Highly selective and sensitive preconcentration of mercury ion and determination by cold vapor atomic absorption spectroscopy

被引:217
|
作者
Ghaedi, M [1 ]
Fathi, MR
Shokrollahi, A
Shajarat, F
机构
[1] Univ Yasuj, Dept Chem, Yasuj 75914353, Iran
[2] Ahvaz Univ, Dept Chem, Ahvaz, Iran
关键词
2-mercaptobenzoxazole; mercury(II) ion; solid phase extraction; cold vapor atomic absorption spectrometry; chromosorb P;
D O I
10.1080/00032710600622167
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and selective method based on a sodium dodecyl sulfate (SDS)-coated chromosorb P modified by 2-mercaptobenzoxazole (MBO) has been developed to selectively separate and concentrate ultra trace amounts of mercury(II) ions for its highly sensitive measurement by cold vapor atomic absorption spectrometry (CVAAS). The mercury ions were adsorbed quantitatively on SDS-coated chromosorb due to its complexation with MBO, while the retained Hg2+ ions were then stripped from the column with minimal amounts of 2 M nitric acid in acetone. The eluting solution was sent to CV-AAS for evaluating Hg2+ ion content and results indicate that the calibration curve was linear for Hg2+ ion in the range of 0.05-85.6 ng mL(-1) and 0.09-9.6 mu g mL(-1) of Hg2+ ions. Maximum capacity of the SDS-coated chromosorb modified with 40 mg of the ligand was found to be 49830 mu g of mercury(II), the limit of detection was 0.01 ng mL(-1) , and enrichment factors were about 300, which make it suitable it for dilute solution analysis. The method was successfully applied to the determination of Hg2+ ion content in real samples.
引用
收藏
页码:1171 / 1185
页数:15