Influence of the distal femoral resection angle on the principal stresses in ceramic total knee components

被引:11
|
作者
Kluess, Daniel [1 ]
Bergschmidt, Philipp [1 ]
Mueller, Iris [1 ]
Mittelmeier, Wolfram [1 ]
Bader, Rainer [1 ]
机构
[1] Univ Rostock, Dept Orthopaed, Biomech & Implant Technol Res Lab, D-18057 Rostock, Germany
来源
KNEE | 2012年 / 19卷 / 06期
关键词
Total knee replacement; Femoral resection; Cutting error; Finite-element-analysis; Ceramics; FINITE-ELEMENT-ANALYSIS; CUTTING ERRORS; ARTHROPLASTY; BONE; FEMUR;
D O I
10.1016/j.knee.2012.03.014
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: A certain failure mode using a newly developed cemented ceramic femoral component in total knee replacement was observed in clinical. application, i.e. fracture of the femoral component during intraoperative impaction. This may be caused by unintentional deflection of the saw blades during cutting with consecutive higher resection angle of the distal femur than desired, leading to bending of the femoral component during implantation. A finite-element-analysis was carried out to simulate implantation of the femoral component and to evaluate the influence of distal femur preparation on implant stress. Scope: We developed and validated a numerical model of the ceramic femoral component including a contact formulation which allowed calculating the principal stresses of the implant during implantation onto the resected femur. The analysis considered different anterior and posterior resection angles with a total of 17 variations. By increasing the femoral resection angle in the finite-element-model it could be shown that a deviation of three degrees from the intended resection angle can cause critical stress amounts during implantation. Conclusions: When implanting the ceramic component in total knee arthroplasty, the femoral resection angles should be prepared very precisely, in particular anterior saw blade deflection has to be avoided. The implant manufacturer increased implant safety through an additional resection template. Moreover, the impaction of the ceramic femoral component during cementing was not further recommended by using a hammer. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:846 / 850
页数:5
相关论文
共 50 条
  • [1] A Comparison of Variable Angle Versus Fixed Angle Distal Femoral Resection in Primary Total Knee Arthroplasty
    Deakin, Angela H.
    Sarungi, Martin
    JOURNAL OF ARTHROPLASTY, 2014, 29 (06): : 1133 - 1137
  • [2] Distal Femoral Valgus Resection Angle in Conventional Total Knee Arthroplasty-a CT Scanogram Study
    Basanagoudar, Praveen L.
    Satishkumar, Bhava R. J.
    Pattabiraman, Kirubakaran
    Kamath, Dinesh
    Ranganadham, A., V
    ARCHIVES OF BONE AND JOINT SURGERY-ABJS, 2023, 11 (03): : 181 - 188
  • [3] Comparison of postoperative alignment using fixed angle versus variable angle distal femoral resection in uncomplicated total knee replacement
    Kehinde Adesola Alatishe
    Wakeel Olaide Lawal
    Michael Ebiyon Ugbeye
    Ranti Oladimeji Babalola
    European Journal of Orthopaedic Surgery & Traumatology, 2023, 33 : 593 - 600
  • [4] Comparison of postoperative alignment using fixed angle versus variable angle distal femoral resection in uncomplicated total knee replacement
    Alatishe, Kehinde Adesola
    Lawal, Wakeel Olaide
    UgbeyenAff, Michael Ebiyon
    Babalola, Ranti Oladimeji
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2023, 33 (03): : 593 - 600
  • [5] SEGMENTAL TOTAL KNEE ARTHROPLASTY AFTER DISTAL FEMORAL RESECTION FOR TUMOR
    SHIH, LY
    SIM, FH
    PRITCHARD, DJ
    ROCK, MG
    CHAO, EYS
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1993, (292) : 269 - 281
  • [6] Computerized tomography scout view for determining distal femoral resection angle in intramedullary instrumentation of total knee arthroplasty
    M. Murad Uslu
    Baris Ozsar
    Meric Cirpar
    Simay Kara
    Fatih Eksioglu
    Ozgur Cetik
    Knee Surgery, Sports Traumatology, Arthroscopy, 2007, 15 : 78 - 82
  • [7] Computerized tomography scout view for determining distal femoral resection angle in intramedullary instrumentation of total knee arthroplasty
    Uslu, M. Murad
    Ozsar, Baris
    Cirpar, Meric
    Kara, Simay
    Eksioglu, Fatih
    Cetik, Ozgur
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2007, 15 (01) : 78 - 82
  • [8] Intraoperative impaction of total knee replacements: An explicit finite-element-analysis of principal stresses in ceramic vs. cobalt-chromium femoral components
    Kluess, Daniel
    Mittelmeier, Wolfram
    Bader, Rainer
    CLINICAL BIOMECHANICS, 2010, 25 (10) : 1018 - 1024
  • [9] Influence of soft tissue balancing and distal femoral resection on flexion contracture in navigated total knee arthroplasty
    Seong Hwan Kim
    Jung-Won Lim
    Ho-Joong Jung
    Han-Jun Lee
    Knee Surgery, Sports Traumatology, Arthroscopy, 2017, 25 : 3501 - 3507
  • [10] Influence of soft tissue balancing and distal femoral resection on flexion contracture in navigated total knee arthroplasty
    Kim, Seong Hwan
    Lim, Jung-Won
    Jung, Ho-Joong
    Lee, Han-Jun
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2017, 25 (11) : 3501 - 3507