Modeling of the flow comparator as calibration device for high pressure natural gas flow metering in Modelica

被引:1
|
作者
Singh, Sukhwinder [1 ]
Schmitz, Gerhard [1 ]
Mickan, Bodo [2 ]
机构
[1] Hamburg Univ Technol, Inst Engn Thermodynam, Hamburg, Germany
[2] Phys Tech Bundesanstalt, Braunschweig, Germany
关键词
Pistons;
D O I
10.1016/j.flowmeasinst.2020.101808
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The German national metrological institute, Physikalisch-Technische Bundesanstalt, is developing a new concept a for a volumetric primary standard to calibrate high pressure gas flow meters. The TUHH is supporting these Research & Development activities with its competence to elaborate computational models for detailed analysis of complex electromechanical systems including fluid flow aspects. The new primary standard is called Flow Comparator and uses an actively driven piston prover to measure the gas flow rate using the time the piston needs to displace a defined enclosed volume of gas in a cylinder. A computational model is developed in Modelica (R) to investigate the Flow Comparator's dynamic behavior and interaction with other components in the loop. The validation of the developed model shows good compliance with measured piston velocity and differential pressure at the piston. The model is used to optimize the frequency inverter's control voltage trajectory to increase the available calibration time.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Using Modelica to investigate the dynamic behaviour of the German national standard for high pressure natural gas flow metering
    von der Heyde, M.
    Schmitz, G.
    Mickan, B.
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2016, 27 (08)
  • [2] Differential flow calibration of turbine flowmeters for air and high pressure natural gas
    Liu Y.
    Li H.
    Wu J.
    Shen Q.
    Quan X.
    1600, Natural Gas Industry Journal Agency (40): : 119 - 124
  • [3] A New Method of Calibration of Nozzles for Flow Metering Gas Installations
    Gutkin, M. B.
    Chistyakov, Yu A.
    MEASUREMENT TECHNIQUES, 2013, 56 (05) : 567 - 569
  • [4] A New Method of Calibration of Nozzles for Flow Metering Gas Installations
    M. B. Gutkin
    Yu. A. Chistyakov
    Measurement Techniques, 2013, 56 : 567 - 569
  • [5] Analysis of high pressure tests on wet gas flow metering with a Venturi meter
    Gajan, P.
    Decaudin, Q.
    Couput, J. P.
    FLOW MEASUREMENT AND INSTRUMENTATION, 2015, 44 : 126 - 131
  • [6] Experimental Evaluation of friction in high speed compressible homenergic flow and development of friction choked gas flow metering device
    Marzouk, El-Sayed M.
    AEJ - Alexandria Engineering Journal, 1995, 34 (02):
  • [7] Gas flow calibration standard device for working site
    Niu, LiNa
    Yang, GuangLei
    Yang, ShiQi
    Zang, XianLe
    Song, YaJing
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2023, 34 (07)
  • [8] Investigation of natural gas hydrate slurry flow properties and flow patterns using a high pressure flow loop
    Ding, Lin
    Shi, Bohui
    Lv, Xiaofang
    Liu, Yang
    Wu, Haihao
    Wang, Wei
    Gong, Jing
    CHEMICAL ENGINEERING SCIENCE, 2016, 146 : 199 - 206
  • [9] Non-invasive flow metering methods for natural gas pipelines
    Zhu, HL
    Min, Z
    PIPELINE & GAS JOURNAL, 2002, 229 (07) : 28 - 32
  • [10] Verification and calibration of natural gas flow measuring system
    Lei, Li
    Zhao, Pujun
    Liu, Wei
    Tianranqi Gongye/Natural Gas Industry, 2002, 22 (04): : 73 - 75