Assessing the Thrombogenic Potential of Heart Valve Prostheses: An Approach for a Standardized In-Vitro Method

被引:7
|
作者
Linde, Torsten [1 ]
Clauser, Johanna [1 ]
Meuris, Bart [3 ]
Steinseifer, Ulrich [1 ,2 ]
机构
[1] Rhein Westfal TH Aachen, Helmholtz Inst Appl Med Engn, Dept Cardiovasc Engn, Aachen, Germany
[2] Monash Univ, Monash Inst Med Engn, Dept Mech & Aerosp Engn, Fac Engn, Melbourne, Vic, Australia
[3] Katholieke Univ Leuven, Univ Hosp Leuven, Dept Cardiovasc Sci, Cardiac Surg, Leuven, Belgium
关键词
Heart valve prostheses; Thrombosis; In vitro testing; Tester validation; PLATELET-ADHESION; MEDTRONIC PARALLEL(TM); PROTAMINE SULFATE; AORTIC ROOT; BLOOD-FLOW; SUBENDOTHELIUM; COMPLICATIONS; FIBRINOLYSIS; THROMBOSIS; HEPARIN;
D O I
10.1007/s13239-019-00408-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
PurposeThrombogenic complications are still a main issue in the general performance of cardiovascular implants, especially heart valves. To date, the thrombogenic potential of those prostheses is pre-clinically assessed in time consuming animal studies with questionable evidence.MethodsIn this study, we present a new in-vitro method to assess and compare deficiencies of heart valve substitutes concerning their thrombogenic performance and locate initial clot formation under physiological conditions using porcine blood. Therefore, an athrombogenic pulse duplicator (THIA3) was developed that simulates the anatomic and hemodynamic conditions in the vicinity of the aortic valve. Validation of this tester was carried out with regard to hemodynamics, reproducibility and using positive and negative control valves and by comparison of clot locations with literature data from chronic animal trials with sheep using a St. Jude bileaflet valve.ResultsValidation of the tester showed quasi-physiological hemodynamics and reproducible clot formation. Identical clot formations were found comparing findings in vitro with chronic animal trials.ConclusionThe THIA 3 has proven its suitability for valid, reproducible evaluation of thrombogenic potential of heart valves in a short period of time.
引用
收藏
页码:216 / 224
页数:9
相关论文
共 50 条
  • [1] Assessing the Thrombogenic Potential of Heart Valve Prostheses: An Approach for a Standardized In-Vitro Method
    Torsten Linde
    Johanna Clauser
    Bart Meuris
    Ulrich Steinseifer
    Cardiovascular Engineering and Technology, 2019, 10 : 216 - 224
  • [2] In vitro Analysis of the Thrombogenic Potential of Heart Valve Prostheses in Blood Testing
    Linde, Torsten
    Mager, Ilona
    Schmitz-Rode, Thomas
    Steinseifer, Ulrich
    CARDIOLOGY, 2016, 134 (02) : 236 - 236
  • [3] Thrombogenic Evaluation of Two Mechanical Heart Valve Prostheses Using a New In-Vitro Test System
    Kim, Chong-Hyon
    Steinseifer, Ulrich
    Schmitz-Rode, Thomas
    JOURNAL OF HEART VALVE DISEASE, 2009, 18 (02): : 207 - 213
  • [4] A Low-Volume Tester for the Thrombogenic Potential of Mechanical Heart Valve Prostheses
    Linde, Torsten
    Hamilton, Kathrin F.
    Timms, Daniel L.
    Schmitz-Rode, Thomas
    Steinseifer, Ulrich
    JOURNAL OF HEART VALVE DISEASE, 2011, 20 (05): : 510 - 517
  • [5] On the hemolytic and thrombogenic potential of occluder prosthetic heart valves from in-vitro measurements
    Figliola, R.S.
    Mueller, T.J.
    Journal of Biomechanical Engineering, 1981, 103 (02) : 83 - 90
  • [6] In-Vitro Assessment Of The Influence Of Oval Annuli On The Hydrodynamic Function Of Percutaneous Heart Valve Prostheses
    Kuetting, Maximilian
    Sedaghat, Alexander
    Utzenrath, Marc
    Schmitz, Christoph
    Werner, Nikos
    Steinseifer, Ulrich
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (17) : B239 - B239
  • [7] MarioHeart: Novel In-Vitro Flow Model for Testing Heart Valve Prostheses and Anticoagulant Therapies
    Devos, Maxime
    Liesdek, Omayra C. D.
    Suyker, Willem J. L.
    van Tuijl, Sjoerd
    Schutgens, Roger E. G.
    van de Vosse, Frans N.
    de Heer, Linda M.
    Rutten, Marcel C. M.
    ASAIO JOURNAL, 2023, 69 (05) : E192 - E198
  • [8] In-vitro measurements of the regurgitation of mechanical mitral heart valve prostheses in case of atrial fibrillation
    Mouret, F
    Garitey, V
    Fuseri, J
    Rieu, R
    JOURNAL OF HEART VALVE DISEASE, 2001, 10 (02): : 264 - 268
  • [9] Aortic root compliance influences hemolysis in mechanical heart valve prostheses: an in-vitro study
    Linde, Torsten
    Hamilton, Kathrin F.
    Navalon, Elena Cuenca
    Schmitz-Rode, Thomas
    Steinseifer, Ulrich
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2012, 35 (07): : 495 - 502
  • [10] The Challenge of Assessing Heart Valve Prostheses by Doppler Echocardiography
    Baumgartner, Helmut
    JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2009, 22 (04) : 394 - 395