Clinical evaluation of a block sequential regularized expectation maximization reconstruction algorithm in 18F-FDG PET/CT studies

被引:41
|
作者
Sah, Bert-Ram [1 ,2 ,4 ]
Stolzmann, Paul [1 ,3 ,4 ]
Delso, Gaspar [1 ,5 ]
Wollenweber, Scott D. [5 ]
Huellner, Martin [1 ,3 ,4 ]
Hakami, Yahya A. [1 ]
Queiroz, Marcelo A. [1 ]
Barbosa, Felipe de Galiza [1 ,4 ]
von Schulthess, Gustav K. [1 ,4 ]
Pietsch, Carsten [1 ,4 ]
Veit-Haibach, Patrick [1 ,2 ,4 ]
机构
[1] Univ Zurich Hosp, Dept Nucl Med, Raemistr 100, CH-8091 Zurich, Switzerland
[2] Univ Zurich Hosp, Dept Diagnost & Intervent Radiol, Zurich, Switzerland
[3] Univ Zurich Hosp, Dept Neuroradiol, Zurich, Switzerland
[4] Univ Zurich, Zurich, Switzerland
[5] GE Healthcare, PET Clin Sci, Waukesha, WI USA
关键词
image reconstruction; penalized likelihood; PET; reconstruction algorithm; TIME-OF-FLIGHT; PENALIZED LIKELIHOOD RECONSTRUCTION; STANDARDIZED UPTAKE VALUES; ITERATIVE RECONSTRUCTION; IMAGE-RECONSTRUCTION; EMISSION TOMOGRAPHY; ORDERED SUBSETS; LESION DETECTION; IMPACT; PERFORMANCE;
D O I
10.1097/MNM.0000000000000604
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo investigate the clinical performance of a block sequential regularized expectation maximization (BSREM) penalized likelihood reconstruction algorithm in oncologic PET/computed tomography (CT) studies.MethodsA total of 410 reconstructions of 41 fluorine-18 fluorodeoxyglucose-PET/CT studies of 41 patients with a total of 2010 lesions were analyzed by two experienced nuclear medicine physicians. Images were reconstructed with BSREM (with four different values) or ordered subset expectation maximization (OSEM) algorithm with/without time-of-flight (TOF/non-TOF) corrections. OSEM reconstruction postfiltering was 4.0mm full-width at half-maximum; BSREM did not use postfiltering. Evaluation of general image quality was performed with a five-point scale using maximum intensity projections. Artifacts (category 1), image sharpness (category 2), noise (category 3), and lesion detectability (category 4) were analyzed using a four-point scale. Size and maximum standardized uptake value (SUVmax) of lesions were measured by a third reader not involved in the image evaluation.ResultsBSREM-TOF reconstructions showed the best results in all categories, independent of different body compartments. In all categories, BSREM non-TOF reconstructions were significantly better than OSEM non-TOF reconstructions (P<0.001). In almost all categories, BSREM non-TOF reconstruction was comparable to or better than the OSEM-TOF algorithm (P<0.001 for general image quality, image sharpness, noise, and P=1.0 for artifact). Only in lesion detectability was OSEM-TOF significantly better than BSREM non-TOF (P<0.001). Both BSREM-TOF and BSREM non-TOF showed a decreasing SUVmax with increasing values (P<0.001) and TOF reconstructions showed a significantly higher SUVmax than non-TOF reconstructions (P<0.001).ConclusionThe BSREM reconstruction algorithm showed a relevant improvement compared with OSEM reconstruction in PET/CT studies in all evaluated categories. BSREM might be used in clinical routine in conjunction with TOF to achieve better/higher image quality and lesion detectability or in PET/CT-systems without TOF-capability for enhancement of overall image quality as well.
引用
收藏
页码:57 / 66
页数:10
相关论文
共 50 条
  • [1] Quantitative and Clinical Evaluation of a Block-Sequential Regularized Expectation Maximization Reconstruction Algorithm for 18F-FDG PET-CT in Oncology
    Tragardh, E.
    Minarik, D.
    Almquist, H.
    Bitzen, U.
    Garpered, S.
    Hvittfelt, E.
    Olsson, B.
    Oddstig, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S59 - S60
  • [2] Clinical evaluation of a new block sequential regularized expectation maximization (BSREM) reconstruction algorithm in PET/CT studies
    Sah, Bert-Ram
    Veit-Haibach, Patrick
    Delso, Gaspar
    Wollenweber, Scott
    Tarrade, Xavier
    Licato, Paul
    von Schulthess, Gustav
    Pietsch, Carsten
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [3] Clinical evaluation of a new block sequential regularized expectation maximization (BSREM) reconstruction algorithm in PET/CT studies
    Sah, Bert-Ram
    Veit-Haibach, Patrick
    Delso, Gaspar
    Wollenweber, Scott
    Tarrade, Xavier
    Licato, Paul
    von Schulthess, Gustav
    Pietsch, Carsten
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [4] Impact of acquisition time and penalizing factor in a block-sequential regularized expectation maximization reconstruction algorithm on a Si-photomultiplier-based PET-CT system for 18F-FDG
    Elin Trägårdh
    David Minarik
    Helén Almquist
    Ulrika Bitzén
    Sabine Garpered
    Erland Hvittfelt
    Berit Olsson
    Jenny Oddstig
    EJNMMI Research, 9
  • [5] Impact of acquisition time and penalizing factor in a block-sequential regularized expectation maximization reconstruction algorithm on a Si-photomultiplier-based PET-CT system for 18F-FDG
    Traegardh, Elin
    Minarik, David
    Almquist, Helen
    Bitzen, Ulrika
    Garpered, Sabine
    Hvittfelt, Erland
    Olsson, Berit
    Oddstig, Jenny
    EJNMMI RESEARCH, 2019, 9 (1)
  • [6] Phantom and clinical evaluation of Block Sequential Regularized Expectation Maximization (BSREM) reconstruction algorithm in 68Ga-PSMA PET-CT studies
    Sadeghi, Fatemeh
    Sheikhzadeh, Peyman
    Kasraie, Nima
    Farzanehfar, Saeed
    Abbasi, Mehrshad
    Salehi, Yalda
    Ay, Mohammadreza
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (03) : 1297 - 1308
  • [7] Phantom and clinical evaluation of Block Sequential Regularized Expectation Maximization (BSREM) reconstruction algorithm in 68Ga-PSMA PET-CT studies
    Fatemeh Sadeghi
    Peyman Sheikhzadeh
    Nima Kasraie
    Saeed Farzanehfar
    Mehrshad Abbasi
    Yalda Salehi
    Mohammadreza Ay
    Physical and Engineering Sciences in Medicine, 2023, 46 : 1297 - 1308
  • [8] Clinical evaluation of regularized OSEM reconstruction algorithm in pediatric total-body 18F-FDG PET/CT
    Zhao, Yu-Mo
    Hu, Ying-Ying
    Feng, Jiatai
    Li, Chenwei
    Zhang, Xu
    Fan, Wei
    JOURNAL OF NUCLEAR MEDICINE, 2021, 62
  • [9] Quantitative performance and optimal regularization parameter in block sequential regularized expectation maximization reconstructions in carotid plaques with 18F-FDG PET /MR
    Yang, Hongwei
    Lu, Jie
    Cui, Bixiao
    Zhang, Yue
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [10] Reproducibility of texture features with block sequential regularized expectation maximization (BSREM) reconstruction on FDG PET/MR
    Nogami, Munenobu
    Zeng, Feibi
    Wakabayashi, Yuichi
    Ueno, Yoshiko
    Kanda, Tomonori
    Sofue, Keitaro
    Kurimoto, Takako
    Kubo, Kazuhiro
    Murakami, Takamichi
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59