Preliminary Results of a Low-Cost 4-Channel Time-Correlated Single Photon Counting System for Time-Domain Diffuse Optical Tomography

被引:0
|
作者
Bouchard, Jonathan [1 ]
Lemaire, William [1 ]
Samson, Arnaud [1 ]
Paulin, Caroline [1 ]
Pratte, Jean-Francois [1 ,2 ]
Berube-Lauziere, Yves [1 ,2 ]
Fontaine, Rejean [1 ,2 ]
机构
[1] Univ Sherbrooke, Dept Elect & Comp Engn, Sherbrooke, PQ J1K 2R1, Canada
[2] Ctr Hosp Univ Sherbrooke CR CHUS, FRQ Sante Funded Ctr Rech, Sherbrooke, PQ, Canada
关键词
Time-Correlated Single Photon Counting (TC-SPC) electronics; Diffuse Optical Tomography (DOT); Data Acquisition (DAQ) system; Time-to-Digital Converter (TDC); Time-of-Flight (TOF) measurement; Medical Imaging;
D O I
暂无
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Time-domain diffuse optical tomography (TD-DOT) provides information-rich data that have not yet been fully exploited for image reconstruction, notably to increase imaging spatial resolution. Current TD-DOT scanners suffer from a very low sensitivity owing to their small number of detection channels. This leads to excessively long acquisition times for in vivo imaging. To obtain a higher number of detection channels, thus increasing detection density, a low-cost time-correlated single photon counting (TCSPC) system dedicated to TD-DOT imaging was designed and developed, resorting solely to off-the-shelf electronic components to reduce costs, in distinction to custom application-specific integrated circuit (ASIC) solutions. It features 4 input channels with a 13.02 ps bin width and a 18.1 ps FWHM accuracy throughout a measurement dynamic range of 12.5 ns. Each channel includes a leading-edge discriminator, with a programmable threshold, for direct interfacing with off-the-shelf photodetector modules. A software-programmable delay line was added to the channel signal path to compensate for undesired propagation delays. The system also supports a virtually unlimited number of TCSPC channels using a daisy-chain configuration through an onboard Ethernet switch.
引用
收藏
页码:200 / 205
页数:6
相关论文
共 50 条
  • [1] A Low-Cost Time-Correlated Single Photon Counting System for Multiview Time-Domain Diffuse Optical Tomography
    Bouchard, Jonathan
    Samson, Arnaud
    Lemaire, William
    Paulin, Caroline
    Pratte, Jean-Francois
    Berube-Lauziere, Yves
    Fontaine, Rejean
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2017, 66 (10) : 2505 - 2515
  • [2] Prospects on Time-Domain Diffuse Optical Tomography Based on Time-Correlated Single Photon Counting for Small Animal Imaging
    Berube-Lauziere, Yves
    Crotti, Matteo
    Boucher, Simon
    Ettehadi, Seyedrohollah
    Pichette, Julien
    Rech, Ivan
    JOURNAL OF SPECTROSCOPY, 2016, 2016 : 1 - 23
  • [3] A Low-Cost Time-Correlated Single Photon Counting Portable DNA Analyzer
    Tian, Yi
    Wei, Liping
    Ho, Derek
    SENSORS, 2019, 19 (13)
  • [4] Ultra-fast time-correlated single photon counting avalanche photodiodes for time-domain non-contact fluorescence diffuse optical tomography
    Robichaud, Vincent
    Lapointe, Eric
    Berube-Lauziere, Yves
    PHOTONICS NORTH 2007, PTS 1 AND 2, 2007, 6796 : 79606 - 79606
  • [5] Multi-dimensional time-correlated single photon counting applied to diffuse optical tomography
    Becker, W
    Bergmann, A
    Gibson, A
    Everdell, N
    Jennions, D
    Schweige, M
    Arridge, SR
    Hebden, JC
    OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE VI, 2005, 5693 : 34 - 42
  • [6] Time-resolved fluorescence measurements for diffuse optical tomography using ultrafast time-correlated single photon counting
    Berube-Lauziere, Yves
    Robichaud, Vincent
    ADVANCED PHOTON COUNTING TECHNIQUES, 2006, 6372
  • [7] 8-channel acquisition system for time-correlated single-photon counting
    Antonioli, S.
    Miari, L.
    Cuccato, A.
    Crotti, M.
    Rech, I.
    Ghioni, M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (06):
  • [8] Time-of-flight optical ranging system based on time-correlated single-photon counting
    Massa, JS
    Buller, GS
    Walker, AC
    Cova, S
    Umasuthan, M
    Wallace, AM
    APPLIED OPTICS, 1998, 37 (31): : 7298 - 7304
  • [9] Development of a high-performance multi-channel system for Time-Correlated Single Photon counting
    Peronio, P.
    Cominelli, A.
    Acconcia, G.
    Rech, I.
    Ghioni, M.
    PHOTON COUNTING APPLICATIONS 2017, 2017, 10229
  • [10] Complete and Compact 32-Channel System for Time-Correlated Single-Photon Counting Measurements
    Cuccato, A.
    Antonioli, S.
    Crotti, M.
    Labanca, I.
    Gulinatti, A.
    Rech, I.
    Ghioni, M.
    IEEE PHOTONICS JOURNAL, 2013, 5 (05):