Evaluation of a Novel Tubular Sample Holder for Dielectric Measurements with Abdulnour's Method

被引:0
|
作者
Rotava, Andre [1 ]
Borges, Luiz E. P. [2 ]
Dias, Mauricio H. C. [1 ]
Santos, Jose C. A. [1 ]
机构
[1] Mil Inst Engn IME, Dept Elect Engn, Rio De Janeiro, Brazil
[2] Mil Inst Engn IME, Dept Chem Engn, Rio De Janeiro, Brazil
关键词
permittivity measurements; high temperature measurements; transmission line method; Abdulnour's method; RECTANGULAR WAVE-GUIDE; PERMITTIVITY;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper assesses the accuracy of a rectangular waveguide pierced by a tubular load sample holder for measuring liquid materials at high temperatures with a formulation based on Abdulnour's method. The measurement technique is derived from the transmission line method. The sample holder is adapted from a section of a rectangular waveguide pierced by a tubular load for high temperature measurements at 2.45 GHz. Simulations of the loaded sample holder are carried out with CST Microwave Studio (R) for several values of sample permittivity. The accuracy of the technique is evaluated by comparing these values to those obtained from Abdulnour's extraction model. It is shown that by properly choosing a set of calibration standards the technique can be quite accurate.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] SEALABLE SAMPLE HOLDER FOR DIELECTRIC MEASUREMENTS ON HYGROSCOPIC SOLIDS
    NANASSY, AJ
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1974, 45 (11): : 1459 - 1460
  • [2] SAMPLE HOLDER FOR DIELECTRIC MEASUREMENTS UP TO 700 DEGREES C
    HRIZO, J
    SUBBARAO, EC
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1963, 34 (01): : 8 - &
  • [3] IMPROVED SAMPLE HOLDER FOR Q-METER DIELECTRIC MEASUREMENTS
    NELSON, SO
    TRANSACTIONS OF THE ASAE, 1979, 22 (04): : 950 - 954
  • [4] Shielded open-circuited sample holder for dielectric measurements of solids and liquids
    Baker-Jarvis, J
    Janezic, MD
    Jones, CA
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1998, 47 (02) : 338 - 344
  • [5] PREFORMED DIELECTRIC SAMPLE HOLDER FOR 3 CENTIMETER WAVE-GUIDE MEASUREMENTS
    REYNOLDS, SI
    FIELDING, J
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1954, 25 (10): : 1035 - 1035
  • [6] TRANSFORMATION METHOD FOR EVALUATION OF DIELECTRIC MEASUREMENTS
    SCHUFMANN, G
    GUNSSER, W
    ZEITSCHRIFT FUR NATURFORSCHUNG SECTION A-A JOURNAL OF PHYSICAL SCIENCES, 1977, 32 (09): : 1059 - 1060
  • [7] One-Port Coaxial Line Sample Holder Characterisation Method of Dielectric Spectra
    Farhat, Iman
    Farrugia, Lourdes
    Bonello, Julian
    Grima, Rafel
    Persico, Raffaele
    Sammut, Charles
    SENSORS, 2024, 24 (17)
  • [8] Flow-through coaxial sample holder design for dielectric properties measurements from 1 to 350 MHz
    Lawrence, KC
    Nelson, SO
    Bartley, PG
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1998, 47 (02) : 354 - 361
  • [9] Non-contact method to reduce contact problems between sample and electrode in dielectric measurements
    Hao, Jing
    Xu, Xiangdong
    Taylor, Nathaniel
    HIGH VOLTAGE, 2020, 5 (06): : 753 - 761
  • [10] AIR GAP EFFECT AND THE IMPORTANCE OF THE ELECTRODES ON THE SAMPLE IN THE LUMPED-CAPACITANCE METHOD FOR DIELECTRIC MEASUREMENTS
    LIAN, AQ
    ZHONG, WL
    CONFERENCE RECORD OF THE 1989 IEEE INDUSTRY APPLICATIONS SOCIETY ANNUAL MEETING, PTS 1-2, 1989, : 2034 - 2037