Determination of temporal reproducibility and variability of cancer biomarkers in serum and EDTA plasma samples using a proximity extension assay

被引:4
|
作者
Christensen, Troels D. [1 ]
Maag, Emil [2 ]
Madsen, Kasper [1 ]
Lindgaard, Sidsel Christy [1 ]
Nielsen, Dorte L. [1 ,3 ]
Johansen, Julia S. [1 ,3 ,4 ]
机构
[1] Copenhagen Univ Hosp Herlev & Gentofte, Dept Oncol, Borgmester Ib Juuls Vej 1, DK-2730 Herlev, Denmark
[2] BioXpedia, Aarhus, Denmark
[3] Univ Copenhagen, Fac Hlth & Med Sci, Dept Clin Med, Copenhagen, Denmark
[4] Copenhagen Univ Hosp Herlev & Gentofte, Dept Med, Herlev, Denmark
关键词
Olink immuno-oncology; Proximity extension assay; Multiplex Immunoassays; Proteins; Tumor markers; Serum-plasma variability;
D O I
10.1186/s12014-022-09380-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background Proximity extension assay (PEA) is a novel antibody-based proteomic technology. Sparse data have been published concerning the matrix effect of serum vs. ethylenediamine tetraacetic acid (EDTA) plasma and the reproducibility of results obtained using PEA technology. Methods We analyzed samples with the PEA-based 92-plex Olink (R) immuno-oncology (I-O) assay. To estimate the matrix effect, we analyzed paired serum and EDTA plasma samples from 12 patients with biliary tract cancer. To evaluate the reproducibility, we used data from 7 studies, where 6-8 serum samples from patients with pancreatic cancer were used as bridging samples on 3 versions of the panel over a 2.5-years period. Results For the study of serum vs. plasma, 80 proteins were evaluable. The mean serum to EDTA plasma ratio ranged from 0.41-3.01. For 36 proteins, the serum and plasma values were not comparable due to high variability of the ratio, poor correlation, or possible concentration effect. For the bridging samples, the mean intra-study inter-assay coefficient of variation (CV) ranged from 11.3% to 26.1%. The mean inter-study CV was 42.0% before normalization and 26.2% after normalization. Inter-study results were well correlated (r >= 0.93), especially for studies using the same version of the panel (r >= 0.99). Conclusion For 44 of 92 proteins included in the Olink (R) I-O panel, the variation between results obtained using serum and EDTA plasma was constant and results were well correlated. Furthermore, samples could be stored for several years and used on different versions of the same PEA panel without it effecting results.
引用
收藏
页数:17
相关论文
共 50 条
  • [1] Determination of temporal reproducibility and variability of cancer biomarkers in serum and EDTA plasma samples using a proximity extension assay
    Troels D. Christensen
    Emil Maag
    Kasper Madsen
    Sidsel Christy Lindgaard
    Dorte L. Nielsen
    Julia S. Johansen
    Clinical Proteomics, 2022, 19
  • [2] Identification of Candidate Plasma Protein Biomarkers for Cervical Cancer Using the Multiplex Proximity Extension Assay
    Berggrund, Malin
    Enroth, Stefan
    Lundberg, Martin
    Assarsson, Erika
    Stalberg, Karin
    Lindquist, David
    Hallmans, Goran
    Grankvist, Kjeii
    Olovsson, Matts
    Gyllensten, Ulf
    MOLECULAR & CELLULAR PROTEOMICS, 2019, 18 (04) : 735 - 743
  • [3] Discovery of novel glioma serum biomarkers by proximity extension assay
    Ghorbani, Atefeh
    Avery, Lisa M.
    Sohaei, Dorsa
    Soosaipillai, Andrea
    Richer, Maxime
    Horbinski, Craig
    McCortney, Katy
    Xu, Wei
    Diamandis, Eleftherios P.
    Prassas, Ioannis
    CLINICAL PROTEOMICS, 2023, 20 (01)
  • [4] Discovery of novel glioma serum biomarkers by proximity extension assay
    Atefeh Ghorbani
    Lisa M. Avery
    Dorsa Sohaei
    Andrea Soosaipillai
    Maxime Richer
    Craig Horbinski
    Katy McCortney
    Wei Xu
    Eleftherios P. Diamandis
    Ioannis Prassas
    Clinical Proteomics, 2023, 20
  • [5] PHOTOCOLORIMETRIC DETERMINATION OF CALCIUM LEVEL IN PLASMA OR SERUM USING EDTA AND MUREXID
    SELOCHNIK, LI
    BRISKIN, AI
    ANTONOVA, EE
    KHIMIKO-FARMATSEVTICHESKII ZHURNAL, 1978, 12 (10): : 138 - 140
  • [6] Procalcitonin detection in human plasma specimens using a fast version of proximity extension assay
    Bedin, Frederic
    Benoit, Vincent
    Ferrazzi, Elsa
    Aufradet, Emeline
    Boulet, Laurent
    Rubens, Agnes
    Dalbon, Pascal
    Imbaud, Pierre
    PLOS ONE, 2023, 18 (02):
  • [7] Detection of biomarkers using a novel proximity extension assay in patients with ST-elevation myocardial infarction
    Kulasingam, Archana
    Hvas, Anne-Mette
    Grove, Erik Lerkevang
    Funck, Kristian Lokke
    Kristensen, Steen Dalby
    THROMBOSIS RESEARCH, 2018, 172 : 21 - 28
  • [8] Proximity Extension Assay Technology Identifies Novel Serum Biomarkers for Predicting Inflammatory Bowel Disease: IBD Character Consortium
    Kalla, Rahul
    Kennedy, Nicholas A.
    Hjelm, Fredrik N.
    Modig, Eddie
    Sundell, Mikael
    Soederholm, Johan
    Andreassen, Bettina K.
    Bergemalm, Daniel
    Ventham, Nicholas T.
    Hjortswang, Henrik
    Ricanek, Petr
    Vatn, Morten H.
    Halfvarson, Jonas
    Gullberg, Mats
    Character, I. B. D.
    Satsangi, Jack
    GASTROENTEROLOGY, 2015, 148 (04) : S436 - S436
  • [9] PROXIMITY EXTENSION ASSAY TECHNOLOGY IDENTIFIES NOVEL SERUM BIOMARKERS FOR PREDICTING INFLAMMATORY BOWEL DISEASE: IBD CHARACTER CONSORTIUM
    Kalla, R.
    Kennedy, N. A.
    Hjelm, F.
    Modig, E.
    Sundell, M.
    Soderholm, J.
    Andreassen, B.
    Bergemalm, D.
    Ventham, N. T.
    Hjortswang, H.
    Petr, R.
    Vatn, M. H.
    Halfvarson, J.
    Gullberg, M.
    Satsangi, J.
    GUT, 2015, 64 : A8 - A8
  • [10] Proximity Extension Assay technology identifies novel serum biomarkers for predicting Inflammatory Bowel Disease: IBD Character Consortium
    Kalla, R.
    Kennedy, N.
    Hjelm, F.
    Modig, E.
    Sundell, M.
    Andreassen, B.
    Bergemalm, D.
    Ricanek, P.
    Soderholm, J.
    Vatn, M.
    Halfvarson, J.
    Gullberg, M.
    Satsangi, J.
    JOURNAL OF CROHNS & COLITIS, 2015, 9 : S146 - S147