELL detector with Diamond-Like Carbon resistive electrode and two-dimensional readout

被引:16
|
作者
Zhou, Yi [1 ,2 ]
Lv, You [1 ,2 ]
Shang, Lunlin [3 ]
Hong, Daojin [1 ,2 ]
Song, Guofeng [1 ,2 ]
Liu, Jianbei [1 ,2 ]
Feng, Jianxin [1 ,2 ]
Shao, Ming [1 ,2 ]
Wang, Xu [1 ,2 ]
Zhang, Zhiyong [1 ,2 ]
机构
[1] Univ Sci & Technol China, State Key Lab Particle Detect & Elect, Hefei 230026, Anhui, Peoples R China
[2] Univ Sci & Technol China, Dept Modern Phys, Hefei 230026, Anhui, Peoples R China
[3] Chinese Acad Sci, Lanzhou Inst Chem Phys, State Key Lab Solid Lubricat, Lanzhou 730000, Gansu, Peoples R China
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 2019年 / 927卷
基金
中国国家自然科学基金;
关键词
MPGD; Micro-resistive WELL; Diamond-like carbon; Resistive electrode; GASEOUS DETECTORS;
D O I
10.1016/j.nima.2019.01.036
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The Micro-Resistive WELL (mu RWELL) detector is a novel spark-protected Micro-Pattern Gaseous Detector (MPGD) with a single well-type amplification stage. By integrating a well-type amplification structure with a resistive electrode on a readout board (mu RWELL PCB), the mu RWELL detector is only composed of a cathode PCB and a mu RWELL PCB, resulting in a very simple assembling process without any gluing, stretching, tensioning or other complicated operation commonly practiced in assemblies of other MPGDs. A critical component of the mu RWELL detector is the resistive electrode that is responsible for spark suppression, and hence maintains stable operation of the detector. In this paper, we present a 10 cm x 10 cm mu RWELL detector with two layers readout strips perpendicular to each other. Its resistive electrode is made of a thin coating of Diamond-Like Carbon by magnetron sputtering technique. The gain performance and rate capability of this mu RWELL detector were systematically studied with X-rays, and its detection efficiency and spatial resolution were measured with a test beam. The detector can achieve a spatial resolution better than 70 mu m in both dimensions while the detection efficiency is higher than 90%.
引用
收藏
页码:31 / 36
页数:6
相关论文
共 50 条
  • [1] Fabrication of diamond-like carbon-based two-dimensional photonic crystals
    Yun, J. G.
    Jung, D. H.
    Kim, H. J.
    Choi, W. S.
    Hong, B. Y.
    Noda, S.
    Song, B. S.
    MICROELECTRONIC ENGINEERING, 2014, 126 : 99 - 102
  • [2] Production and performance study of Diamond-Like Carbon resistive electrode in MPGD
    Lv, You
    Zhou, Yi
    Liu, Jianbei
    Shao, Ming
    Zhang, Zhiyong
    Song, Guofeng
    Wang, Xu
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 958
  • [3] Development of Diamond-like Carbon Resistive Electrode for Micro-pattern Gas Detector
    Shang L.
    Zhou Y.
    Lyu Y.
    Zhang G.
    Li W.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2020, 54 (06): : 968 - 974
  • [4] Atomic hydrogen storage in a two-dimensional hydrogenated diamond-like carbon film
    Ray, Nihar Ranjan
    CURRENT SCIENCE, 2017, 112 (07): : 1564 - 1567
  • [5] Two-dimensional imaging triple-GEM detector with resistive anode readout
    Ju, X. D.
    Dong, M. Y.
    Zhao, Y. C.
    Zhou, C. X.
    Dong, J.
    Zhao, Y. B.
    Zhang, H. Y.
    OuYang, Q.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [6] Superior stiffness and vibrational spectroscopic signature of two-dimensional diamond-like carbon nitrides
    Longuinhos, R.
    Ribeiro-Soares, J.
    PHYSICA E-LOW-DIMENSIONAL SYSTEMS & NANOSTRUCTURES, 2020, 119 (119):
  • [7] Monopolar Resistive Switching in Diamond-Like Carbon Films
    Vedeneev, A. S.
    Luzanov, V. A.
    Rylkov, V. V.
    SEMICONDUCTORS, 2019, 53 (14) : 1970 - 1973
  • [8] Monopolar Resistive Switching in Diamond-Like Carbon Films
    A. S. Vedeneev
    V. A. Luzanov
    V. V. Rylkov
    Semiconductors, 2019, 53 : 1970 - 1973
  • [9] Signal characteristics of a resistive-strip micromegas detector with an integrated two-dimensional readout
    Lin, Tai-Hua
    Duedder, Andreas
    Schott, Matthias
    Valderanis, Chrysostomos
    Wehner, Laura
    Westenberger, Robert
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 767 : 281 - 288
  • [10] A multi-electrode two-dimensional position sensitive diamond detector
    Ditalia Tchernij, S.
    Siciliano, D.
    Provatas, G.
    Forneris, J.
    Picollo, F.
    Campostrini, M.
    Rigato, V.
    Siketic, Z.
    Jaksic, M.
    Vittone, E.
    APPLIED PHYSICS LETTERS, 2024, 124 (22)