Use of a Novel Diagonal Pump in an In Vitro Neonatal Pulsatile Extracorporeal Life Support Circuit

被引:10
|
作者
Evenson, Alissa [1 ]
Wang, Shigang [1 ]
Kunselman, Allen R. [2 ]
Uendar, Akif [1 ,3 ,4 ]
机构
[1] Penn State Hershey Childrens Hosp, Pediat Cardiovasc Res Ctr, Penn State Milton S Hershey Med Ctr, Dept Pediat,Penn State Hershey Coll Med, Hershey, PA USA
[2] Penn State Hershey Coll Med, Dept Publ Hlth Sci, Hershey, PA USA
[3] Penn State Hershey Coll Med, Penn State Milton S Hershey Med Ctr, Dept Surg, Hershey, PA USA
[4] Penn State Univ, Dept Bioengn, Coll Engn, University Pk, PA 16802 USA
关键词
Extracorporeal life support; Diagonal pump; Pediatrics; Pulsatile; Surplus hemodynamic energy; NONPULSATILE PERFUSION; CARDIOPULMONARY BYPASS; OXYGEN-SATURATION; FLOW;
D O I
10.1111/aor.12240
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
One approach with the potential to improve morbidity and mortality rates following extracorporeal life support (ECLS) is the use of pulsatile perfusion. Currently, no ECLS pumps used in the United States can produce pulsatile flow. The objective of this experiment is to evaluate a novel diagonal pump used in Europe to determine whether it provides physiological pulsatility in a neonatal model. The ECLS circuit consisted of a Medos Deltastream DP3 diagonal pump, a Hilite 800LT polymethylpentene diffusion membrane oxygenator, and arterial/venous tubing. A 300-mL pseudopatient was connected to the circuit using an 8Fr arterial cannula and a 10Fr venous cannula. A clamp maintained constant pressure entering the pseudopatient. Trials (64 total) were conducted in nonpulsatile and pulsatile modes at flow rates of 200 mL/min to 800 mL/min. Flow and pressure data were collected using a custom-based data acquisition system. The Deltastream DP3 pump was capable of producing an adequate quality of pulsatility. Pulsatile flow produced increased mean arterial pressure, energy equivalent pressure (EEP), and surplus hemodynamic energy (SHE) at all flow rates compared to nonpulsatile flow. Pressure drop across the cannula accounted for the majority of pressure loss in the circuit. The greatest loss of SHE and total hemodynamic energy occurred across the arterial cannula due to its small diameter. The Deltastream DP3 pump produced physiological pulsatile flow without backflow while providing EEP and SHE to our neonatal pseudopatient. Further experiments are necessary to determine the impact of this pulsatile pump in an in vivo model prior to clinical use.
引用
收藏
页码:E1 / E9
页数:9
相关论文
共 50 条
  • [1] Novel Pulsatile Diagonal Pump for Pediatric Extracorporeal Life Support System
    Wang, Shigang
    Kunselman, Allen R.
    Uendar, Akif
    ARTIFICIAL ORGANS, 2013, 37 (01) : 37 - 47
  • [2] In Vitro Performance Analysis of a Novel Pulsatile Diagonal Pump in a Simulated Pediatric Mechanical Circulatory Support System
    Wang, Shigang
    Kunselman, Allen R.
    Uendar, Akif
    ARTIFICIAL ORGANS, 2014, 38 (01) : 64 - 72
  • [3] In Vitro Hemodynamic Evaluation of a Novel Pulsatile Extracorporeal Life Support System: Impact of Perfusion Modes and Circuit Components on Energy Loss
    Wang, Shigang
    Kunselman, Allen R.
    Clark, Joseph B.
    Uendar, Akif
    ARTIFICIAL ORGANS, 2015, 39 (01) : 59 - 66
  • [4] Pediatric Extracorporeal Life Support Using a Third Generation Diagonal Pump
    Speth, Marlene
    Muench, Frank
    Purbojo, Ariawan
    Gloeckler, Martin
    Toka, Okan
    Cesnjevar, Robert A.
    Rueffer, Andre
    ASAIO JOURNAL, 2016, 62 (04) : 482 - 490
  • [5] Development of a new pulsatile extracorporeal life support device incorporating a dual pulsatile blood pump
    Choi, S. W.
    Nam, K. W.
    Chung, J.
    Lee, J. C.
    Hwang, C. M.
    Won, Y. S.
    Sun, K.
    Min, B. G.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2006, 29 (06): : 583 - 590
  • [6] In Vitro Evaluation of an Alternative Neonatal Extracorporeal Life Support Circuit on Hemodynamic Performance and Bubble Trap
    Spencer, Shannon B.
    Wang, Shigang
    Woitas, Karl
    Glass, Kristen
    Kunselman, Allen R.
    Undar, Akif
    ARTIFICIAL ORGANS, 2017, 41 (01) : 17 - 24
  • [7] Impact of Pulsatile Flow Settings on Hemodynamic Energy Levels Using the Novel Diagonal Medos DP3 Pump in a Simulated Pediatric Extracorporeal Life Support System
    Adedayo, Pelumi
    Wang, Shigang
    Kunselman, Allen R.
    Undar, Akif
    WORLD JOURNAL FOR PEDIATRIC AND CONGENITAL HEART SURGERY, 2014, 5 (03) : 440 - 448
  • [8] Evaluation of Conventional Nonpulsatile and Novel Pulsatile Extracorporeal Life Support Systems in a Simulated Pediatric Extracorporeal Life Support Model
    Wang, Shigang
    Evenson, Alissa
    Chin, Brian J.
    Kunselman, Allen R.
    Uendar, Akif
    ARTIFICIAL ORGANS, 2015, 39 (01) : E1 - E9
  • [9] Comparison of Two Types of Neonatal Extracorporeal Life Support Systems With Pulsatile and Nonpulsatile Flow
    Haines, Nikkole
    Wang, Shigang
    Myers, John L.
    Undar, Akif
    ARTIFICIAL ORGANS, 2009, 33 (11) : 958 - 966
  • [10] Impact of Tubing Length on Hemodynamics in a Simulated Neonatal Extracorporeal Life Support Circuit
    Qiu, Feng
    Uluer, Mehmet C.
    Kunselman, Allen
    Clark, J. Brian
    Myers, John L.
    Uendar, Akif
    ARTIFICIAL ORGANS, 2010, 34 (11) : 1003 - 1009