Optimization of HIFU Treatment on the Basis of Temperature Distributions Measured by a Thin-film Thermocouple Array

被引:1
|
作者
Matsuki, Kosuke [1 ]
Narumi, Ryuta [1 ]
Azuma, Takashi [1 ]
Yoshinaka, Kiyoshi
Sasaki, Akira [1 ]
Okita, Kohei
Takagi, Shu [1 ]
Matsumoto, Yoichiro [1 ]
机构
[1] Univ Tokyo, Dept Mech Eng, Tokyo 1138654, Japan
关键词
FOCUSED ULTRASOUND; SURGERY;
D O I
10.1109/ULTSYM.2012.0405
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
It is required to shorten HIFU treatment time to reduce the burden of patients. With this view, we considered focus switching method at two points. To optimize switching parameters, it was necessary to measure temperature field under the exposure of ultrasound. In this study, we measured it by a prototype thin-film thermocouple array. This way has advantage to minimize the influence of thermocouple to acoustic and temperature field. Then we enlarged the temperature rise area by focus switching. Therefore, we could say that focus switching has the potential to reduce HIFU treatment time.
引用
收藏
页数:3
相关论文
共 50 条
  • [1] Thin-Film Thermocouple Array for Time-Resolved Local Temperature Mapping
    Liu, Haixiao
    Sun, Weiqiang
    Chen, Qing
    Xu, Shengyong
    IEEE ELECTRON DEVICE LETTERS, 2011, 32 (11) : 1606 - 1608
  • [2] TRANSPARENT THIN-FILM THERMOCOUPLE
    YUST, M
    KREIDER, KG
    THIN SOLID FILMS, 1989, 176 (01) : 73 - 78
  • [3] THIN-FILM THERMOCOUPLE GAUGE
    KUO, TC
    FLATTERY, J
    GHOSH, PK
    KORNREICH, PG
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1988, 6 (03): : 1150 - 1152
  • [4] Design, Fabrication, and Characterization of a Pt/Au Thin-Film Thermocouple Array
    Ding, Jiong
    Wang, Jiayin
    Xu, Qiyue
    Yang, Suijun
    Ye, Shuliang
    IEEE SENSORS JOURNAL, 2020, 20 (12) : 6287 - 6294
  • [5] Temperature Distributions Measurement of High Intensity Focused Ultrasound using a Thin-film Thermocouple Array and Estimation of Thermal Error caused by Viscous Heating
    Matsuki, Kosuke
    Narumi, Ryuta
    Azuma, Takashi
    Yoshinaka, Kiyoshi
    Sasaki, Akira
    Okita, Kohei
    Takagi, Shu
    Matsumoto, Yoichiro
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 3722 - 3725
  • [6] Metallic thin-film thermocouple for thermoelectric microgenerators
    Castano, E
    Revuelto, E
    Martin, MC
    GarciaAlonso, A
    Gracia, FJ
    SENSORS AND ACTUATORS A-PHYSICAL, 1997, 60 (1-3) : 65 - 67
  • [7] Research on a dedicated thin-film thermocouple testing system for transient temperature measurement
    Xie, Yingnan
    Cui, Yunxian
    Wang, Haoyu
    Feng, Wenrui
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2024, 35 (08)
  • [9] Dynamic Calibration of the Cutting Temperature Sensor of NiCr/NiSi Thin-film Thermocouple
    Cui Yunxian
    Yang Deshun
    Jia Ying
    Zeng Qiyong
    Sun Baoyuan
    CHINESE JOURNAL OF MECHANICAL ENGINEERING, 2011, 24 (01) : 73 - 77
  • [10] Development of NiCr/NiSi thin-film thermocouple sensor for cutting temperature measurement
    Zeng, QY
    Sun, BY
    Xu, J
    Deng, XL
    Xu, J
    Jia, Y
    ISTM/2005: 6th International Symposium on Test and Measurement, Vols 1-9, Conference Proceedings, 2005, : 3859 - 3862