Mini EXPLORER II: a prototype high-sensitivity PET/CT scanner for companion animal whole body and human brain scanning

被引:31
|
作者
Lv, Yang [1 ]
Lv, Xinyu [1 ]
Liu, Weiping [1 ]
Judenhofer, Martin S. [2 ,3 ]
Zwingenberger, Allison [4 ]
Wisner, Erik [4 ]
Berg, Eric [3 ]
McKenney, Sarah [2 ]
Leung, Edwin [3 ]
Spencer, Benjamin A. [2 ,3 ]
Cherry, Simon R. [2 ,3 ]
Badawi, Ramsey D. [2 ,3 ]
机构
[1] Shanghai United Imaging Healthcare Co Ltd, Mol Imaging Business Unit, Shanghai, Peoples R China
[2] Univ Calif Davis, Sch Med, Dept Radiol, Davis, CA 95616 USA
[3] Univ Calif Davis, Sch Engn, Dept Biomed Engn, Davis, CA 95616 USA
[4] Univ Calif Davis, Dept Surg & Radiol Sci, Sch Vet Med, Davis, CA 95616 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2019年 / 64卷 / 07期
关键词
total-body PET; PET/CT; companion animal imaging; EXPLORER; brain PET; AXIAL-FIELD; SYSTEM; FEASIBILITY;
D O I
10.1088/1361-6560/aafc6c
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
As part of the EXPLORER total-body positron emission tomography (PET) project, we have designed and built a high-resolution, high-sensitivity PET/CT scanner, which is expected to have excellent performance for companion animal whole body and human brain imaging. The PET component has a ring diameter of 52 cm and an axial field of view of 48.3 cm. The detector modules are composed of arrays of lutetium (yttrium) oxyorthosilicate (LYSO) crystals of dimensions 2.76 x 2.76 x 18.1 mm(3) coupled to silicon photomultipliers (SiPMs) for read-out. The CT component is a 24 detector row CT scanner with a 50 kW x-ray tube. PET system time-of-flight resolution was measured to be 409 +/- 39 ps and average system energy resolution was 11.7% +/- 1.5% at 511 keV. The NEMA NU2-2012 system sensitivity was found to be 52-54 kcps MBq(-1). Spatial resolution was 2.6 mm at 10 mm from the center of the FOV and 2.0 mm rods were clearly resolved on a mini-Derenzo phantom. Peak noise-equivalent count (NEC) rate, using the NEMA NU 2-2012 phantom, was measured to be 314 kcps at 9.2 kBq cc(-1). The CT scanner passed the technical components of the American College of Radiology (ACR) accreditation tests. We have also performed scans of a Hoffman brain phantom and we show images from the first canine patient imaged on this device.
引用
收藏
页数:12
相关论文
共 34 条
  • [1] Prism-PET II: a prototype ultra-high resolution and high-sensitivity PET scanner for human brain imaging
    Wang, Zipai
    Zeng, Xinjie
    Tan, Wanbin
    Li, Yixin
    Petersen, Eric
    LaBella, Andy
    Zhang, Furu
    Qi, Jinyi
    Ajay, Gupta
    de Leon, Mony
    Chiang, Gloria
    Goldan, Amirhossein
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [2] Measurement accuracy and practicality of 15O-gas PET using a high-sensitivity whole-body PET/CT scanner
    Omasa, Wataru
    Matsumoto, Keiichi
    Shimizu, Keiji
    Akamatsu, Go
    Ikari, Yasuhiko
    Fukui, Tatsuya
    Sasaki, Masahiro
    Senda, Michio
    Yamane, Tomohiko
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [3] Performance Evaluation of a High-Sensitivity Preclinical PET/CT Scanner
    Cheng, Y.
    Hu, T.
    Li, X.
    Zeng, L.
    Zhang, S.
    Li, A.
    Hua, Y.
    Xie, Q.
    Zhou, X.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S704 - S704
  • [4] Performance characterization of a high-sensitivity small-animal PET scanner
    Kao, Chien-Min
    Xie, Qingguo
    Dong, Yun
    Wan, Lu
    Chen, Chin-Tu
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4176 - +
  • [5] Use of a whole body PET/CT scanner as a combined animal PET/CT imaging device.
    Adler, LP
    Doss, M
    Manjeshwar, RM
    Trotter, DEG
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (05) : 288P - 288P
  • [6] First prototype of a compact helmetchin PET for high-sensitivity brain imaging
    Yamaya, Taiga
    Yoshida, Eiji
    Tashima, Hideaki
    Inadama, Naoko
    Shinaji, Tetsuya
    Wakizaka, Hidekatsu
    Nitta, Munetaka
    Tazawa, Shusaku
    Suhara, Tetsuya
    Fujibayashi, Yasuhisa
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [7] A High-Sensitivity Small-Animal PET Scanner: Development and Initial Performance Measurements
    Kao, Chien-Min
    Xie, Qingguo
    Dong, Yun
    Wan, Lu
    Chen, Chin-Tu
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) : 2678 - 2688
  • [8] A prototype PET/SPET/X-rays scanner dedicated for whole body small animal studies
    Rouchota, Maritina
    Georgiou, Maria
    Fysikopoulos, Eleftherios
    Fragogeorgi, Eirini
    Mikropoulos, Konstantinos
    Papadimitroulas, Panagiotis
    Kagadis, George C.
    Loudos, George
    HELLENIC JOURNAL OF NUCLEAR MEDICINE, 2017, 20 (02): : 146 - 153
  • [9] An Image Quality Evaluation of Short Acquisition on a High-Sensitivity PET-CT Scanner
    Lv, Yang
    Deng, Zilin
    Wang, Fang
    Li, Chenwei
    Dong, Yun
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [10] Design and performance of a prototype whole-body PET/CT scanner with fiber-optic readout
    Worstell, W
    Adler, S
    Domigan, P
    Johnson, O
    Kudrolli, H
    Lazuka, D
    Monteverde, P
    Nevin, J
    Rohatgi, R
    Romanov, L
    Starsja, S
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3280 - 3284