Assessment of the Speech Intelligibility Performance of Post Lingual Cochlear Implant Users at Different Signal-to-Noise Ratios Using the Turkish Matrix Test

被引:3
|
作者
Polat, Zahra [1 ]
Bulut, Erdogan [2 ]
Atas, Ahmet [3 ]
机构
[1] Istanbul Univ, Sch Med, Dept Audiol, Istanbul, Turkey
[2] Trakya Univ, Sch Med, Dept Audiol, Istanbul, Turkey
[3] Cerrahpasa Med Sch Hosp, Dept Otorhinolaryngol, Istanbul, Turkey
关键词
Turkish Matrix Test; noise; speech intelligibility; cochlear implant; SENTENCE TEST;
D O I
10.5152/balkanmedj.2016.160180
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Spoken word recognition and speech perception tests in quiet are being used as a routine in assessment of the benefit which children and adult cochlear implant users receive from their devices. Cochlear implant users generally demonstrate high level performances in these test materials as they are able to achieve high level speech perception ability in quiet situations. Although these test materials provide valuable information regarding Cochlear Implant (CI) users' performances in optimal listening conditions, they do not give realistic information regarding performances in adverse listening conditions, which is the case in the everyday environment. Aims: The aim of this study was to assess the speech intelligibility performance of post lingual CI users in the presence of noise at different signal-to-noise ratio with the Matrix Test developed for Turkish language. Study Design: Cross-sectional study. Methods: The thirty post lingual implant user adult subjects, who had been using implants for a minimum of one year, were evaluated with Turkish Matrix test. Subjects' speech intelligibility was measured using the adaptive and non-adaptive Matrix Test in quiet and noisy environments. Results: The results of the study show a correlation between Pure Tone Average (PTA) values of the subjects and Matrix test Speech Reception Threshold (SRT) values in the quiet. Hence, it is possible to asses PTA values of CI users using the Matrix Test also. However, no correlations were found between Matrix SRT values in the quiet and Matrix SRT values in noise. Similarly, the correlation between PTA values and intelligibility scores in noise was also not significant. Therefore, it may not be possible to assess the intelligibility performance of CI users using test batteries performed in quiet conditions. Conclusion: The Matrix Test can be used to assess the benefit of CI users from their systems in everyday life, since it is possible to perform intelligibility test with the Matrix test using a material that CI users experience in their everyday life and it is possible to assess their difficulty in speech discrimination in noisy conditions they have to cope with.
引用
收藏
页码:532 / 538
页数:7
相关论文
共 19 条
  • [1] Comparing the Speech Perception of Cochlear Implant Users with Three Different Finnish Speech Intelligibility Tests in Noise
    Willberg, Tytti
    Sivonen, Ville
    Linder, Pia
    Dietz, Aarno
    JOURNAL OF CLINICAL MEDICINE, 2021, 10 (16)
  • [2] Chinese speech intelligibility at different speech sound pressure levels and signal-to-noise ratios in simulated classrooms
    Peng Jianxin
    APPLIED ACOUSTICS, 2010, 71 (04) : 386 - 390
  • [3] Music perception and speech intelligibility in noise performance by Italian-speaking cochlear implant users
    Hilal Dincer D’Alessandro
    Patrick J. Boyle
    Ginevra Portanova
    Patrizia Mancini
    European Archives of Oto-Rhino-Laryngology, 2022, 279 : 3821 - 3829
  • [4] Music perception and speech intelligibility in noise performance by Italian-speaking cochlear implant users
    Dincer D'Alessandro, Hilal
    Boyle, Patrick J.
    Portanova, Ginevra
    Mancini, Patrizia
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2022, 279 (08) : 3821 - 3829
  • [5] Subjective Assessment of Cochlear Implant Users' Signal-to-Noise Ratio Requirements for Different Levels of Wireless Device Usability
    Julstrom, Stephen
    Kozma-Spytek, Linda
    JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2014, 25 (10) : 952 - 968
  • [6] Speech intelligibility and recall of first and second language words heard at different signal-to-noise ratios
    Hygge, Staffan
    Kjellberg, Anders
    Nostl, Anatole
    FRONTIERS IN PSYCHOLOGY, 2015, 6
  • [7] Assessment of Speech Intelligibility in Free-field Sound Chamber at Different Signal Noise Ratios
    Polat, Zahra
    Atas, Ahmet
    JOURNAL OF ACADEMIC RESEARCH IN MEDICINE-JAREM, 2020, 10 (02): : 155 - 159
  • [8] Beamforming and Single-Microphone Noise Reduction: Effects on Signal-to-Noise Ratio and Speech Recognition of Bimodal Cochlear Implant Users
    Stronks, H. Christiaan
    Briaire, Jeroen
    Frijns, Johan
    TRENDS IN HEARING, 2022, 26
  • [9] Beamforming and Single-Microphone Noise Reduction: Effects on Signal-to-Noise Ratio and Speech Recognition of Bimodal Cochlear Implant Users
    Stronks, H. Christiaan
    Briaire, Jeroen J.
    Frijns, Johan H. M.
    TRENDS IN HEARING, 2022, 26
  • [10] Speech Intelligibility and Recall of Spoken Material Heard at Different Signal-to-noise Ratios and the Role Played by Working Memory Capacity
    Ljung, Robert
    Israelsson, Karl
    Hygge, Staffan
    APPLIED COGNITIVE PSYCHOLOGY, 2013, 27 (02) : 198 - 203