Effects of a powered ankle-foot prosthesis on kinetic loading of the contralateral limb: A case series

被引:0
|
作者
Hill, David [1 ]
Herr, Hugh [1 ]
机构
[1] MIT, Media Lab, Cambridge, MA 02139 USA
关键词
amputee; ankle; biomechanics; external adduction moment; gait; loading rate; pressure; prosthesis; resultant force; transtibial; TO-STEP TRANSITIONS; TRANSTIBIAL PROSTHESIS; WALKING; AMPUTEE; GAIT; OSTEOARTHRITIS; PATTERNS; KNEE;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Lower-extremity amputees encounter a series of stress-related challenges. Among them is an increased risk of chronic joint disorders. For unilateral, transtibial amputees, we hypothesize that increasing the power output of the trailing, ankle-foot prosthesis during powered plantar flexion could mitigate kinetic loading applied to the leading, contralateral leg during walking. Here, we present a case series that analyzes kinetic factors of unilateral, transtibial amputee gait and forms a comparison between two types of ankle prostheses with varying power outputs. The factors examined here are impact resultant force, peak foot pressure at heel-strike, step-to-step transition work, and knee external adduction moment. The two prostheses are the amputee participant's daily-use passive ankle-foot prosthesis and the BiOM powered ankle-foot prosthesis capable of biologically accurate powered plantar flexion during late stance. In a preliminary study on two transtibial amputees walking over level terrain at a controlled speed (1.25 m/s), we observed average reductions of 8% in peak impact resultant force, 18% in impact resultant force loading rate, 8% in peak heel-strike foot pressure, and 15% in the 1st peak knee external adduction moment when the powered ankle-foot prosthesis was compared to the conventional passive prosthesis. Overall, our preliminary results suggest that more biomimetic prosthetic ankle-foot push-off during late stance may limit leading-leg musculoskeletal stress in walking.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Powered ankle-foot prosthesis
    Biomechatronics Group, MIT Media Lab., Cambridge, MA 02139, United States
    IEEE Rob Autom Mag, 2008, 3 (52-59):
  • [2] Effects of a powered ankle-foot prosthesis on kinetic loading of the unaffected leg during level-ground walking
    Grabowski, Alena M.
    D'Andrea, Susan
    JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2013, 10
  • [3] Effects of a powered ankle-foot prosthesis on kinetic loading of the unaffected leg during level-ground walking
    Alena M Grabowski
    Susan D’Andrea
    Journal of NeuroEngineering and Rehabilitation, 10
  • [4] Powered ankle-foot prosthesis - The importance of series and parallel motor elasticity
    Au, Samuel K.
    Herr, Hugh M.
    IEEE ROBOTICS & AUTOMATION MAGAZINE, 2008, 15 (03) : 52 - 59
  • [5] Active control of a powered ankle-foot prosthesis
    Ashwin Needham
    Andrzej Ordys
    Journal of Foot and Ankle Research, 7 (Suppl 1)
  • [6] Biomechanical design of a powered ankle-foot prosthesis
    Au, Samuel K.
    Weber, Jeff
    Herr, Hugh
    2007 IEEE 10TH INTERNATIONAL CONFERENCE ON REHABILITATION ROBOTICS, VOLS 1 AND 2, 2007, : 298 - 303
  • [7] Actuator Design for Robotic Powered an Ankle-Foot Prosthesis
    Ahmed, M. Hannan
    Wahid, Fawaz
    Ali, Ahsan
    Tiwana, Mohsin I.
    Iqbal, Javaid
    Lovell, Nigel H.
    2015 International Symposium on Bioelectronics and Bioinformatics (ISBB), 2015, : 136 - 139
  • [8] Biomechanical Design and Prototyping of a Powered Ankle-Foot Prosthesis
    Alleva, Stefano
    Antonelli, Michele Gabrio
    Beomonte Zobel, Pierluigi
    Durante, Francesco
    MATERIALS, 2020, 13 (24) : 1 - 15
  • [9] Powered Ankle-Foot Prosthesis for the improvement of amputee ambulation
    Au, Samuel K.
    Herr, Hugh
    Weber, Jeff
    Martinez-Villalpando, Ernesto C.
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 3020 - 3026
  • [10] GAIT EMULATOR FOR EVALUATION OF A POWERED ANKLE-FOOT PROSTHESIS
    Ficanha, Evandro
    Dallali, Houman
    Rastgaar, Mo
    PROCEEDINGS OF THE ASME 10TH ANNUAL DYNAMIC SYSTEMS AND CONTROL CONFERENCE, 2017, VOL 1, 2017,