The influence of lens power and center thickness on the intraocular pressure measured through soft lenses: A comparison of two noncontact tonometers

被引:6
|
作者
Ogbuehi, Kelechi C. [1 ]
机构
[1] King Saud Univ, Dept Optometry & Vis Sci, Coll Appl Med Sci, Cornea Res Chair, Riyadh 11433, Saudi Arabia
来源
CONTACT LENS & ANTERIOR EYE | 2012年 / 35卷 / 03期
关键词
Daily disposable soft contact lenses; Intraocular pressure; Noncontact tonometer; CT80; PT100; CENTRAL CORNEAL THICKNESS; APPLANATION TONOMETRY; GOLDMANN; RELIABILITY; AGREEMENT; ACCURACY;
D O I
10.1016/j.clae.2012.01.002
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To quantify the influence of soft contact lens power and thickness on the intraocular pressure (IOP). Methods: Thirty-nine young, healthy adult volunteers completed this study. One eye of each subject was randomly assigned either a +6D or a -6D high water content daily disposable lens. The other eye was fitted with the second lens. Triplicate measurements of IOP were taken before, during, and after contact lens wear. Each time, IOP was assessed in a randomized order with two noncontact tonometers. The lenses were swapped between eyes during a second session of measurements, one week later. Results: In the first session with the +6D lenses, the average IOPs (+/- SDs) before, with the lenses fitted, and after the lenses were removed, were: 14.3 +/- 2.9 mmHg, 17.0 +/- 3.3 mmHg and 13.9 +/- 3.1 mmHg, respectively, for the CT80 and 13.6 +/- 3.1 mmHg, 17.1 +/- 4.5 mmHg and 13.3 +/- 2.9 mmHg, respectively, for the PT100. The corresponding values for the first session with the -6D lenses were: 14.3 +/- 3.1 mmHg, 13.1 +/- 3.1 mmHg and 14.1 +/- 3.3 mmHg, respectively, for the CT80 and 13.6 +/- 3.2 mmHg, 13.0 +/- 3.0 mmHg and 13.6 +/- 3.2 mmHg, respectively, for the PT100. IOP significantly (P<0.05) increased (+Delta IOP= 2.7 +/- 0.4 mmHg with the CT80 in the first session) with the +6D lenses, but decreased (P<0.05) when the -6D lenses were fitted (-Delta IOP = 0.6 +/- 0.2 mmHg with the PT100 in the first session). The soft contact lens-induced changes were consistent between sessions but varied between tonometers. Conclusions: The measurement of IOP through soft contact lenses resulted in consistent, statistically significant differences in IOP, which were not uniform across tonometers and which did not appear to be solely related to the central thickness of the soft contact lenses. (C) 2012 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:118 / 128
页数:11
相关论文
共 10 条
  • [1] The Influence of Central Corneal Thickness and Corneal Curvature Radius on The Intraocular Pressure as Measured By Different Tonometers: Noncontact and Goldmann Applanation Tonometers
    Harada, Yukinori
    Hirose, Naofumi
    Kubota, Toshiaki
    Tawara, Akihiko
    JOURNAL OF GLAUCOMA, 2008, 17 (08) : 619 - 625
  • [2] Intraocular Pressure Measurement With the Noncontact Tonometer Through Soft Contact Lenses
    Liu, Yi-Chun
    Huang, Jehn-Yu
    Wang, I-Jong
    Hu, Fung-Rong
    Hou, Yu-Chih
    JOURNAL OF GLAUCOMA, 2011, 20 (03) : 179 - 182
  • [3] Comparison of intraocular pressure adjusted by central corneal thickness or corneal biomechanical properties as measured in glaucomatous eyes using noncontact tonometers and the Goldmann applanation tonometer
    Yaoeda, Kiyoshi
    Fukushima, Atsushi
    Shirakashi, Motohiro
    Fukuchi, Takeo
    CLINICAL OPHTHALMOLOGY, 2016, 10 : 829 - 834
  • [4] Accuracy of intraocular pressure measurements in dogs using two different tonometers and plano therapeutic soft contact lenses
    Ahn, Jeong-Taek
    Jeong, Man-Bok
    Park, Young-Woo
    Kim, Se-Eun
    Ahn, Jae-Sang
    Lee, Yes-Ran
    Lee, Eui-Ri
    Seo, Kangmoon
    VETERINARY OPHTHALMOLOGY, 2012, 15 : 83 - 88
  • [5] The effects of the modulus of the lens material on intraocular pressure measurement through soft contact lenses
    Boyraz, S.
    Gungor, I.
    IRISH JOURNAL OF MEDICAL SCIENCE, 2013, 182 (03) : 331 - 335
  • [6] The effects of the modulus of the lens material on intraocular pressure measurement through soft contact lenses
    S. Boyraz
    İ. Güngör
    Irish Journal of Medical Science, 2013, 182 : 331 - 335
  • [7] Comparison of two handheld applanation tonometers and the association of central corneal thickness, age, and intraocular pressure in normal and diseased canine eyes
    Kato, Kumiko
    VETERINARY OPHTHALMOLOGY, 2014, 17 (06) : 417 - 425
  • [8] Intraocular Pressure Measurements in Relation to Head Position and Through Soft Contact Lenses: Comparison of Three Portable Instruments
    Klein, Ainat
    Shemesh, Gabi
    Loewenstein, Anat
    Kurtz, Shimon
    OPHTHALMIC SURGERY LASERS & IMAGING, 2011, 42 (01) : 64 - 71
  • [9] The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, noncontact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry
    Tonnu, PA
    Ho, T
    Newson, T
    El Sheikh, A
    Sharma, K
    White, E
    Bunce, C
    Garway-Heath, D
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2005, 89 (07) : 851 - 854
  • [10] Comparison of central corneal thickness and intraocular pressure measured with two different tono/pachymeter devices in non-glaucomatouse children
    Teberik, Kuddusi
    Eski, Mehmet Tahir
    Ankarali, Handan
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2021, 31 (06) : 3010 - 3017