Optimising image quality and radiation dose for neonatal incubator imaging

被引:4
|
作者
Tugwell-Allsup, J. [1 ]
Morris, R. W. [1 ]
Hibbs, R. [2 ]
England, A. [3 ]
机构
[1] Besti Cadwaladr Univ Hlth Board, Penrhosgarnedd Rd, Bangor LL57 2PW, Gwynedd, Wales
[2] Integral Business Support Ltd, Wrexham, Wales
[3] Salford Univ, Manchester, Greater Manches, England
关键词
Optimisation; Neonate; Incubator; Visual image quality; Effective dose; Balance; COMFORT PADS; DESIGN; IMPACT;
D O I
10.1016/j.radi.2020.03.011
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Neonates often require imaging within incubators however limited evidence exists as to the optimal method and acquisition parameters to achieve these examinations. This study aims to standardise and optimise neonatal chest radiography within incubators. Methods: A neonatal anthropomorphic phantom was imaged on two different incubators under controlled conditions using a DR system. Exposure factors, SID and placement of image receptor (direct v tray) were explored whilst keeping all other parameters consistent. Image quality was evaluated using absolute visual grading analysis (VGA) with contrast-to-noise ratio (CNR) also calculated for comparison. Effective dose was established using Monte Carlo simulation using entrance surface dose within its calculations. Results: VGA and CNR reduced significantly (p < 0.05) whilst effective dose increased significantly (p < 0.05) for images acquired using the incubator tray. The optimal combinations of parameters for incubator imaging were: image receptor directly behind neonate, 0.5 mAs, 60 kV at 100 cm SID, however, if tray needs to be used then these need to be adapted to: 1 mAs at maximum achievable SID. Effective dose was highest for images acquired using both incubator tray and 100 cm SID owing to a decrease in focus to skin distance. There is significant increase (p < 0.01) in VGA between using 0.5 mAs and 1 mAs but an apparent lack of increase between 1 and 1.5 mAs. Conclusion: Using the incubator tray has an adverse effect on both image quality and radiation dose for incubator imaging. Direct exposure is optimal for this type of examination but if tray needs to be used, both mAs and SID need to be increased slightly to compensate. Implications for practice: This study can help inform practice in order to both standardise and optimise chest imaging for neonates in incubators. Crown Copyright (C) 2020 Published by Elsevier Ltd on behalf of The College of Radiographers. All rights reserved.
引用
收藏
页码:E258 / E263
页数:6
相关论文
共 50 条
  • [1] Neonatal radiography: Radiation dose in association with image quality
    Dougeni, E
    Karatza, A
    Delis, H
    Kalogeropoulou, C
    Panayiotakis, G
    Mantagos, S
    PEDIATRIC RESEARCH, 2005, 58 (02) : 387 - 387
  • [2] Neonatal radiography: radiation dose in association with image quality: 192
    Dougeni E.
    Karatza A.
    Delis H.
    Kalogeropoulou C.
    Panayiotakis G.
    Mantagos S.
    Pediatric Research, 2005, 58 (2) : 387 - 387
  • [3] Radiation dose and image quality optimisation in medical imaging
    Hogg, Peter
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2015, 62 (03) : 236 - 237
  • [4] Radiation dose and image quality in pediatric bitewing imaging
    Rajalingam, Raheis
    Brage, Karen
    Nielsen, Louise Kjaerby
    Eriksen, Freja Bogh
    Jorgensen, Helene Hviid
    Mikkelsen, Anne Sofie
    Scholer, Gitte
    Mekis, Nejc
    Bruvo, Maja
    Precht, Helle
    RADIATION PROTECTION DOSIMETRY, 2024, 201 (02) : 113 - 121
  • [5] Optimisation of radiation dose and image quality in mobile neonatal chest radiography
    Hinojos-Armendariz, V. I.
    Mejia-Rosales, S. J.
    Franco-Cabrera, M. C.
    RADIOGRAPHY, 2018, 24 (02) : 104 - 109
  • [6] Image quality and radiation dose in planar imaging - Image quality figure of merits from the CDRAD phantom
    Konst, Bente
    Weedon-Fekjaer, Harald
    Bath, Magnus
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (07): : 151 - 159
  • [7] Image quality associated with the use of an MR-compatible incubator in neonatal neuroimaging
    O'Regan, K.
    Filan, P.
    Pandit, N.
    Maher, M.
    Fanning, N.
    BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1012): : 363 - 367
  • [8] Reduced radiation dose and improved image quality using a mini mobile digital imaging system in a neonatal intensive care unit
    Kim, Tae-Hoon
    Ryu, Jong-Hyun
    Jeong, Chang-Won
    Jun, Hong Young
    Heo, Dong-Woon
    Lee, Seung Hyun
    Oh, Yeon Kyun
    Lee, Mi-Jung
    Yoon, Kwon-Ha
    CLINICAL IMAGING, 2017, 42 : 165 - 171
  • [9] Dose and image quality optimization in neonatal radiography
    Dougeni, E. D.
    Delis, H. B.
    Karatza, A. A.
    Kalogeropoulou, C. P.
    Skiadopoulos, S. G.
    Mantagos, S. P.
    Panayiotakis, G. S.
    BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (958): : 807 - 815
  • [10] Editorial: Radiation dose and image quality
    Manning, David
    RADIOGRAPHY, 2014, 20 (04) : 291 - 292