CMOS absorbance detection system for capillary electrophoresis

被引:8
|
作者
Pittet, P
Lu, GN
Bertoncini, N
Renaud, L
机构
[1] Univ Lyon 1, Elect Lab, EA 3730, F-69622 Villeurbanne, France
[2] Univ Lyon 1, CNRS, Sci Analyt Lab, UMR 5180, F-69622 Villeurbanne, France
关键词
capillary electrophoresis; absorptiometry; CMOS BDJ detector; signal processing;
D O I
10.1016/j.msec.2005.10.048
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper presents a cost-effective portable photodetection system for capillary electrophoresis absorptiometry. By using a CMOS BDJ (buried double p-n junction) detector, a dual-wavelength method for absorbance measurement is implemented. This system includes associated electronics for low-noise pre-amplification and A/D conversion, followed by digital signal acquisition and processing. Two signal processing approaches are adopted to enhance the signal to noise ratio. One is variable time synchronous detection, which optimizes the sensitivity and measuring rate compared to a conventional synchronous detection technique. The other is a statistical approach based on principal component analysis, which allows optimal estimation of detected signal. This system has been designed and tested in capillary electrophoresis conditions. Its operation has been verified with performances comparable to those of a commercialized spectrophotometric system (HP-3D CE). With potential on-chip integration of associated electronics, it may be operated as an integrable detection module for microchip electrophoresis and other microanalysis systems. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:282 / 289
页数:8
相关论文
共 50 条